- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- CDC 2019-Novel Coronavirus (2019-nCoV)
- #WakeUp
- 👿💩👎
- Real-Time RT-PCR Diagnostic Panel
- For Emergency Use Only
- Instructions for Use
- Catalog # 2019-nCoVEUA-01
- 1000 reactions
- For In-vitro Diagnostic (IVD) Use
- Rx Only
- https://www.fda.gov/media/134922/downloadCenters for Disease Control and Prevention
- Division of Viral Diseases
- 1600 Clifton Rd NE
- Atlanta GA 30329
- 1
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Table of Contents
- Intended Use ............................................................................................................................. 2
- Summary and Explanation....................................................................................................... 2
- Principles of the Procedure ..................................................................................................... 3
- Materials Required (Provided)................................................................................................. 5
- Materials Required (But Not Provided)................................................................................... 6
- Warnings and Precautions .................................................................................................... 10
- Reagent Storage, Handling, and Stability............................................................................. 11
- Specimen Collection, Handling, and Storage....................................................................... 12
- Specimen Referral to CDC ..................................................................................................... 13
- Reagent and Controls Preparation........................................................................................ 13
- General Preparation ............................................................................................................... 14
- Nucleic Acid Extraction.......................................................................................................... 14
- Assay Set Up........................................................................................................................... 16
- Create a Run Template on the Applied Biosystems 7500 Fast Dx Real-time PCR
- Instrument (Required if no template exists)......................................................................... 20
- Defining the Instrument Settings .......................................................................................... 26
- Running a Test........................................................................................................................ 29
- Interpretation of Results and Reporting............................................................................... 34
- 2019-nCoV rRT-PCR Diagnostic Panel Results Interpretation Guide ................................ 36
- Quality Control........................................................................................................................ 37
- Limitations .............................................................................................................................. 37
- Conditions of Authorization for the Laboratory................................................................... 38
- Performance Characteristics................................................................................................. 39
- Disposal................................................................................................................................... 49
- References .............................................................................................................................. 49
- Revision History ..................................................................................................................... 50
- Contact Information, Ordering, and Product Support......................................................... 50
- Appendix A: Heat Treatment Alternative to Extraction ....................................................... 51
- 2
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Intended Use
- The CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is a real-time RT-PCR
- test intended for the qualitative detection of nucleic acid from the 2019-nCoV in upper and lower
- respiratory specimens (such as nasopharyngeal or oropharyngeal swabs, sputum, lower respiratory tract
- aspirates, bronchoalveolar lavage, and nasopharyngeal wash/aspirate or nasal aspirate) collected from
- individuals who meet 2019-nCoV clinical and/or epidemiological criteria (for example, clinical signs and
- symptoms associated with 2019-nCoV infection, contact with a probable or confirmed 2019-nCoV case,
- history of travel to geographic locations where 2019-nCoV cases were detected, or other epidemiologic
- links for which 2019-nCoV testing may be indicated as part of a public health investigation). Testing in
- the United States is limited to laboratories certified under the Clinical Laboratory Improvement
- Amendments of 1988 (CLIA), 42 U.S.C. § 263a, to perform high complexity tests.
- Results are for the identification of 2019-nCoV RNA. The 2019-nCoV RNA is generally detectable in upper
- and lower respiratory specimens during infection. Positive results are indicative of active infection with
- 2019-nCoV but do not rule out bacterial infection or co-infection with other viruses. The agent detected
- may not be the definite cause of disease. Laboratories within the United States and its territories are
- required to report all positive results to the appropriate public health authorities.
- Negative results do not preclude 2019-nCoV infection and should not be used as the sole basis for
- treatment or other patient management decisions. Negative results must be combined with clinical
- observations, patient history, and epidemiological information.
- Testing with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is intended for use by trained
- laboratory personnel who are proficient in performing real-time RT-PCR assays. The CDC 2019-Novel
- Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is only for use under a Food and Drug
- Administration’s Emergency Use Authorization.
- Summary and Explanation
- An outbreak of pneumonia of unknown etiology in Wuhan City, Hubei Province, China was initially
- reported to WHO on December 31, 2019. Chinese authorities identified a novel coronavirus (2019-
- nCoV), which has resulted in millions of confirmed human infections globally. Cases of asymptomatic
- infection, mild illness, severe illness, and deaths have been reported.
- The CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is a molecular in vitro diagnostic test that aids in
- the detection and diagnosis 2019-nCoV and is based on widely used nucleic acid amplification
- technology. The product contains oligonucleotide primers and dual-labeled hydrolysis probes (TaqMan®)
- and control material used in rRT-PCR for the in vitro qualitative detection of 2019-nCoV RNA in
- respiratory specimens.
- The term “qualified laboratories” refers to laboratories in which all users, analysts, and any person
- reporting results from use of this device should be trained to perform and interpret the results from this
- procedure by a competent instructor prior to use.
- 3
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Principles of the Procedure
- The oligonucleotide primers and probes for detection of 2019-nCoV were selected from regions of the
- virus nucleocapsid (N) gene. The panel is designed for specific detection of the 2019-nCoV (two
- primer/probe sets). An additional primer/probe set to detect the human RNase P gene (RP) in control
- samples and clinical specimens is also included in the panel.
- RNA isolated and purified from upper and lower respiratory specimens is reverse transcribed to cDNA
- and subsequently amplified in the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument with SDS
- version 1.4 software. In the process, the probe anneals to a specific target sequence located between
- the forward and reverse primers. During the extension phase of the PCR cycle, the 5’ nuclease activity of
- Taq polymerase degrades the probe, causing the reporter dye to separate from the quencher dye,
- generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved from
- their respective probes, increasing the fluorescence intensity. Fluorescence intensity is monitored at
- each PCR cycle by Applied Biosystems 7500 Fast Dx Real-Time PCR System with SDS version 1.4 software.
- Detection of viral RNA not only aids in the diagnosis of illness but also provides epidemiological and
- surveillance information.
- 4
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Summary of Preparation and Testing Process
- Resuspend
- primer/probe mix,
- aliquot and store at
- ≤ -20°C
- Resuspend and
- aliquot nCoVPC,
- store at -70°C
- Extract sample RNA
- and HSC RNA
- Prepare master mix
- (15 µL)
- Prepare rRT-PCR
- plate (5 µL RNA)
- Run assay on
- ABI 7500Fast Dx
- Analyze data
- Report results
- Upon receipt of
- rRT-PCR Panel
- reagents
- Upon obtaining
- sample
- 5
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Materials Required (Provided)
- Note: CDC will maintain on its website a list of commercially available lots of primer and probe sets
- and/or positive control materials that are acceptable alternatives to the CDC primer and probe set
- and/or positive control included in the Diagnostic Panel. Only material distributed through the CDC
- International Reagent Resource and specific lots of material posted to the CDC website are acceptable
- for use with this assay under CDC’s Emergency Use Authorization.
- This list of acceptable alternative lots of primer and probe materials and/or positive control materials
- will be available at:
- https://www.cdc.gov/coronavirus/2019-nCoV/lab/virus-requests.html
- Primers and Probes:
- Catalog #2019-nCoVEUA-01 Diagnostic Panel Box #1:
- Reagent Label Part # Description Quantity /
- Tube
- Reactions /
- Tube
- 2019-nCoV_N1 RV202001
- RV202015 2019-nCoV_N1 Combined Primer/Probe Mix 22.5 nmol 1000
- 2019-nCoV_N2 RV202002
- RV202016 2019-nCoV_N2 Combined Primer/Probe Mix 22.5 nmol 1000
- RP RV202004
- RV202018 Human RNase P Combined Primer/Probe Mix 22.5 nmol 1000
- Positive Control (either of the following products are acceptable)
- Catalog #2019-nCoVEUA-01 Diagnostic Panel Box #2:
- Reagent
- Label Part # Description Quantity Notes
- nCoVPC RV202005
- 2019-nCoV Positive Control (nCoVPC)
- For use as a positive control with the CDC 2019-
- nCoV Real-Time RT-PCR Diagnostic Panel
- procedure. The nCoVPC contains noninfectious
- positive control material supplied in a dried state
- and must be resuspended before use. nCoVPC
- consists of in vitro transcribed RNA. nCoVPC will
- yield a positive result with each assay in the
- 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- including RP.
- 4 tubes
- Provides
- (800) 5 µL
- test reactions
- 6
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Catalog #VTC-04 CDC 2019-nCoV Positive Control (nCoVPC)
- Reagent
- Label Part # Description Quantity Notes
- nCoVPC RV202005
- 2019-nCoV Positive Control (nCoVPC)
- For use as a positive control with the CDC 2019-
- nCoV Real-Time RT-PCR Diagnostic Panel
- procedure. The nCoVPC contains noninfectious
- positive control material supplied in a dried state
- and must be resuspended before use. nCoVPC
- consists of in vitro transcribed RNA. nCoVPC will
- yield a positive result with each assay in the
- 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- including RP.
- 4 tubes
- Provides
- (800) 5 µL
- test reactions
- Materials Required (But Not Provided)
- Human Specimen Control (HSC)
- Description Quantity CDC Catalog No.
- Manufactured by CDC. For use as a nucleic acid extraction procedural
- control to demonstrate successful recovery of nucleic acid as well as
- extraction reagent integrity. The HSC consists of noninfectious (betaPropiolactone treated) cultured human cell material supplied as a liquid
- suspended in 0.01 M PBS at pH 7.2-7.4.
- 10 vials x 500uL KT0189
- Acceptable alternatives to HSC:
- • Negative human specimen material: Laboratories may prepare a volume of human specimen
- material (e.g., human sera or pooled leftover negative respiratory specimens) to extract and
- run alongside clinical samples as an extraction control. This material should be prepared in
- sufficient volume to be used across multiple runs. Material should be tested prior to use as
- the extraction control to ensure it generates the expected results for the HSC listed in these
- instructions for use.
- • Contrived human specimen material: Laboratories may prepare contrived human specimen
- materials by suspending any human cell line (e.g., A549, Hela, or 293) in PBS. This material
- should be prepared in sufficient volume to be used across multiple runs. Material should be
- tested prior to use as the extraction control to ensure it generates the expected results for
- the HSC listed in these instructions for use.
- CDC will maintain on its website a list of commercially alternative extraction controls, if applicable,
- that are acceptable for use with this assay under CDC’s Emergency Use Authorization, at:
- https://www.cdc.gov/coronavirus/2019-nCoV/lab/virus-requests.html
- 7
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- rRT-PCR Enzyme Mastermix Options
- Reagent Quantity Catalog No.
- Quantabio qScript XLT One-Step RT-qPCR ToughMix
- 100 x 20 μL rxns
- (1 x 1 mL) 95132-100
- 2000 x 20 μL rxns
- (1 x 20 mL) 95132-02K
- 500 x 20 μL rxns
- (5 x 1 mL) 95132-500
- Quantabio UltraPlex 1-Step ToughMix (4X)
- 100 x 20 µL rxns
- (500 µL) 95166-100
- 500 x 20 μL rxns
- (5 x 500 µL) 95166-500
- 1000 x 20 μL rxns
- (1 x 5 mL) 95166-01K
- Promega GoTaq® Probe 1- Step RT-qPCR System
- 200 x 20 μL rxns
- (2 mL) A6120
- 1250 x 20 μL rxns
- 12.5 mL A6121
- Thermofisher TaqPath™ 1-Step RT-qPCR Master Mix, CG
- 1000 reactions A15299
- 2000 reactions A15300
- 8
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- RNA Extraction Options
- For each of the kits listed below, CDC has confirmed that the external lysis buffer is effective for
- inactivation of SARS-CoV-2.
- Instrument/Manufacturer Extraction Kit Catalog No.
- QIAGEN
- 2
- QIAmp DSP Viral RNA Mini Kit 50 extractions (61904)
- 2
- QIAamp Viral RNA Mini Kit 50 extractions (52904)
- 250 extractions (52906)
- QIAGEN EZ1 Advanced XL
- 2
- EZ1 DSP Virus Kit
- 48 extractions (62724)
- Buffer AVL (19073 or 19089)
- EZ1 Advanced XL DSP Virus Card (9018703)
- 2
- EZ1 Virus Mini Kit v2.0
- 48 extractions (955134)
- Buffer AVL (19073 or 19089)
- EZ1 Advanced XL Virus Card v2.0 (9018708)
- Roche MagNA Pure 24 2
- MagNA Pure 24 Total NA
- Isolation Kit
- 96 extractions (07 658 036 001)
- External Lysis Buffer (06 374 913 001, 12 239 469
- 103, 03 246 779 001 or 03 246 752 001)
- Roche MagNA Pure 96 2
- DNA and Viral NA Small Volume
- Kit
- 576 extractions (06 543 588 001)
- External Lysis Buffer (06 374 913 001, 12 239 469
- 103, 03 246 779 001 or 03 246 752 001)
- 1
- Roche MagNA Pure LC 2
- Total Nucleic Acid Kit 192 extractions (03 038 505 001)
- 1
- Roche MagNA Pure Compact 2
- Nucleic Acid Isolation Kit I 32 extractions (03 730 964 001)
- Promega Maxwell® RSC 48 3
- Maxwell® RSC Viral Total
- Nucleic Acid Purification Kit
- 48 extractions (AS1330)
- 144 extractions (ASB1330)
- 1
- QIAGEN QIAcube
- 2
- QIAmp DSP Viral RNA Mini Kit 50 extractions (61904)
- 2
- QIAamp Viral RNA Mini Kit
- 50 extractions (52904)
- 250 extractions (52906)
- 1, 3bioMérieux NucliSENS®
- easyMAG®
- and
- 1, 3bioMérieux EMAG®
- (Automated magnetic
- extraction reagents sold
- separately. Both instruments
- use the same reagents and
- disposables, with the
- exception of tips.)
- EasyMAG® Magnetic Silica (280133)
- EasyMAG® Lysis Buffer (280134)
- EasyMAG® Lysis Buffer, 2 mL (200292)
- EasyMAG® Wash Buffers 1,2, and 3
- (280130, 280131, 280132)
- EasyMAG® Disposables (280135)
- Biohit Pipette Tips (easyMAG® only) (280146)
- EMAG®1000μL Tips (418922)
- 1
- Equivalence and performance of these extraction platforms for extraction of viral RNA were demonstrated with the CDC
- Human Influenza Virus Real-Time RT-PCR Diagnostic Panel (K190302). Performance characteristics of these extraction
- platforms with 2019-nCoV (SARS CoV-2) have not been demonstrated.
- 9
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 2 CDC has confirmed that the external lysis buffer used with this extraction method is effective for inactivation of SARSCoV-2. 3 CDC has compared the concentration of inactivating agent in the lysis buffer used with this extraction method and has
- determined the concentration to be within the range of concentrations found effective in inactivation of SARS-CoV-2.
- Alternative to Extraction:
- If a laboratory cannot access adequate extraction reagents to support testing demand due to the
- global shortage of reagents, CDC has evaluated a heat treatment procedure for upper respiratory
- specimens using the Quantabio UltraPlex 1-Step ToughMix (4X), CG. Though performance was
- comparable, this method has been evaluated with a limited number of clinical specimens and a
- potential reduction in sensitivity due to carryover of inhibitory substances or RNA degradation cannot
- be ruled out. It should only be used when a jurisdiction determines that the testing need is great
- enough to justify the risk of a potential loss of sensitivity. Heat-treated specimens generating
- inconclusive or invalid results should be extracted with an authorized extraction method prior to
- retesting. Details and procedure for the heat treatment alternative to extraction may be found in
- Appendix A.
- Equipment and Consumables Required (But Not Provided)
- Vortex mixer
- Microcentrifuge
- Micropipettes (2 or 10 μL, 200 μL and 1000 μL)
- Multichannel micropipettes (5-50 μl)
- Racks for 1.5 mL microcentrifuge tubes
- 2 x 96-well -20°C cold blocks
- 7500 Fast Dx Real-Time PCR Systems with SDS 1.4 software (Applied Biosystems; catalog
- #4406985 or #4406984)
- Extraction systems (instruments): QIAGEN EZ1 Advanced XL, QIAGEN QIAcube, Roche MagNA
- Pure 24, Roche MagNA Pure 96, Promega Maxwell® RSC 48, Roche MagNA Pure LC, Roche
- MagNA Pure Compact, bioMérieux easyMAG, and bioMérieux EMAG
- Molecular grade water, nuclease-free
- 10% bleach (1:10 dilution of commercial 5.25-6.0% hypochlorite bleach)
- DNAZapTM (Ambion, cat. #AM9890) or equivalent
- RNase AWAY™ (Fisher Scientific; cat. #21-236-21) or equivalent
- Disposable powder-free gloves and surgical gowns
- Aerosol barrier pipette tips
- 1.5 mL microcentrifuge tubes (DNase/RNase free)
- 0.2 mL PCR reaction plates (Applied Biosystems; catalog #4346906 or #4366932)
- MicroAmp Optical 8-cap Strips (Applied Biosystems; catalog #4323032)
- Qualifying Alternative Components:
- If a laboratory modifies this test by using unauthorized, alternative components (e.g., extraction
- methods or PCR instruments), the modified test is not authorized under this EUA. FDA’s Policy for
- Diagnostic Tests for Coronavirus Disease-2019 during the Public Health Emergency, updated May 11,
- 2020, does not change this. As part of this policy, FDA does not intend to object when a laboratory
- modifies an EUA-authorized test, which could include using unauthorized components, without
- 10
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- obtaining an EUA or EUA amendment, where the modified test is validated using a bridging study to the
- EUA-authorized test .
- Warnings and Precautions
- • For in vitro diagnostic use (IVD).
- • This test has not been FDA cleared or approved; this test has been authorized by FDA under
- an EUA for use by laboratories certified under CLIA, 42 U.S.C. § 263a, to perform high
- complexity tests.
- • This test has been authorized only for the detection of nucleic acid from SARSCoV-2, not
- for any other viruses or pathogens.
- • This test is only authorized for the duration of the declaration that circumstances exist
- justifying the authorization of emergency use of in vitro diagnostic tests for detection
- and/or diagnosis of COVID-19 under Section 564(b)(1) of the Act, 21 U.S.C. § 360bbb3(b)(1), unless the authorization is terminated or revoked sooner.
- • Follow standard precautions. All patient specimens and positive controls should be considered
- potentially infectious and handled accordingly.
- • Do not eat, drink, smoke, apply cosmetics or handle contact lenses in areas where reagents and
- human specimens are handled.
- • Handle all specimens as if infectious using safe laboratory procedures. Refer to Interim
- Laboratory Biosafety Guidelines for Handling and Processing Specimens Associated with 2019-
- nCoV https://www.cdc.gov/coronavirus/2019-nCoV/lab-biosafety-guidelines.html.
- • Specimen processing should be performed in accordance with national biological safety
- regulations.
- • If infection with 2019-nCoV is suspected based on current clinical and epidemiological screening
- criteria recommended by public health authorities, specimens should be collected with
- appropriate infection control precautions.
- • Performance characteristics have been determined with human upper respiratory specimens and
- lower respiratory tract specimens from human patients with signs and symptoms of respiratory
- infection.
- • Perform all manipulations of live virus samples within a Class II (or higher) biological safety
- cabinet (BSC).
- • Use personal protective equipment such as (but not limited to) gloves, eye protection, and lab
- coats when handling kit reagents while performing this assay and handling materials including
- samples, reagents, pipettes, and other equipment and reagents.
- • Amplification technologies such as PCR are sensitive to accidental introduction of PCR product
- from previous amplifications reactions. Incorrect results could occur if either the clinical
- specimen or the real-time reagents used in the amplification step become contaminated by
- accidental introduction of amplification product (amplicon). Workflow in the laboratory should
- proceed in a unidirectional manner.
- Maintain separate areas for assay setup and handling of nucleic acids.
- Always check the expiration date prior to use. Do not use expired reagents. Do not
- substitute or mix reagents from different kit lots or from other manufacturers.
- Change aerosol barrier pipette tips between all manual liquid transfers.
- During preparation of samples, compliance with good laboratory techniques is essential
- to minimize the risk of cross-contamination between samples and the inadvertent
- 11
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- introduction of nucleases into samples during and after the extraction procedure. Proper
- aseptic technique should always be used when working with nucleic acids.
- Maintain separate, dedicated equipment (e.g., pipettes, microcentrifuges) and supplies
- (e.g., microcentrifuge tubes, pipette tips) for assay setup and handling of extracted
- nucleic acids.
- Wear a clean lab coat and powder-free disposable gloves (not previously worn) when
- setting up assays.
- Change gloves between samples and whenever contamination is suspected.
- Keep reagent and reaction tubes capped or covered as much as possible.
- Primers, probes (including aliquots), and enzyme master mix must be thawed and
- maintained on a cold block at all times during preparation and use.
- Work surfaces, pipettes, and centrifuges should be cleaned and decontaminated with
- cleaning products such as 10% bleach, DNAZap™, or RNase AWAY™ to minimize risk of
- nucleic acid contamination. Residual bleach should be removed using 70% ethanol.
- • RNA should be maintained on a cold block or on ice during preparation and use to ensure
- stability.
- • Dispose of unused kit reagents and human specimens according to local, state, and federal
- regulations.
- Reagent Storage, Handling, and Stability
- • Store all dried primers and probes and the positive control, nCoVPC, at 2-8°C until re-hydrated for
- use. Store liquid HSC control materials at ≤ -20°C.
- Note: Storage information is for CDC primer and probe materials obtained through the
- International Reagent Resource. If using commercial primers and probes, please refer to the
- manufacturer’s instructions for storage and handling.
- • Always check the expiration date prior to use. Do not use expired reagents.
- • Protect fluorogenic probes from light.
- • Primers, probes (including aliquots), and enzyme master mix must be thawed and kept on a cold
- block at all times during preparation and use.
- • Do not refreeze probes.
- • Controls and aliquots of controls must be thawed and kept on ice at all times during preparation
- and use.
- 12
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Specimen Collection, Handling, and Storage
- Inadequate or inappropriate specimen collection, storage, and transport are likely to yield false test
- results. Training in specimen collection is highly recommended due to the importance of specimen
- quality. CLSI MM13-A may be referenced as an appropriate resource.
- Collecting the Specimen
- • Refer to Interim Guidelines for Collecting, Handling, and Testing Clinical Specimens from
- Patients Under Investigation (PUIs) for 2019 Novel Coronavirus (2019-nCoV)
- https://www.cdc.gov/coronavirus/2019-nCoV/guidelines-clinical-specimens.html
- • Follow specimen collection device manufacturer instructions for proper collection methods.
- • Swab specimens should be collected using only swabs with a synthetic tip, such as nylon or
- Dacron®, and an aluminum or plastic shaft. Calcium alginate swabs are unacceptable and
- cotton swabs with wooden shafts are not recommended. Place swabs immediately into
- sterile tubes containing 1-3 ml of appropriate transport media, such as viral transport media
- (VTM).
- Transporting Specimens
- • Specimens must be packaged, shipped, and transported according to the current edition of
- the International Air Transport Association (IATA) Dangerous Goods Regulation. Follow
- shipping regulations for UN 3373 Biological Substance, Category B when sending potential
- 2019-nCoV specimens. Store specimens at 2-8°C and ship overnight to CDC on ice pack. If a
- specimen is frozen at -70°C or lower, ship overnight to CDC on dry ice.
- Storing Specimens
- • Specimens can be stored at 2-8oC for up to 72 hours after collection.
- • If a delay in extraction is expected, store specimens at -70oC or lower.
- • Extracted nucleic acid should be stored at -70oC or lower.
- 13
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Specimen Referral to CDC
- For state and local public health laboratories:
- • Ship all specimens overnight to CDC.
- • Ship frozen specimens on dry ice and non-frozen specimens on cold packs.
- • Refer to the International Air Transport Association (IATA - www.iata.org) for requirements
- for shipment of human or potentially infectious biological specimens. Follow shipping
- regulations for UN 3373 Biological Substance, Category B when sending potential 2019-nCoV
- specimens.
- • Prior to shipping, notify CDC Division of Viral Diseases (see contact information below) that
- you are sending specimens.
- • Send all samples to the following recipient:
- Centers for Disease Control and Prevention
- c/o STATT
- Attention: Unit 66
- 1600 Clifton Rd., Atlanta, GA 30329-4027
- Phone: (404) 639-3931
- The emergency contact number for CDC Emergency Operations Center (EOC) is
- 770-488-7100.
- All other laboratories that are CLIA certified and meet requirements to perform high complexity
- testing:
- • Please notify your state and/or local public health laboratory for specimen referral and
- confirmatory testing guidance.
- Reagent and Controls Preparation
- NOTE: Storage information is for materials obtained through the CDC International Regent Resource.
- If using commercial products for testing, please refer to the manufacturer’s instructions for storage,
- handling, and preparation instructions.
- Primer and Probe Preparation:
- 1) Upon receipt, store dried primers and probes at 2-8°C.
- 2) Precautions: These reagents should only be handled in a clean area and stored at appropriate
- temperatures (see below) in the dark. Freeze-thaw cycles should be avoided. Maintain cold
- when thawed.
- 3) Using aseptic technique, suspend dried reagents in 1.5 mL of nuclease-free water and allow
- to rehydrate for 15 min at room temperature in the dark.
- 4) Mix gently and aliquot primers/probe in 300 μL volumes into 5 pre-labeled tubes. Store a
- single, working aliquot of primers/probes at 2-8oC in the dark. Store remaining aliquots at ≤ -
- 20oC in a non-frost-free freezer. Do not refreeze thawed aliquots (stable for up to 4 months at
- 2-8oC).
- 14
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 2019-nCoV Positive Control (nCoVPC) Preparation:
- 1) Precautions: This reagent should be handled with caution in a dedicated nucleic acid handling
- area to prevent possible contamination. Freeze-thaw cycles should be avoided. Maintain on
- ice when thawed.
- 2) Resuspend dried reagent in each tube in 1 mL of nuclease-free water to achieve the proper
- concentration. Make single use aliquots (approximately 30 μL) and store at ≤ -70oC.
- 3) Thaw a single aliquot of diluted positive control for each experiment and hold on ice until
- adding to plate. Discard any unused portion of the aliquot.
- Human Specimen Control (HSC) (not provided)
- 1) Human Specimen Control (HSC) or one of the listed acceptable alternative extraction controls
- must be extracted and processed with each specimen extraction run.
- 2) Refer to the Human Specimen Control (HSC) package insert for instructions for use.
- No Template Control (NTC) (not provided)
- 1) Sterile, nuclease-free water
- 2) Aliquot in small volumes
- 3) Used to check for contamination during specimen extraction and/or plate set-up
- General Preparation
- Equipment Preparation
- Clean and decontaminate all work surfaces, pipettes, centrifuges, and other equipment prior to use.
- Decontamination agents should be used including 10% bleach, 70% ethanol, and DNAzap™, or RNase
- AWAY™ to minimize the risk of nucleic acid contamination.
- Nucleic Acid Extraction
- Performance of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is dependent upon the amount
- and quality of template RNA purified from human specimens. The following commercially available
- RNA extraction kits and procedures have been qualified and validated for recovery and purity of RNA
- for use with the panel:
- Qiagen QIAamp® DSP Viral RNA Mini Kit or QIAamp® Viral RNA Mini Kit
- Recommendation(s): Utilize 100 μL of sample and elute with 100 μL of buffer or utilize 140 μL of
- sample and elute with 140 μL of buffer.
- Qiagen EZ1 Advanced XL
- Kit: Qiagen EZ1 DSP Virus Kit and Buffer AVL (supplied separately) for offboard lysis
- Card: EZ1 Advanced XL DSP Virus Card
- Recommendation(s): Add 120 μL of sample to 280 μL of pre-aliquoted Buffer AVL (total input sample
- volume is 400 μL). Proceed with the extraction on the EZ1 Advanced XL. Elution volume is 120 μL.
- 15
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Kit: Qiagen EZ1 Virus Mini Kit v2.0 and Buffer AVL (supplied separately) for offboard lysis
- Card: EZ1 Advanced XL Virus Card v2.0
- Recommendation(s): Add 120 μL of sample to 280 μL of pre-aliquoted Buffer AVL (total input sample
- volume is 400 μL). Proceed with the extraction on the EZ1 Advanced XL. Elution volume is 120 μL.
- Roche MagNA Pure 96
- Kit: Roche MagNA Pure 96 DNA and Viral NA Small Volume Kit
- Protocol: Viral NA Plasma Ext LysExt Lys SV 4.0 Protocol or Viral NA Plasma Ext Lys SV Protocol
- Recommendation(s): Add 100 μL of sample to 350 μL of pre-aliquoted External Lysis Buffer (supplied
- separately) (total input sample volume is 450 μL). Proceed with the extraction on the MagNA Pure 96.
- (Internal Control = None). Elution volume is 100 μL.
- Roche MagNA Pure 24
- Kit: Roche MagNA Pure 24 Total NA Isolation Kit
- Protocol: Pathogen 1000 2.0 Protocol
- Recommendation(s): Add 100 µL of sample to 400 µL of pre-aliquoted External Lysis Buffer (supplied
- separately) (total input sample volume is 500 µL). Proceed with the extraction on the MagNA Pure 24.
- (Internal Control = None). Elution volume is 100 µL.
- Promega Maxwell® RSC 48
- Kit: Promega Maxwell® Viral Total Nucleic Acid Purification Kit
- Protocol: Viral Total Nucleic Acid
- Recommendation(s): Add 120 µL of sample to 330 µL of pre-aliquoted External Lysis Buffer (300 µL
- Lysis Buffer plus 30 µL Proteinase K; supplied within the kit) (total input volume is 450 µL). Proceed
- with the extraction on the Maxwell® RSC 48. Elution volume is 75 µL.
- Equivalence and performance of the following extraction platforms were demonstrated with the CDC
- Human Influenza Virus Real-Time RT-PCR Diagnostic Panel (K190302) and based on those data are
- acceptable for use with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel.
- QIAGEN QIAcube
- Kit: QIAGEN QIAamp® DSP Viral RNA Mini Kit or QIAamp® Viral RNA Mini Kit
- Recommendations: Utilize 140 μL of sample and elute with 100 μL of buffer.
- Roche MagNA Pure LC
- Kit: Roche MagNA Pure Total Nucleic Acid Kit
- Protocol: Total NA External_lysis
- Recommendation(s): Add 100 μL of sample to 300 μL of pre-aliquoted TNA isolation kit lysis buffer
- (total input sample volume is 400 μL). Elution volume is 100 μL.
- Roche MagNA Pure Compact
- Kit: Roche MagNA Pure Nucleic Acid Isolation Kit I
- Protocol: Total_NA_Plasma100_400
- Recommendation(s): Add 100 μL of sample to 300 μL of pre-aliquoted TNA isolation kit lysis buffer
- (total input sample volume is 400 μL). Elution volume is 100 μL.
- 16
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- bioMérieux NucliSENS® easyMAG® Instrument
- Protocol: General protocol (not for blood) using “Off-board Lysis” reagent settings.
- Recommendation(s): Add 100 μL of sample to 1000 μL of pre-aliquoted easyMAG lysis buffer (total
- input sample volume is 1100 μL). Incubate for 10 minutes at room temperature. Elution volume is 100
- μL.
- bioMérieux EMAG® Instrument
- Protocol: Custom protocol: CDC Flu V1 using “Off-board Lysis” reagent settings.
- Recommendation(s): Add 100 μL of samples to 2000 μL of pre-aliquoted easyMAG lysis buffer (total
- input sample volume is 2100 μL). Incubate for 10 minutes at room temperature. Elution volume is 100
- μL. The custom protocol, CDC Flu V1, is programmed on the bioMérieux EMAG® instrument with the
- assistance of a bioMérieux service representative. Installation verification is documented at the time of
- installation. Laboratories are recommended to retain a record of the step-by-step verification of the
- bioMérieux custom protocol installation procedure.
- Manufacturer’s recommended procedures (except as noted in recommendations above) are to be
- followed for sample extraction. HSC must be included in each extraction batch.
- Disclaimer: Names of vendors or manufacturers are provided as examples of suitable product sources. Inclusion
- does not imply endorsement by the Centers for Disease Control and Prevention.
- Assay Set Up
- Reaction Master Mix and Plate Set Up
- Note: Plate set-up configuration can vary with the number of specimens and workday
- organization. NTCs and nCoVPCs must be included in each run.
- 1) In the reagent set-up room clean hood, place rRT-PCR buffer, enzyme, and primer/probes on
- ice or cold-block. Keep cold during preparation and use.
- 2) Mix buffer, enzyme, and primer/probes by inversion 5 times.
- 3) Centrifuge reagents and primers/probes for 5 seconds to collect contents at the bottom of
- the tube, and then place the tube in a cold rack.
- 4) Label one 1.5 mL microcentrifuge tube for each primer/probe set.
- 5) Determine the number of reactions (N) to set up per assay. It is necessary to make excess
- reaction mix for the NTC, nCoVPC, HSC (if included in the RT-PCR run), and RP reactions and
- for pipetting error. Use the following guide to determine N:
- • If number of samples (n) including controls equals 1 through 14, then N = n + 1
- • If number of samples (n) including controls is 15 or greater, then N = n + 2
- 7) For each primer/probe set, calculate the amount of each reagent to be added for each
- reaction mixture (N = # of reactions).
- 17
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Thermofisher TaqPath™ 1-Step RT-qPCR Master Mix
- Step # Reagent Vol. of Reagent Added
- per Reaction
- 1 Nuclease-free Water N x 8.5 µL
- 2 Combined Primer/Probe Mix N x 1.5 µL
- 3 TaqPathTM 1-Step RT-qPCR Master Mix (4x) N x 5.0 µL
- Total Volume N x 15.0 µL
- Promega GoTaq® Probe 1- Step RT-qPCR System
- Step # Reagent Vol. of Reagent Added
- per Reaction
- 1 Nuclease-free Water N x 3.1 µL
- 2 Combined Primer/Probe Mix N x 1.5 µL
- 3 GoTaq Probe qPCR Master Mix with dUTP N x 10.0 µL
- 4 Go Script RT Mix for 1-Step RT-qPCR N x 0.4 µL
- Total Volume N x 15.0 µL
- Quantabio qScript XLT One-Step RT-qPCR ToughMix
- Step # Reagent Vol. of Reagent Added
- per Reaction
- 1 Nuclease-free Water N x 3.5 µL
- 2 Combined Primer/Probe Mix N x 1.5 µL
- 3 qScript XLT One-Step RT-qPCR ToughMix
- (2X)
- N x 10.0 µL
- Total Volume N x 15.0 µL
- Quantabio UltraPlex 1-Step ToughMix (4X)
- Step # Reagent Vol. of Reagent Added
- per Reaction
- 1 Nuclease-free Water N x 8.5 µL
- 2 Combined Primer/Probe Mix N x 1.5 µL
- 3 UltraPlex 1-Step ToughMix (4X) N x 5.0 µL
- Total Volume N x 15.0 µL
- 18
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 8) Dispense reagents into each respective labeled 1.5 mL microcentrifuge tube. After addition of
- the reagents, mix reaction mixtures by pipetting up and down. Do not vortex.
- 9) Centrifuge for 5 seconds to collect contents at the bottom of the tube, and then place the
- tube in a cold rack.
- 10) Set up reaction strip tubes or plates in a 96-well cooler rack.
- 11) Dispense 15 µL of each master mix into the appropriate wells going across the row as shown
- below (Figure 1):
- Figure 1: Example of Reaction Master Mix Plate Set-Up
- 1 2 3 4 5 6 7 8 9 10 11 12
- A N1 N1 N1 N1 N1 N1 N1 N1 N1 N1 N1 N1
- B N2 N2 N2 N2 N2 N2 N2 N2 N2 N2 N2 N2
- C RP RP RP RP RP RP RP RP RP RP RP RP
- D
- E
- F
- G
- H
- 12) Prior to moving to the nucleic acid handling area, prepare the No Template Control (NTC)
- reactions for column #1 in the assay preparation area.
- 13) Pipette 5 µL of nuclease-free water into the NTC sample wells (Figure 2, column 1). Securely
- cap NTC wells before proceeding.
- 14) Cover the entire reaction plate and move the reaction plate to the specimen nucleic acid
- handling area.
- Nucleic Acid Template Addition
- 1) Gently vortex nucleic acid sample tubes for approximately 5 seconds.
- 2) Centrifuge for 5 seconds to collect contents at the bottom of the tube.
- 3) After centrifugation, place extracted nucleic acid sample tubes in the cold rack.
- 4) Samples should be added to columns 2-11 (column 1 and 12 are for controls) to the specific
- assay that is being tested as illustrated in Figure 2. Carefully pipette 5.0 µL of the first sample
- into all the wells labeled for that sample (i.e. Sample “S1” down column #2). Keep other
- sample wells covered during addition. Change tips after each addition.
- 5) Securely cap the column to which the sample has been added to prevent cross contamination
- and to ensure sample tracking.
- 6) Change gloves often and when necessary to avoid contamination.
- 7) Repeat steps #4 and #5 for the remaining samples.
- 19
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 8) If necessary, add 5 µL of Human Specimen Control (HSC) extracted sample to the HSC wells
- (Figure 2, column 11). Securely cap wells after addition. NOTE: Per CLIA regulations, HSC
- must be tested at least once per day.
- 9) Cover the entire reaction plate and move the reaction plate to the positive template control
- handling area.
- Assay Control Addition
- 1) Pipette 5 µL of nCoVPC RNA to the sample wells of column 12 (Figure 2). Securely cap wells
- after addition of the control RNA.
- NOTE: If using 8-tube strips, label the TAB of each strip to indicate sample position. DO NOT
- LABEL THE TOPS OF THE REACTION TUBES!
- 2) Briefly centrifuge reaction tube strips for 10-15 seconds. After centrifugation return to cold
- rack.
- NOTE: If using 96-well plates, centrifuge plates for 30 seconds at 500 x g, 4°C.
- Figure 2. 2019-nCoV rRT-PCR Diagnostic Panel: Example of Sample and Control Set-up
- 1 2 3 4 5 6 7 8 9 10 11a 12
- A NTC S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 nCoV PC
- B NTC S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 nCoV PC
- C NTC S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 nCoV PC
- D
- E
- F
- G
- H
- a
- Replace the sample in this column with extracted HSC if necessary
- 20
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Create a Run Template on the Applied Biosystems 7500 Fast Dx Real-time PCR
- Instrument (Required if no template exists)
- If the template already exists on your instrument, please proceed to the RUNNING A TEST section.
- 1) Launch the Applied Biosystems 7500 Fast Dx Real-time PCR Instrument by double clicking on the
- Applied Biosystems 7500 Fast Dx System icon on the desktop.
- 2) A new window should appear, select Create New Document from the menu.
- Figure 3. New Document Wizard Window
- 3) The New Document Wizard screen in Figure 3 will appear. Select:
- a. Assay: Standard Curve (Absolute Quantitation)
- b. Container: 96-Well Clear
- c. Template: Blank Document
- d. Run Mode: Standard 7500
- e. Operator: Your Name
- f. Comments: SDS v1.4
- g. Plate Name: Your Choice
- 4) After making selections click Next at the bottom of the window.
- Make sure to change
- Run Mode to
- STANDARD 7500
- 21
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Figure 4. Creating New Detectors
- NOTE: ROX is the default passive reference. This will be changed to “none” in step 12.
- 5) After selecting next, the Select Detectors screen (Figure 4) will appear.
- 6) Click the New Detector button (see Figure 4).
- 7) The New Detector window will appear (Figure 5). A new detector will need to be defined for
- each primer and probe set. Creating these detectors will enable you to analyze each primer and
- probe set individually at the end of the reaction.
- 22
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Figure 5. New Detector Window
- 8) Start by creating the N1 Detector. Include the following:
- a. Name: N1
- b. Description: leave blank
- c. Reporter Dye: FAM
- d. Quencher Dye: (none)
- e. Color: to change the color of the detector indicator do the following:
- ⇒ Click on the color square to reveal the color chart
- ⇒ Select a color by clicking on one of the squares
- ⇒ After selecting a color click OK to return to the New Detector screen
- f. Click the OK button of the New Detector screen to return to the screen shown in Figure 4.
- 9) Repeat step 6-8 for each target in the panel.
- Name Reporter Dye Quencher Dye
- N1 FAM (none)
- N2 FAM (none)
- RP FAM (none)
- 23
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 10) After each Detector is added, the Detector Name, Description, Reporter and Quencher fields
- will become populated in the Select Detectors screen (Figure 6).
- 11) Before proceeding, the newly created detectors must be added to the document. To add the
- new detectors to the document, click ADD (see Figure 6). Detector names will appear on the
- right-hand side of the Select Detectors window (Figure 6).
- Figure 6. Adding New Detectors to Document
- 12) Once all detectors have been added, select (none) for Passive Reference at the top right-hand
- drop-down menu (Figure 7).
- Figure 7. Select Passive Reference
- Passive reference should be set to “(none)” as described above.
- 24
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 13)Click Next at the bottom of the Select Detectors window to proceed to the Set Up Sample Plate
- window (Figure 8).
- 14)In the Set Up Sample Plate window (Figure 8), use your mouse to select row A from the lower
- portion of the window, in the spreadsheet (see Figure 8).
- 15)In the top portion of the window, select detector N1. A check will appear next to the detector
- you have selected (Figure 8). You will also notice the row in the spreadsheet will be populated
- with a colored “U” icon to indicate which detector you’ve selected.
- 16)Repeat step 14-15 for each detector that will be used in the assay.
- Figure 8. Sample Plate Set-up
- 17) Select Finish after detectors have been assigned to their respective rows. (Figure 9).
- Figure 9. Finished Plate Set-up
- 25
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 18) After clicking “Finish”, there will be a brief pause allowing the Applied Biosystems 7500 Fast Dx
- to initialize. This initialization is followed by a clicking noise. Note: The machine must be turned
- on for initialization.
- 19) After initialization, the Plate tab of the Setup (Figure 10) will appear.
- 20)Each well of the plate should contain colored U icons that correspond with the detector labels
- that were previously chosen. To confirm detector assignments, select Tools from the file menu,
- then select Detector Manager.
- Figure 10. Plate Set-up Window
- 26
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 21) The Detector Manager window will appear (Figure 11).
- Figure 11. Detector Manager Window
- 22)Confirm all detectors are included and that each target has a Reporter set to FAM and the
- Quencher is set to (none).
- 23)If all detectors are present, select Done. The detector information has been created and
- assigned to wells on the plate.
- Defining the Instrument Settings
- 1) After detectors have been created and assigned, proceed to instrument set up.
- 2) Select the Instrument tab to define thermal cycling conditions.
- 3) Modify the thermal cycling conditions as follows (Figure 12):
- Thermofisher TaqPath™ 1-Step RT-qPCR Master Mix, CG
- a. In Stage 1, Set to 2 min at 25°C; 1 Rep.
- b. In Stage 2, Set to 15 min at 50°C; 1 Rep.
- c. In Stage 3, Set to 2 min at 95°C, 1 Rep.
- d. In Stage 4, Step 1 set to 3 sec at 95°C.
- e. In Stage 4, Step 2 set to 30 sec at 55.0°C.
- f. In Stage 4, Reps should be set to 45.
- g. Under Settings (Figure 12), bottom left-hand box, change volume to 20 µL.
- h. Under Settings, Run Mode selection should be Standard 7500.
- i. Step 2 of Stage 4 should be highlighted in yellow to indicate data collection (see Figure
- 12).
- OR
- 27
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Quantabio qScriptTM XLT One-Step RT-qPCR ToughMix or UltraPlex 1-Step ToughMix (4X)
- a. In Stage 1, Set to 10 min at 50°C; 1 Rep.
- b. In Stage 2, Set to 3 min at 95°C, 1 Rep.
- c. In Stage 3, Step 1 set to 3 sec at 95°C.
- d. In Stage 3, Step 2 set to 30 sec at 55.0°C.
- e. In Stage 3, Reps should be set to 45.
- f. Under Settings (Figure 12), bottom left-hand box, change volume to 20 µL.
- g. Under Settings, Run Mode selection should be Standard 7500.
- h. Step 2 of Stage 3 should be highlighted in yellow to indicate data collection (see Figure
- 12).
- OR
- Promega GoTaq® Probe 1-Step RT-qPCR System
- a. In Stage 1, Set to 15 min at 45°C; 1 Rep.
- b. In Stage 2, Set to 2 min at 95°C, 1 Rep.
- c. In Stage 3, Step 1 set to 3 sec at 95°C.
- d. In Stage 3, Step 2 set to 30 sec at 55.0°C.
- e. In Stage 3, Reps should be set to 45.
- f. Under Settings (Figure 12), bottom left-hand box, change volume to 20 µL.
- g. Under Settings, Run Mode selection should be Standard 7500.
- h. Step 2 of Stage 3 should be highlighted in yellow to indicate data collection (see Figure
- 12).
- 28
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Figure 12. Instrument Window
- 4) After making changes to the Instrument tab, the template file is ready to be saved. To save the template,
- select File from the top menu, then select Save As. Since the enzyme options have different instrument
- settings, it is recommended that the template be saved with a name indicating the enzyme option.
- 5) Save the template as 2019-nCoV Dx Panel TaqPath or 2019-nCoV Dx Panel Quanta or 2019-nCoV Dx
- Panel Promega as appropriate in the desktop folder labeled “ABI Run Templates” (you must create this
- folder). Save as type should be SDS Templates (*.sdt) (Figure 13).
- Figure 13. Saving Template
- 29
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Running a Test
- 1) Turn on the ABI 7500 Fast Dx Real-Time PCR Instrument.
- 2) Launch the Applied Biosystems 7500 Fast Dx Real-time PCR System by double clicking on the
- 7500 Fast Dx System icon on the desktop.
- 3) A new window should appear, select Open Existing Document from the menu.
- 4) Navigate to select your ABI Run Template folder from the desktop.
- 5) Double click on the appropriate template file (2019-nCoV Dx Panel TaqPath or 2019-nCoV Dx
- Panel Quanta or 2019-nCoV Dx Panel Promega)
- 6) There will be a brief pause allowing the Applied Biosystems 7500 Fast Dx Real-Time PCR
- Instrument to initialize. This initialization is followed by a clicking noise. Note: The machine must
- be turned on for initialization.
- Figure 14. Plate Set-up Window
- 7) After the instrument initializes, a plate map will appear (Figure 14). The detectors and controls
- should already be labeled as they were assigned in the original template.
- 30
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 8) Click the Well Inspector icon from the top menu.
- 9) Highlight specimen wells of interest on the plate map.
- 10) Type sample identifiers to Sample Name box in the Well Inspector window (Figure 15).
- Figure 15. Labeling Wells
- 11) Repeat steps 9-10 until all sample identifiers are added to the plate setup.
- 31
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 12) Once all specimen and control identifiers are added click the Close button on the Well Inspector
- window to return to the Plate set up tab.
- 13) Click the Instrument tab at the upper left corner.
- 14) The reaction conditions, volumes, and type of 7500 reaction should already be loaded (Figure
- 16).
- Figure 16. Instrument Settings
- 15) Ensure settings are correct (refer to the Defining Instrument Settings).
- 16) Before proceeding, the run file must be saved; from the main menu, select File, then Save As.
- Save in appropriate run folder designation.
- 17) Load the plate into the plate holder in the instrument. Ensure that the plate is properly aligned
- in the holder.
- 18) Once the run file is saved, click the Start button. Note: The run should take approximately 1 hour
- and 20 minutes to complete.
- 32
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Data Analysis
- 1) After the run has completed, select the Results tab at the upper left corner of the software.
- 2) Select the Amplification Plot tab to view the raw data (Figure 17).
- Figure 17. Amplification Plot Window
- 3) Start by highlighting all the samples from the run; to do this, click on the upper left-hand box (a)
- of the sample wells (Figure 17). All the growth curves should appear on the graph.
- 4) On the right-hand side of the window (b), the Data drop down selection should be set to Delta
- Rn vs. Cycle.
- 5) Select N1 from (c), the Detector drop down menu, using the downward arrow.
- a. Please note that each detector is analyzed individually to reflect different
- performance profiles of each primer and probe set.
- 6) In the Line Color drop down (d), Detector Color should be selected.
- 7) Under Analysis Settings select Manual Ct (e).
- b. Do not change the Manual Baseline default numbers.
- 8) Using the mouse, click and drag the red threshold line until it lies within the exponential phase
- of the fluorescence curves and above any background signal (Figure 18).
- c
- c
- a
- b
- c
- d
- e
- 33
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Figure 18. Amplification Plot
- 9) Click the Analyze button in the lower right corner of the window. The red threshold line will turn
- to green, indicating the data has been analyzed.
- 10) Repeat steps 5-9 to analyze results generated for each set of markers (N1, N2, RP).
- 11) Save analysis file by selecting File then Save As from the main menu.
- 12) After completing analysis for each of the markers, select the Report tab above the graph to
- display the Ct values (Figure 19). To filter report by sample name in ascending or descending
- order, simply click on Sample Name in the table.
- Figure 19. Report
- Exponential
- PCR Phase
- Background noise
- Threshold adjusted
- to fall within the
- PCR exponential
- phase.
- 34
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Interpretation of Results and Reporting
- Extraction and Positive Control Results and Interpretation
- No Template Control (NTC)
- The NTC consists of using nuclease-free water in the rRT-PCR reactions instead of RNA. The NTC
- reactions for all primer and probe sets should not exhibit fluorescence growth curves that cross the
- threshold line. If any of the NTC reactions exhibit a growth curve that crosses the cycle threshold, sample
- contamination may have occurred. Invalidate the run and repeat the assay with strict adherence to the
- guidelines.
- 2019-nCoV Positive Control (nCoVPC)
- The nCoVPC consists of in vitro transcribed RNA. The nCoVPC will yield a positive result with the
- following primer and probe sets: N1, N2, and RP.
- Human Specimen Control (HSC) (Extraction Control)
- When HSC is run with the CDC 2019-nCoV rRT-PCR Diagnostic Panel (see previous section on Assay Set
- Up), the HSC is used as an nucleic acid extraction procedural control to demonstrate successful recovery
- of nucleic acid as well as extraction reagent integrity. The HSC control consists of noninfectious cultured
- human cell (A549) material. Purified nucleic acid from the HSC should yield a positive result with the RP
- primer and probe set and negative results with all 2019-nCoV markers.
- Expected Performance of Controls Included in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Control
- Type
- External
- Control
- Name
- Used to
- Monitor
- 2019
- nCoV_N1
- 2019
- nCoV_N2 RP Expected Ct
- Values
- Positive nCoVPC
- Substantial
- reagent failure
- including
- primer and
- probe integrity
- + + + < 40.00 Ct
- Negative NTC
- Reagent and/or
- environmental
- contamination
- - - - None
- detected
- Extraction HSC
- Failure in lysis
- and extraction
- procedure,
- potential
- contamination
- during
- extraction
- - - + < 40.00 Ct
- 35
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- If any of the above controls do not exhibit the expected performance as described, the assay may have
- been set up and/or executed improperly, or reagent or equipment malfunction could have occurred.
- Invalidate the run and re-test.
- RNase P (Extraction Control)
- All clinical samples should exhibit fluorescence growth curves in the RNase P reaction that cross
- the threshold line within 40.00 cycles (< 40.00 Ct), thus indicating the presence of the human
- RNase P gene. Failure to detect RNase P in any clinical specimens may indicate:
- − Improper extraction of nucleic acid from clinical materials resulting in loss of RNA and/or
- RNA degradation.
- − Absence of sufficient human cellular material due to poor collection or loss of specimen
- integrity.
- − Improper assay set up and execution.
- − Reagent or equipment malfunction.
- If the RP assay does not produce a positive result for human clinical specimens, interpret as
- follows:
- − If the 2019-nCoV N1 and N2are positive even in the absence of a positive RP, the result
- should be considered valid. It is possible, that some samples may fail to exhibit RNase P
- growth curves due to low cell numbers in the original clinical sample. A negative RP signal
- does not preclude the presence of 2019-nCoV virus RNA in a clinical specimen.
- − If all 2019-nCoV markers AND RNase P are negative for the specimen, the result should be
- considered invalid for the specimen. If residual specimen is available, repeat the extraction
- procedure and repeat the test. If all markers remain negative after re-test, report the
- results as invalid and a new specimen should be collected if possible.
- 2019-nCoV Markers (N1 and N2)
- • When all controls exhibit the expected performance, a specimen is considered negative if all
- 2019-nCoV marker (N1, N2) cycle threshold growth curves DO NOT cross the threshold line
- within 40.00 cycles (< 40.00 Ct) AND the RNase P growth curve DOES cross the threshold line
- within 40.00 cycles (< 40.00 Ct).
- • When all controls exhibit the expected performance, a specimen is considered positive for 2019-
- nCoV if all 2019-nCoV marker (N1, N2) cycle threshold growth curves cross the threshold line
- within 40.00 cycles (< 40.00 Ct). The RNase P may or may not be positive as described above, but
- the 2019-nCoV result is still valid.
- • When all controls exhibit the expected performance and the growth curves for the 2019-nCoV
- markers (N1, N2) AND the RNase P marker DO NOT cross the cycle threshold growth curve within
- 40.00 cycles (< 40.00 Ct), the result is invalid. The extracted RNA from the specimen should be retested. If residual RNA is not available, re-extract RNA from residual specimen and re-test. If the
- re-tested sample is negative for all markers and RNase P, the result is invalid and collection of a
- new specimen from the patient should be considered.
- • When all controls exhibit the expected performance and the cycle threshold growth curve for any
- one marker (N1 or N2, but not both markers) crosses the threshold line within 40.00 cycles (<
- 40.00 Ct) the result is inconclusive. The extracted RNA should be retested. If residual RNA is not
- available, re-extract RNA from residual specimen and re-test. If the same result is obtained,
- 36
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- report the inconclusive result. Consult with your state public health laboratory or CDC, as
- appropriate, to request guidance and/or to coordinate transfer of the specimen for additional
- analysis.
- • If HSC is positive for N1 or N2, then contamination may have occurred during extraction or
- sample processing. Invalidate all results for specimens extracted alongside the HSC. Re-extract
- specimens and HSC and re-test.
- 2019-nCoV rRT-PCR Diagnostic Panel Results Interpretation Guide
- The table below lists the expected results for the 2019-nCoV rRT-PCR Diagnostic Panel. If a laboratory
- obtains unexpected results for assay controls or if inconclusive or invalid results are obtained and cannot
- be resolved through the recommended re-testing, please contact CDC for consultation and possible
- specimen referral. See pages 13 and 50 for referral and contact information.
- 2019
- nCoV_N1
- 2019
- nCoV_N2 RP Result
- Interpretationa Report Actions
- + + ±
- 2019-nCoV
- detected Positive 2019-nCoV
- Report results to CDC and
- sender.
- If only one of the two
- targets is positive ±
- Inconclusive
- Result Inconclusive
- Repeat testing of nucleic acid
- and/or re-extract and repeat
- rRT-PCR. If the repeated result
- remains inconclusive, contact
- your State Public Health
- Laboratory or CDC for
- instructions for transfer of the
- specimen or further guidance.
- - - +
- 2019-nCoV not
- detected Not Detected
- Report results to sender.
- Consider testing for other
- respiratory viruses.b
- - - - Invalid Result Invalid
- Repeat extraction and rRT-PCR.
- If the repeated result remains
- invalid, consider collecting a
- new specimen from the patient. aLaboratories should report their diagnostic result as appropriate and in compliance with their specific reporting
- system.
- bOptimum specimen types and timing for peak viral levels during infections caused by 2019-nCoV have not been
- determined. Collection of multiple specimens from the same patient may be necessary to detect the virus. The
- possibility of a false negative result should especially be considered if the patient’s recent exposures or clinical
- presentation suggest that 2019-nCoV infection is possible, and diagnostic tests for other causes of illness (e.g.,
- other respiratory illness) are negative. If 2019-nCoV infection is still suspected, re-testing should be considered
- in consultation with public health authorities.
- 37
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Quality Control
- • Quality control requirements must be performed in conformance with local, state, and federal
- regulations or accreditation requirements and the user’s laboratory’s standard quality control
- procedures. For further guidance on appropriate quality control practices, refer to 42 CFR
- 493.1256.
- • Quality control procedures are intended to monitor reagent and assay performance.
- • Test all positive controls prior to running diagnostic samples with each new kit lot to ensure all
- reagents and kit components are working properly.
- • Good laboratory practice (cGLP) recommends including a positive extraction control in each
- nucleic acid isolation batch.
- • Although HSC is not included with the 2019-nCov rRT-PCR Diagnostic Panel, the HSC extraction
- control must proceed through nucleic acid isolation per batch of specimens to be tested.
- • Always include a negative template control (NTC) and the appropriate positive control (nCoVPC)
- in each amplification and detection run. All clinical samples should be tested for human RNase P
- gene to control for specimen quality and extraction.
- Limitations
- • All users, analysts, and any person reporting diagnostic results should be trained to perform this
- procedure by a competent instructor. They should demonstrate their ability to perform the test
- and interpret the results prior to performing the assay independently.
- • Performance of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel has only been established
- in upper and lower respiratory specimens (such as nasopharyngeal or oropharyngeal swabs,
- sputum, lower respiratory tract aspirates, bronchoalveolar lavage, and nasopharyngeal
- wash/aspirate or nasal aspirate).
- • Negative results do not preclude 2019-nCoV infection and should not be used as the sole basis
- for treatment or other patient management decisions. Optimum specimen types and timing for
- peak viral levels during infections caused by 2019-nCoV have not been determined. Collection of
- multiple specimens (types and time points) from the same patient may be necessary to detect
- the virus.
- • A false-negative result may occur if a specimen is improperly collected, transported or handled.
- False-negative results may also occur if amplification inhibitors are present in the specimen or if
- inadequate numbers of organisms are present in the specimen.
- • Positive and negative predictive values are highly dependent on prevalence. False-negative test
- results are more likely when prevalence of disease is high. False-positive test results are more
- likely when prevalence is moderate to low.
- • Do not use any reagent past the expiration date.
- • If the virus mutates in the rRT-PCR target region, 2019-nCoV may not be detected or may be
- detected less predictably. Inhibitors or other types of interference may produce a false-negative
- result. An interference study evaluating the effect of common cold medications was not
- performed.
- • Test performance can be affected because the epidemiology and clinical spectrum of infection
- caused by 2019-nCoV is not fully known. For example, clinicians and laboratories may not know
- 38
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- the optimum types of specimens to collect, and, during the course of infection, when these
- specimens are most likely to contain levels of viral RNA that can be readily detected.
- • Detection of viral RNA may not indicate the presence of infectious virus or that 2019-nCoV is the
- causative agent for clinical symptoms.
- • The performance of this test has not been established for monitoring treatment of 2019-nCoV
- infection.
- • The performance of this test has not been established for screening of blood or blood products
- for the presence of 2019-nCoV.
- • This test cannot rule out diseases caused by other bacterial or viral pathogens.
- Conditions of Authorization for the Laboratory
- The CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel Letter of Authorization, along with the
- authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for Patients, and authorized
- labeling are available on the FDA website:
- https://www.fda.gov/MedicalDevices/Safety/EmergencySituations/ucm161496.htm
- Use of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel must follow the procedures outlined in
- these manufacturer’s Instructions for Use and the conditions of authorization outlined in the Letter of
- Authorization. Deviations from the procedures outlined are not permitted under the Emergency Use
- Authorization (EUA). To assist clinical laboratories running the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel, the relevant Conditions of Authorization are listed verbatim below, and are required to
- be met by laboratories performing the EUA test.
- • Authorized laboratories1 will include with reports of the results of the CDC 2019-nCoV Real-Time
- RT-PCR Diagnostic Panel, all authorized Fact Sheets. Under exigent circumstances, other
- appropriate methods for disseminating these Fact Sheets may be used, which may include mass
- media.
- • Authorized laboratories will perform the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel as
- outlined in the CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel
- Instructions for Use. Deviations from the authorized procedures, including the authorized RT-PCR
- instruments, authorized extraction methods, authorized clinical specimen types, authorized
- control materials, authorized other ancillary reagents and authorized materials required to
- perform the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel are not permitted. 2
- • Authorized laboratories that receive the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel must
- notify the relevant public health authorities of their intent to run the test prior to initiating
- testing.
- 1Authorized Laboratories: For ease of reference, the Letter of Authorization refers to “laboratories certified under the Clinical
- Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. § 263a, to perform high complexity tests” as “authorized
- laboratories.”
- 2If an authorized laboratory is interested in implementing changes to the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- that are not in the scope (Section II) of this letter of authorization FDA recommends you discuss with FDA after considering
- the policy outlined in Immediately in Effect Guidance for Clinical Laboratories and Food and Drug Administration Staff: Policy
- for Diagnostics Testing in Laboratories Certified to Perform High Complexity Testing under CLIA prior to Emergency Use
- Authorization for Coronavirus Disease-2019 during the Public Health Emergency
- (https://www.fda.gov/media/135659/download).
- 39
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- • Authorized laboratories will have a process in place for reporting test results to healthcare
- providers and relevant public health authorities, as appropriate.
- • Authorized laboratories will collect information on the performance of the test and report to
- DMD/OHT7-OIR/OPEQ/CDRH (via email: CDRH-EUA-Reporting@fda.hhs.gov) and CDC
- (respvirus@cdc.gov) any suspected occurrence of false-positive or false-negative results and
- significant deviations from the established performance characteristics of the test of which they
- become aware.
- • Authorized laboratories will report adverse events, including problems with test performance or
- results, to MedWatch by submitting the online FDA Form 3500
- (https://www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home) or by
- calling 1-800-FDA-1088
- • All laboratory personnel using the test must be appropriately trained in RT-PCR techniques and
- use appropriate laboratory and personal protective equipment when handling this kit and use
- the test in accordance with the authorized labeling.
- • CDC, IRR, manufacturers and distributors of commercial materials identified as acceptable on the
- CDC website, and authorized laboratories will ensure that any records associated with this EUA
- are maintained until otherwise notified by FDA. Such records will be made available to FDA for
- inspection upon request.
- Performance Characteristics
- Analytical Performance:
- Limit of Detection (LoD):
- LoD studies determine the lowest detectable concentration of 2019-nCoV at which approximately 95%
- of all (true positive) replicates test positive. The LoD was determined by limiting dilution studies using
- characterized samples.
- The analytical sensitivity of the rRT-PCR assays contained in the CDC 2019 Novel Coronavirus (2019-
- nCoV) Real-Time RT-PCR Diagnostic Panel were determined in Limit of Detection studies. Since no
- quantified virus isolates of the 2019-nCoV are currently available, assays designed for detection of the
- 2019-nCoV RNA were tested with characterized stocks of in vitro transcribed full length RNA (N gene;
- GenBank accession: MN908947.2) of known titer (RNA copies/µL) spiked into a diluent consisting of a
- suspension of human A549 cells and viral transport medium (VTM) to mimic clinical specimen. Samples
- were extracted using the QIAGEN EZ1 Advanced XL instrument and EZ1 DSP Virus Kit (Cat# 62724) and
- manually with the QIAGEN DSP Viral RNA Mini Kit (Cat# 61904). Real-Time RT-PCR assays were
- performed using the ThemoFisher Scientific TaqPath™ 1-Step RT-qPCR Master Mix, CG (Cat# A15299) on
- the Applied Biosystems™ 7500 Fast Dx Real-Time PCR Instrument according to the CDC 2019-nCoV RealTime RT-PCR Diagnostic Panel instructions for use.
- 40
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- A preliminary LoD for each assay was determined testing triplicate samples of RNA purified using each
- extraction method. The approximate LoD was identified by extracting and testing 10-fold serial dilutions
- of characterized stocks of in vitro transcribed full-length RNA. A confirmation of the LoD was determined
- using 3-fold serial dilution RNA samples with 20 extracted replicates. The LoD was determined as the
- lowest concentration where ≥ 95% (19/20) of the replicates were positive.
- Table 4. Limit of Detection Confirmation of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- with QIAGEN EZ1 DSP
- Targets 2019-nCoV_N1 2019-nCoV_N2
- RNA Concentration1 10 0.5 10 0.0 10 -0.5 10 0.5 10 0.0 10 -0.5
- Positives/Total 20/20 19/20 13/20 20/20 17/20 9/20
- Mean Ct2 32.5 35.4 NA 35.8 NA NA
- Standard Deviation
- (Ct) 0.5 0.8 NA 1.3 NA NA
- 1 Concentration is presented in RNA copies/µL
- 2 Mean Ct reported for dilutions that are ≥ 95% positive. Calculations only include positive results.
- NA not applicable
- Table 5. Limit of Detection Confirmation CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel with
- QIAGEN QIAmp DSP Viral RNA Mini Kit
- Targets 2019-nCoV_N1 2019-nCoV_N2
- RNA Concentration1 10 0.5 10 0.0 10 -0.5 10 0.5 10 0.0 10 -0.5 10 -1.0
- Positives/Total 20/20 20/20 6/20 20/20 20/20 20/20 8/20
- Mean Ct2 32.0 32.8 NA 33.0 35.4 36.2 NA
- Standard Deviation
- (Ct) 0.7 0.8 NA 1.4 0.9 1.9 NA
- 1 Concentration is presented in RNA copies/µL
- 2 Mean Ct reported for dilutions that are ≥ 95% positive. Calculations only include positive results.
- NA not applicable
- Table 6. Limit of Detection of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Virus Material
- Limit of Detection (RNA copies/µL)
- QIAGEN EZ1
- Advanced XL
- QIAGEN DSP Viral
- RNA Mini Kit
- 2019 Novel
- Coronavirus
- N Gene RNA
- Transcript 100.5 100
- FDA Sensitivity Evaluation: The analytical sensitivity of the test will be further assessed by evaluating an
- FDA-recommended reference material using an FDA developed protocol if applicable and/or when
- available.
- 41
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- In Silico Analysis of Primer and Probe Sequences:
- The oligonucleotide primer and probe sequences of the CDC 2019 nCoV Real-Time RT-PCR Diagnostic
- Panel were evaluated against 31,623 sequences available in the Global Initiative on Sharing All Influenza
- Data (GISAID, https://www.gisaid.org) database as of June 20, 2020, to demonstrate the predicted
- inclusivity of the 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Nucleotide mismatches in the
- primer/probe regions with frequencies > 0.1% are shown below. With the exception of one nucleotide
- mismatch with frequency > 1% (2.00%) at the third position of the N1 probe, the frequency of all
- mismatches was < 1%, indicating that prevalence of the mismatches were sporadic. Only one sequence
- (0.0032%) had two nucleotide mismatches in the N1 probe, and one other sequence from a different
- isolate (0.0032%) had two nucleotide mismatches in the N1 reverse primer. No sequences were found to
- have more than one mismatch in any N2 primer/probe region. The risk of these mismatches resulting in
- a significant loss in reactivity causing a false negative result is extremely low due to the design of the
- primers and probes, with melting temperatures > 60°C and with annealing temperature at 55°C that can
- tolerate up to two mismatches.
- Table 7. In Silico Inclusivity Analysis of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel Among
- 31,623 Genome Sequences Available from GISAID as of June 20, 2020
- Primer/probe N1 probe N1 reverse N2 probe
- Location (5'>3') 3 15 21 13
- Mismatch Nucleotide C>T G>T T>C C>T
- Mismatch No. 632 34 71 46
- Mismatch Frequency (%) 2.00 0.11 0.22 0.15
- Specificity/Exclusivity Testing: In Silico Analysis
- BLASTn analysis queries of the 2019-nCoV rRT-PCR assays primers and probes were performed against
- public domain nucleotide sequences. The database search parameters were as follows: 1) The nucleotide
- collection consists of GenBank+EMBL+DDBJ+PDB+RefSeq sequences, but excludes EST, STS, GSS, WGS,
- TSA, patent sequences as well as phase 0, 1, and 2 HTGS sequences and sequences longer than 100Mb;
- 2) The database is non-redundant. Identical sequences have been merged into one entry, while
- preserving the accession, GI, title and taxonomy information for each entry; 3) Database was updated on
- 10/03/2019; 4) The search parameters automatically adjust for short input sequences and the expect
- threshold is 1000; 5) The match and mismatch scores are 1 and -3, respectively; 6) The penalty to create
- and extend a gap in an alignment is 5 and 2 respectively.
- 2019-nCoV_N1 Assay:
- Probe sequence of 2019-nCoV rRT-PCR assay N1 showed high sequence homology with SARS
- coronavirus and Bat SARS-like coronavirus genome. However, forward and reverse primers showed no
- sequence homology with SARS coronavirus and Bat SARS-like coronavirus genome. Combining primers
- and probe, there is no significant homologies with human genome, other coronaviruses or human
- microflora that would predict potential false positive rRT-PCR results.
- 42
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 2019-nCoV_N2 Assay:
- The forward primer sequence of 2019-nCoV rRT-PCR assay N2 showed high sequence homology to Bat
- SARS-like coronaviruses. The reverse primer and probe sequences showed no significant homology with
- human genome, other coronaviruses or human microflora. Combining primers and probe, there is no
- prediction of potential false positive rRT-PCR results.
- In summary, the 2019-nCoV rRT-PCR assay N1 and N2, designed for the specific detection of 2019-nCoV,
- showed no significant combined homologies with human genome, other coronaviruses, or human
- microflora that would predict potential false positive rRT-PCR results.
- In addition to the in silico analysis, several organisms were extracted and tested with the CDC 2019-nCoV
- Real-Time RT-PCR Diagnostic Panel to demonstrate analytical specificity and exclusivity. Studies were
- performed with nucleic acids extracted using the QIAGEN EZ1 Advanced XL instrument and EZ1 DSP
- Virus Kit. Nucleic acids were extracted from high titer preparations (typically ≥ 105 PFU/mL or ≥ 106
- CFU/mL). Testing was performed using the ThemoFisher Scientific TaqPath™ 1-Step RT-qPCR Master Mix,
- CG on the Applied Biosystems™ 7500 Fast Dx Real-Time PCR instrument. The data demonstrate that the
- expected results are obtained for each organism when tested with the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel.
- 43
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Table 8. Specificity/Exclusivity of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Virus Strain Source
- 2019-
- nCoV_
- N1
- 2019-
- nCoV_
- N2
- Final
- Result
- Human coronavirus 229E Isolate 0/3 0/3 Neg.
- Human coronavirus OC43 Isolate 0/3 0/3 Neg.
- Human coronavirus NL63 clinical specimen 0/3 0/3 Neg.
- Human coronavirus HKU1 clinical specimen 0/3 0/3 Neg.
- MERS-coronavirus Isolate 0/3 0/3 Neg.
- SARS-coronavirus Isolate 0/3 0/3 Neg.
- bocavirus - clinical specimen 0/3 0/3 Neg.
- Mycoplasma pneumoniae Isolate 0/3 0/3 Neg.
- Streptococcus Isolate 0/3 0/3 Neg.
- Influenza A(H1N1) Isolate 0/3 0/3 Neg.
- Influenza A(H3N2) Isolate 0/3 0/3 Neg.
- Influenza B Isolate 0/3 0/3 Neg.
- Human adenovirus, type 1 Ad71 Isolate 0/3 0/3 Neg.
- Human metapneumovirus - Isolate 0/3 0/3 Neg.
- respiratory syncytial virus Long A Isolate 0/3 0/3 Neg.
- rhinovirus Isolate 0/3 0/3 Neg.
- parainfluenza 1 C35 Isolate 0/3 0/3 Neg.
- parainfluenza 2 Greer Isolate 0/3 0/3 Neg.
- parainfluenza 3 C-43 Isolate 0/3 0/3 Neg.
- parainfluenza 4 M-25 Isolate 0/3 0/3 Neg.
- Endogenous Interference Substances Studies:
- The CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel uses conventional well-established nucleic acid
- extraction methods and based on our experience with CDC’s other EUA assays, including the CDC Novel
- Coronavirus 2012 Real-time RT-PCR Assay for the presumptive detection of Middle East Respiratory
- Syndrome Coronavirus (MERS-CoV) and the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic
- Panel-Influenza A/H7 (Eurasian Lineage) Assay for the presumptive detection of novel influenza A (H7N9)
- virus that are both intended for use with a number of respiratory specimens, we do not anticipate
- interference from common endogenous substances.
- Specimen Stability and Fresh-frozen Testing:
- To increase the likelihood of detecting infection, CDC recommends collection of lower respiratory and
- upper respiratory specimens for testing. If possible, additional specimen types (e.g., stool, urine) should
- be collected and should be stored initially until decision is made by CDC whether additional specimen
- sources should be tested. Specimens should be collected as soon as possible once a PUI is identified
- regardless of symptom onset. Maintain proper infection control when collecting specimens. Store
- specimens at 2-8°C and ship overnight to CDC on ice pack. Label each specimen container with the
- patient’s ID number (e.g., medical record number), unique specimen ID (e.g., laboratory requisition
- number), specimen type (e.g., nasal swabs) and the date the sample was collected. Complete a CDC
- Form 50.34 for each specimen submitted.
- 44
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Clinical Performance:
- As of February 22, 2020, CDC has tested 2071 respiratory specimens from persons under investigation
- (PUI) in the U.S. using the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Specimen types include
- bronchial fluid/wash, buccal swab, nasal wash/aspirate, nasopharyngeal swab, nasopharyngeal/throat
- swab, oral swab, sputum, oropharyngeal (throat) swab, swab (unspecified), and throat swab.
- Table 9: Summary of CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel Data
- Generated by Testing Human Respiratory Specimens Collected from PUI Subjects in the U.S.
- Specimen Type
- 2019 nCoV
- Negative
- 2019 nCoV
- Positive Inconclusive Invalid Total
- Bronchial
- fluid/wash 2 0 0 0 2
- Buccal swab 5 1 0 0 6
- Nasal
- wash/aspirate 6 0 0 0 6
- Nasopharyngeal
- swab 927 23 0 0 950
- Nasopharyngeal
- swab/throat
- swab 4 0 0 0 4
- Oral swab 476 9 0 0 485
- Pharyngeal
- (throat) swab 363 10 0 1 374
- Sputum 165 5 0 0 170
- Swab
- (unspecified)1 71 1 0 0 72
- Tissue (lung) 2 0 0 0 2
- Total 2021 49 0 1 2071
- 1Actual swab type information was missing from these upper respiratory tract specimens.
- Two thousand twenty-one (2021) respiratory specimens of the 2071 respiratory specimens tested
- negative by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Forty-nine (49) of the 2071
- respiratory specimens tested positive by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Only
- one specimen (oropharyngeal (throat) swab) was invalid. Of the 49 respiratory specimens that tested
- positive by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel, seventeen (17) were confirmed by
- genetic sequencing and/or virus culture (positive percent agreement = 17/17, 95% CI: 81.6%-100%)
- During the early phase of the testing, a total of 117 respiratory specimens collected from 46 PUI subjects
- were also tested with two analytically validated real-time RT-PCR assays that target separate and
- independent regions of the nucleocapsid protein gene of the 2019-nCoV, N4 and N5 assays. The
- nucleocapsid protein gene targets for the N4 and N5 assays are different and independent from the
- nucleocapsid protein gene targets for the two RT-PCR assays included in the CDC 2019-nCoV Real-Time
- RT-PCR Diagnostic Panel, N1 and N2. Any positive result from the N4 and/or the N5 assay was further
- investigated by genetic sequencing.
- 45
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Performance of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel testing these 117 respiratory
- specimens was estimated against a composite comparator. A specimen was considered comparator
- negative if both the N4 and the N5 assays were negative. A specimen was considered comparator
- positive when the N4 and/or the N5 assay generated a positive result, and the comparator positive
- result(s) were further investigated and confirmed to be 2019-nCoV RNA positive by genetic sequencing.
- Table 10: Percent Agreement of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel with the
- Composite Comparator
- CDC 2019-nCoV
- Panel Result
- Composite Comparator Result
- Positive Negative
- Positive 131 0
- Inconclusive 0 0
- Negative 0 104
- 1Composite comparator results were available for 13 of 49 CDC 2019-nCoV Panel positive specimens
- only.
- Positive percent agreement = 13/13 = 100% (95% CI: 77.2% - 100%)
- Negative percent agreement = 104/104 = 100% (95% CI: 96.4% - 100%)
- Enzyme Master Mix Evaluation:
- The limit of detection equivalence between the ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix and
- the following enzyme master mixes was evaluated: Quantabio qScript XLT One-Step RT-qPCR ToughMix,
- Quantabio UltraPlex 1-Step ToughMix (4X), and Promega GoTaq® Probe 1- Step RT-qPCR System. Serial
- dilutions of 2019 novel coronavirus (SARS CoV-2) transcript were tested in triplicate with the CDC 2019-
- nCoV Real-time RT-PCR Diagnostic Panel using all four enzyme master mixes. Both manufactured
- versions of oligonucleotide probe, BHQ and ZEN, were used in the comparison. The lowest detectable
- concentration of transcript at which all replicates tested positive using the Quantabio qScript XLT OneStep RT-qPCR ToughMix and Quantabio UltraPlex 1-Step ToughMix (4X) was similar to that observed for
- the ThemoFisher TaqPath™ 1-Step RT-qPCR Master Mix. The lowest detectable concentration of
- transcript when using the Promega GoTaq® Probe 1- Step RT-qPCR System was one dilution above that
- observed for the other candidates when evaluated with the BHQ version of the CDC assays. The
- candidate master mixes all performed equivalently or at one dilution below the ThemoFisher TaqPath™
- 1-Step RT-qPCR Master Mix when evaluated with the ZEN version of the CDC assays.
- 46
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Table 11: Limit of Detection Comparison for Enzyme Master Mixes – BHQ Probe Summary Results
- Copy Number
- ThemoFisher TaqPath™
- 1-Step RT-qPCR Master
- Mix
- Quantabio qScript XLT
- One-Step RT-qPCR
- ToughMix
- Quantabio UltraPlex 1-
- Step ToughMix (4X)
- Promega GoTaq® Probe
- 1- Step RT-qPCR System
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 102 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
- 101 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
- 100 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 2/3
- 10-1 copies µL 2/3 0/3 1/3 1/3 1/3 1/3 0/3 0/3
- Table 12: Limit of Detection Comparison for Enzyme Master Mixes – ZEN Probe Summary Results
- Copy Number
- ThemoFisher TaqPath™
- 1-Step RT-qPCR Master
- Mix
- Quantabio qScript XLT
- One-Step RT-qPCR
- ToughMix
- Quantabio UltraPlex 1-
- Step ToughMix (4X)
- Promega GoTaq® Probe
- 1- Step RT-qPCR System
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 2019-
- nCoV_N1
- 2019-
- nCoV_N2
- 102 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
- 101 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
- 100 copies/µL 3/3 2/3 3/3 3/3 3/3 2/3 3/3 3/3
- 10-1 copies µL 1/3 1/3 0/3 0/3 0/3 1/3 1/3 1/3
- Retrospective positive (18) and negative (17) clinical respiratory specimens were extracted using the
- QIAGEN EZ1 Advanced XL instrument and EZ1 DSP Virus Kit and were tested with the CDC 2019-nCoV
- Real-time RT-PCR Diagnostic Panel using the Quantabio qScript XLT One-Step RT-qPCR ToughMix,
- Quantabio UltraPlex 1-Step ToughMix (4X), and Promega GoTaq® Probe 1- Step RT-qPCR System master
- mixes. All three enzyme master mixes performed equivalently, demonstrating 100% positive and 100%
- negative agreement with expected results and a 95% confidence interval of 82.4%-100% and 81.6%-
- 100%, respectively.
- Table 13: Clinical Comparison – Retrospective Study Summary Results
- CDC 2019-nCoV
- Real-time RTPCR Diagnostic
- Panel Result
- Quantabio qScript XLT
- One-Step RT-qPCR
- ToughMix
- Quantabio UltraPlex 1-Step
- ToughMix (4X)
- Promega GoTaq® Probe 1-
- Step RT-qPCR System
- Positive Negative Positive Negative Positive Negative
- Positive 18 0 18 0 18 0
- Negative 0 17 0 17 0 17
- 47
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Roche MagNA Pure 24 and MagNA Pure 96 Extraction Platform Evaluation:
- Performance of the 2019-CoV Real-time RT-PCR Diagnostic Panel using the Roche MagNA Pure 24 and
- MagNA Pure 96 extraction platforms was compared to performance with an authorized extraction
- method. Serial dilutions of quantified inactivated SARS-CoV-2 virus (USA-WA1/2020; 100 RNA copies/µL)
- in lysis buffer were added to pooled negative upper respiratory tract specimen matrix. Five samples of
- each dilution were extracted in parallel with the QIAGEN EZ1 Advanced XL (EZ1 DSP Virus Kit Cat# 62724)
- and the Roche MagNA Pure 24 (MagNA Pure 24 Total NA Isolation Kit Cat# 07658036001) and Roche
- MagNA Pure 96 (MagNA Pure 96 DNA and Viral Nucleic Acid Small Volume Kit Cat# 06543588001)
- extraction platforms and evaluated using the 2019-nCoV Real-Time RT-PCR Diagnostic Panel and
- ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix. The observed LoD was defined as the lowest
- concentration at which 100% (5 out of 5 total) of all replicates tested positive for both primer/probe sets
- (N1 and N2) in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. The acceptance criteria for
- equivalence were defined as demonstrating an observed LoD either at the same endpoint or within a 3-
- fold dilution. The results showed that both the MagNA Pure 24 and MagNA Pure 96 extraction platforms
- performed equivalently or within one 3-fold dilution of the LoD observed when using the QIAGEN EZ1
- Advanced XL extraction platform.
- Table 14. Limit of Detection Comparison between the QIAGEN EZ1 Advanced XL, Roche MagNA Pure
- 96, and Roche MagNA Pure 24 Extraction Platforms using the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel
- Platform Parameter 2019-nCoV_N1 Assay 2019-nCoV_N2 Assay Observed
- LoD1
- QIAGEN EZ1
- Advanced XL
- RNA copies/µL 101.0 100.5 100.0 101.0 100.5 100.0
- 100.5 # pos./total 5/5 5/5 5/5 5/5 5/5 3/5
- Mean Ct2 34.0 35.0 36.3 33.9 36.6 NA
- Std. Deviation 0.2 0.8 0.2 0.4 0.9 NA
- Roche MagNA Pure
- 96
- RNA copies/µL 101.0 100.5 100.0 101.0 100.5 100.0
- 100.5 # pos./total 5/5 5/5 5/5 5/5 5/5 2/5
- Mean Ct2 33.3 34.6 36.1 33.2 35.7 NA
- Std. Deviation 0.5 0.5 0.3 0.3 0.4 NA
- Roche MagNA Pure
- 24
- RNA copies/µL 101.0 100.5 100.0 101.0 100.5 100.0
- 101.0 # pos./total 5/5 3/5 3/5 5/5 5/5 5/5
- Mean Ct2 34.4 NA NA 35.2 36.9 36.2
- Std. Deviation 0.6 NA NA 0.5 1.0 0.8
- 1
- Concentration is presented in RNA copies/µL. The observed LoD is the lowest concentration where both assays showed 100%
- positive detection.
- 2
- Mean Ct reported for dilutions that show 100% positivity. Calculations only include positive results.
- NA = not applicable
- Previously characterized clinical remainder specimens (14 positive and 15 negative) were extracted using
- both the Roche MagNA Pure 96 and MagNA Pure 24 extraction platforms and evaluated using the 2019-
- nCoV Real-Time RT-PCR Diagnostic Panel and ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix.
- Acceptance criteria for clinical equivalence was defined as demonstrating 100% concurrence with
- qualitative results shown with the authorized comparator method (QIAGEN EZ1 Advanced XL). Results
- from this study showed 100% concurrence with the comparator method for both the Roche MagNA Pure
- 48
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- 96 and Roche MagNA Pure 24 extraction platforms when used with the CDC 2019-nCoV Real-Time RTPCR Diagnostic panel.
- Table 15. Clinical Comparison Results – Retrospective Study Results
- Test Platform
- Test
- Platform
- Result
- QIAGEN EZ1
- Advanced XL Result Positive % Agreement (CI)1 Negative % Agreement (CI)1
- Positive Negative
- Roche MagNA
- Pure 96
- Positive 14 0 100.0 (78.5 – 100.0) 100.0 (79.6 – 100.0) Negative 0 15
- Roche MagNA
- Pure 24
- Positive 14 0 100.0 (78.5 – 100.0) 100.0 (79.6 – 100.0) Negative 0 15 1 CI = 95% confidence interval
- Promega Maxwell® RSC 48 Extraction Platform Evaluation:
- Performance of the 2019-CoV Real-time RT-PCR Diagnostic Panel using the Promega Maxwell® RSC 48
- extraction platform was compared to performance with an authorized extraction method. Serial
- dilutions of quantified inactivated SARS-CoV-2 virus (USA-WA1/2020; 100 RNA copies/µL) in VTM were
- added to pooled negative upper respiratory tract specimen matrix. Five samples of each dilution were
- extracted in parallel with the QIAGEN EZ1® Advanced XL (EZ1 DSP Virus Kit Cat# 62724) and the Promega
- Maxwell® RSC 48 (Promega Maxwell® Viral Total Nucleic Acid Purification Kit Cat# AS1330) extraction
- platforms and evaluated using the 2019-nCoV Real-Time RT-PCR Diagnostic Panel and ThermoFisher
- TaqPath™ 1-Step RT-qPCR Master Mix. The observed LoD was defined as the lowest concentration at
- which 100% (5 out of 5 total) of all replicates tested positive for both primer/probe sets (N1 and N2) in
- the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. The acceptance criteria for equivalence were
- defined as demonstrating an observed LoD either at the same endpoint or within a 3-fold dilution. The
- results showed that the performance of the Maxwell® RSC 48 extraction platform performed
- equivalently or within one 3-fold dilution of the LoD observed when using the QIAGEN EZ1® Advanced XL
- extraction platform.
- Table 16. Limit of Detection Comparison Between the QIAGEN EZ1® Advanced XL and Promega
- Maxwell® RSC 48 Extraction Platforms Using the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Platform Parameter 2019-nCoV_N1 Assay 2019-nCoV_N2 Assay Observed
- LoD1
- QIAGEN EZ1®
- Advanced XL
- RNA copies/µL 100.5 100.0 10-0.5 100.5 100.0 10-0.5
- 10 # pos./total 5/5 5/5 0/5 5/5 5/5 3/5 0.0
- Mean Ct2 32.27 33.80 NA 35.13 36.41 NA
- Std. Deviation 0.81 0.40 NA 0.81 0.40 NA
- Promega Maxwell®
- RSC 48
- RNA copies/µL 100.5 100.0 10-0.5 100.5 100.0 10-0.5
- 10 # pos./total 5/5 5/5 3/5 5/5 5/5 5/5 0.0
- Mean Ct2 31.11 32.97 NA 31.89 33.95 35.17
- Std. Deviation 0.24 0.34 NA 0.24 0.35 0.65
- 1
- Concentration is presented in RNA copies/µL. The observed LoD is the lowest concentration where both assays showed 100%
- positive detection.
- 2
- Mean cycle threshold (Ct) reported for dilutions that show 100% positivity. Calculations only include positive results.
- NA = not applicable
- 49
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Previously characterized clinical remainder specimens (15 positive and 15 negative) were extracted using
- the Promega Maxwell® RSC 48 extraction platform alongside the currently authorized QIAGEN EZ1®
- Advanced XL extraction platform and evaluated using the 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- and ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix. Results from the Maxwell® RSC 48 were
- compared with the QIAGEN EZ1® Advanced XL extraction performed in parallel showing 100% (15/15)
- qualitative concurrence on positive samples and 93.3% (14/15) qualitative concurrence on negative
- samples. This evaluation showed that two originally negative (QIAGEN QIAamp® DSP Viral RNA Mini Kit)
- specimens (Specimens 16 and 24) yielded an inconclusive result after extraction using the QIAGEN EZ1®
- Advanced XL. Repeat of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel resolved one of the two
- specimens (Specimen 24, negative result). The second specimen (Specimen 16) remained inconclusive.
- Both these specimens yielded a negative result on the Maxwell® RSC 48.
- Table 17. Clinical Comparison Results – Retrospective Study Results
- Test Platform
- Promega Maxwell® RSC 48 Positive %
- Agreement (CI)1
- Negative %
- Agreement (CI)1
- Result Positive Negative Inconclusive
- QIAGEN EZ1®
- Advanced XL
- Positive 15 0 0
- 100.0
- (79.6-100.0)
- 93.3
- (70.2-98.9) Negative 0 14 0
- Inconclusive 0 1 0
- 1 CI = 95% confidence interval
- Disposal
- Dispose of hazardous or biologically contaminated materials according to the practices of your
- institution.
- References
- 1. Ballew, H. C., et al. “Basic Laboratory Methods in Virology,” DHHS, Public Health Service 1975
- (Revised 1981), Centers for Disease Control and Prevention, Atlanta, Georgia 30333.
- 2. Clinical Laboratory Standards Institute (CLSI), “Collection, Transport, Preparation and Storage of
- Specimens for Molecular Methods: Proposed Guideline,” MM13-A
- 3. Lieber, M., et al. "A Continuous Tumor Cell Line from a Human Lung Carcinoma with Properties of
- Type II Alveolar Epithelial Cells." International Journal of Cancer 1976, 17(1), 62-70.
- 50
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Revision History
- Revision # Effective Date Summary of Revisions
- 1 February 4, 2020 Original Instructions for Use
- 2 March 15, 2020 • Intended use update
- • Removal of N3 primer and probe set from Diagnostic Panel
- • Performance data update
- • Addition of alternative nucleic acid extraction platforms
- • Addition of acceptable alternatives to HSC and addition of
- QIAGEN RUO extraction reagents
- • Positive results no longer presumptive. No confirmation of
- positive results required
- 3 March 30, 2020 • Addition of alternative enzyme master mix options
- 4 June 12, 2020 • Addition of MagNA Pure 24 extraction method
- • Addition of performance data for the MagNA Pure 96
- extraction method with SARS-CoV-2
- • Addition of heat treatment alternative to specimen
- extraction
- • Addition of Roche and QIAGEN external lysis buffer
- alternatives
- • Acknowledgment of FDA policy permitting end users to
- qualify alternative components without seeking an EUA or
- EUA amendment
- 5 July 13, 2020 • Addition of Promega Maxwell® RSC 48 extraction method
- • Update to in silico inclusivity analyses
- Contact Information, Ordering, and Product Support
- For technical and product support, contact the CDC Division of Viral Diseases directly.
- Send email to: respvirus@cdc.gov
- Note: If your laboratory is using reagents sourced from someone other than the CDC International
- Reagent Resource, please refer to the manufacturer’s instructions provided with the commercial
- materials.
- 51
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Appendix A: Heat Treatment Alternative to Extraction
- UltraPlex 1-Step ToughMix (4X)
- This procedure is only for use by public health laboratories.
- Purpose:
- In response to a global shortage of nucleic acid extraction reagents causing significant delays in testing,
- the CDC has investigated the use of a heat treatment method requiring minimal reagents as a specimen
- processing alternative to nucleic acid extraction for use with the 2019-nCoV Real-Time RT-PCR Diagnostic
- Panel.
- Where possible, laboratories should use qualified RNA or total nucleic acid extraction methods for
- processing of specimens for subsequent testing by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic
- Panel. Extraction removes inhibitory substances from specimens that could negatively impact PCR
- performance.
- This procedure for use of heat treatment for specimen processing is only recommended when a
- shortage of qualified extraction reagents is a limiting factor in a laboratory’s ability to meet urgent
- COVID-19 testing demand.
- Precautions/Warnings/Limitations:
- • CDC has evaluated this heat treatment process and has determined that this process is effective
- for inactivation of SARS-CoV-2 in patient specimens.
- • Performance was evaluated with only upper respiratory specimens. Heat treatment of lower
- respiratory specimens for subsequent testing by the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel has not been evaluated.
- • This procedure for heat treatment of specimens is only for use with the Quantabio UltraPlex 1-
- Step ToughMix (4X).
- • Heat treatment should only be conducted when a lab is ready to test the specimens by PCR.
- Testing of heat-treated specimens must be conducted the same day.
- Acceptable Specimens:
- • Upper respiratory specimens
- Note: Do not use heat treatment to process specimens that appear bloody or that contain
- particulate matter. Such specimens should be extracted using a qualified RNA or TNA extraction
- method prior to testing.
- Materials Required (not provided):
- • 70% ethanol
- • 10% bleach, freshly prepared
- • 96-well PCR reaction plates (Applied Biosystems catalog # 4346906, 4366932, 4346907, or
- equivalent)
- • Optical strip caps (Applied Biosystems 4323032, or equivalent)
- • 1.5 mL Sarstedt tubes or equivalent
- 52
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- • Aerosol resistant micropipette tips
- • Micropipettes
- • 96-well cold block
- • Cold blocks for 1.5 mL - 2.0 mL tubes
- • Vortex mixer
- • 96-well plate centrifuge or equivalent
- • Thermal cycler or equivalent
- • Class II Biological Safety Cabinet (BSC)
- Procedure:
- Sample Preparation
- 1) Decontaminate BSC with 10% bleach followed by 70% ethanol.
- 2) If samples are frozen, thaw on ice or at 4°C. Wipe the outside of the sample tube with 70%
- ethanol. Place thawed sample on cold rack or ice in BSC.
- 3) Pulse vortex each sample and briefly spin down in a centrifuge to collect the liquid at the
- bottom of the tube.
- Heat Treatment
- 1) Place a thermal cycler in the BSC, turn on, and program for 95°C for 1 min followed by 4°C
- hold.
- 2) Place a 96-well PCR plate onto a cold rack or ice in the BSC.
- 3) Transfer 100 µL of each sample to the 96-well PCR plate and securely cap each well using
- optical strip caps.
- NOTE: Ensure that an HSC extraction control is included in each batch run as required under
- CLIA.
- 4) Place this 96-well PCR plate on the pre-heated thermal cycler and start run. Leave plate on
- thermal cycler at 4°C, or place on ice or a cold block.
- 5) Remove plate and centrifuge for 1 minute at 500 x g to pellet cellular debris.
- 6) Place plate on a cold rack or ice and proceed to testing the supernatant by rRT-PCR.
- 7) Testing of heat-treated specimens must be conducted the same day heat treatment is
- performed. For long term storage, keep the original specimen at ≤-70°C.
- Special Testing Considerations for Heat Treated Specimens:
- • Enzyme Master Mix
- Testing of specimens that have been processed with heat treatment should be conducted
- with the Quantabio UltraPlex 1-Step ToughMix (4X), which demonstrated the best
- performance with heat treated specimens. PCR testing of heat-treated specimens should
- follow the instructions in the main body of this Instructions for Use document.
- • Resolution of Inconclusive and Invalid Results
- Retesting of heat-treated specimens that generated an inconclusive or invalid result must
- include extraction of the original specimen with a qualified RNA or total nucleic acid (TNA)
- extraction method, if available. Do not re-test the heat-treated specimen material to resolve
- inconclusive or invalid test results.
- 53
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Verification:
- CDC recommends performance of verification studies for the heat treatment method prior to diagnostic
- use that includes side-by-side preparation of a panel of positive and negative clinical specimens using a
- qualified extraction method and this heat treatment method with subsequent testing by the CDC 2019-
- nCoV Real-Time RT-PCR Diagnostic Panel.
- Performance Characteristics:
- Quantabio UltraPlex 1-Step ToughMix (4X)
- Limit of Detection Comparison
- Serial dilutions of inactivated SARS-CoV-2 [SARS-CoV-2 USA-WA1/2020] were prepared in simulated
- specimen material (human A549 cells suspended in viral transport medium). Each concentration was
- prepared side-by-side five times by both EZ1 extraction and by heat treatment. Each extracted or heattreated sample was subsequently tested by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel using
- the Quantabio UltraPlex 1-Step ToughMix (4X) on the Applied Biosystems 7500 Fast Dx instrument.
- Observed detection was similar between the two specimen preparation methods.
- Table B1: UltraPlex Limit of Detection Comparison between QIAGEN EZ1 Advanced XL extraction and heat
- treatment (95°C for 1 min) method – Summary Results
- Enzyme Platform Parameter 2019-nCoV_N1 Assay 2019-nCoV_N2 Assay Observed
- LoD1
- Quantabio UltraPlex 1-
- Step ToughMix (4X)
- 5 µL Template Addition
- QIAGEN
- EZ1
- Advanced
- XL
- RNA copies/µL 101.0 100.5 100.0 10-0.5 10-1.0 101.0 100.5 100.0 10-0.5 10-1.0
- 100.5 # pos./total 5/5 5/5 4/5 4/5 3/5 5/5 5/5 5/5 2/5 2/5
- Mean Ct2 34.11 34.59 NA NA NA 32.97 33.76 34.70 NA NA
- Std. Deviation 0.75 0.99 NA NA NA 0.33 0.72 0.98 NA NA
- Heat
- Treatment
- 95°C
- for 1 min
- RNA copies/µL 101.0 100.5 100.0 10-0.5 10-1.0 101.0 100.5 100.0 10-0.5 10-1.0
- 100.5 # pos./total 5/5 5/5 4/5 5/5 1/5 5/5 5/5 4/5 2/5 1/5
- Mean Ct2 33.41 34.32 NA 36.73 NA 33.45 35.25 NA NA NA
- Std. Deviation 0.62 0.40 NA 0.82 NA 0.40 0.80 NA NA NA
- 1Concentration is presented in RNA copies/µL. The observed LoD is the lowest concentration where both assays showed 100%
- positive detection.
- 2Mean Ct reported for dilutions that show 100% positivity. Calculations only include positive results.
- NA = not applicable
- 54
- CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
- Clinical Comparison
- A panel of 39 upper respiratory specimens were tested side-by-side using extraction with the Qiagen EZ1
- extraction instrument and heat treatment. Extracted and heat-treated specimens were subsequently
- tested with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel using the Quantabio UltraPlex 1-Step
- ToughMix (4X). Qualitative results were compared to demonstrate agreement.
- Table B2: Clinical Comparison Results Summary – Heat Treatment versus QIAGEN EZ1 Advanced XL
- Test Result Heat Treatment Total
- Positive %
- Agreement (CI)1
- Negative %
- Agreement (CI)1
- Positive Inconclusive Negative
- QIAGEN EZ1
- Advanced XL
- Positive 18 1 0 19
- Inconclusive 0 0 0 0 94.7 (75.4-99.1) 100 (83.9-100)
- Negative 0 0 20 20
- Total 18 1 20 39
- 1 CI = 95% confidence interval
- Questions and Comments:
- If you have questions or comments about this procedure, please send by email to: respvirus@cdc.gov
- Division of Viral Diseases / Respiratory Viruses Branch
- CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
- Requirements
- *** DO NOT DISCARD: Important product-specific information ***
- Document #: CDC-006-00005 Revision #: 05 Effective Date: 07/13/2020 Page 1 of 4
- CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel –
- Verification Requirements
- Please consult the following guidance from the Centers for Medicare & Medicaid Services
- (CMS) regarding diagnostic tests under Emergency Use Authorization (EUA):
- https://www.cms.gov/Medicare/Provider-Enrollment-andCertification/SurveyCertificationGenInfo/Policy-and-Memos-to-States-and-RegionsItems/QSO18-19-CLIA
- INTENDED USE
- The CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is a realtime RT-PCR test intended for the qualitative detection of nucleic acid from the 2019-nCoV in
- upper and lower respiratory specimens (such as nasopharyngeal or oropharyngeal swabs,
- sputum, lower respiratory tract aspirates, bronchoalveolar lavage, and nasopharyngeal
- wash/aspirate or nasal aspirate) collected from individuals who meet 2019-nCoV clinical
- and/or epidemiological criteria (for example, clinical signs and symptoms associated with
- 2019-nCoV infection, contact with a probable or confirmed 2019-nCoV case, history of travel
- to a geographic locations where 2019-nCoV cases were detected, or other epidemiologic links
- for which 2019-nCoV testing may be indicated as part of a public health investigation). Testing
- in the United States is limited to laboratories certified under the Clinical Laboratory
- Improvement Amendments of 1988 (CLIA), 42 U.S.C. § 263a, to perform high complexity
- tests.
- Results are for the identification of 2019-nCoV RNA. The 2019-nCoV RNA is generally
- detectable in upper and lower respiratory specimens during infection. Positive results are
- indicative of active infection with 2019-nCoV but do not rule out bacterial infection or coinfection with other viruses. The agent detected may not be the definite cause of disease.
- Laboratories within the United States and its territories are required to report all positive
- results to the appropriate public health authorities.
- Negative results do not preclude 2019-nCoV infection and should not be used as the sole basis
- for treatment or other patient management decisions. Negative results must be combined with
- clinical observations, patient history, and epidemiological information.
- Testing with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is intended for use by
- trained laboratory personnel who are proficient in performing real-time RT-PCR assays. The
- CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is only for use
- under a Food and Drug Administration’s Emergency Use Authorization.
- REQUIRED MATERIALS
- The 2019 novel coronavirus positive control (nCoVPC) is provided with the CDC 2019-nCoV
- Real-Time RT-PCR Diagnostic Panel and should be prepared according to the Instructions for
- Use. The nCoVPC consists of an RNA transcript of the 2019-nCoV N gene as well as human
- RNase P gene segment. nCoVPC will yield a positive result with the following primer and probe
- sets: 2019-nCoV_N1, 2019-nCoV_N2, and RP.
- Approximately 2 mL of an upper respiratory specimen (e.g. nasopharyngeal swab (NPS) in
- transport media) are needed for testing. Specimens may be pooled if less than 2 mL of one
- specimen is available.
- Refer to CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel package insert (manufacturer
- instructions) for additional reagents, materials, and instructions.
- PRECAUTIONS
- This reagent should be handled in an approved biosafety level 2 (BSL-2) handling area to
- avoid contamination of laboratory equipment and reagents that could cause false positive
- Division of Viral Diseases / Respiratory Viruses Branch
- CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
- Requirements
- *** DO NOT DISCARD: Important product-specific information ***
- Document #: CDC-006-00005 Revision #: 05 Effective Date: Page 2 of 4
- results. This product is an RNA transcript and is non-infectious. However, the nCoVPC should
- be handled in accordance with Good Laboratory Practices.
- Store reagent at appropriate temperatures (see Instructions for Use) and hold on ice when
- thawed.
- Please use standard precautions when handling respiratory specimens.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE
- QIAamp® DSP VIRAL RNA MINI KIT OR THE QIAamp® VIRAL RNA MINI KIT
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 560 µL of lysis buffer into each of nine tubes labeled 1-9.
- • Add 140 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- • To prepare samples at a moderate concentration, spike 14 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 14 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 14 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE QIAGEN
- EZ1® ADVANCED XL
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 280 µL of lysis buffer into each of nine tubes labeled 1-9.
- • Add 120 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE ROCHE
- MagNA PURE TOTAL NUCLEIC ACID KIT OR THE ROCHE MagNA PURE NUCLEIC ACID
- ISOLATION KIT I
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 300 µL of lysis buffer into each of nine tubes labeled 1-9.
- • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- Division of Viral Diseases / Respiratory Viruses Branch
- CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
- Requirements
- *** DO NOT DISCARD: Important product-specific information ***
- Document #: CDC-006-00005 Revision #: 05 Effective Date: Page 3 of 4
- • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE ROCHE
- MagNA PURE 24 AND TOTAL NUCLEIC ACID ISOLATION KIT
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 400 µL of lysis buffer into each of nine tubes labeled 1-9.
- • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE ROCHE
- MagNA PURE 96 DNA AND VIRAL NA SMALL VOLUME KIT
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 350 µL of lysis buffer into each of nine tubes labeled 1-9.
- • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE
- PROMEGA MAXWELL® RSC 48
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 330 µL of lysis buffer (300 µL of lysis buffer + 30 µL Proteinase K, included in the
- kit) into each of nine tubes labeled 1-9.
- Division of Viral Diseases / Respiratory Viruses Branch
- CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
- Requirements
- *** DO NOT DISCARD: Important product-specific information ***
- Document #: CDC-006-00005 Revision #: 05 Effective Date: Page 4 of 4
- • Add 120 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE
- BIOMÉRIEUX NucliSENS easyMAG OR THE BIOMÉRIEUX EMAG
- • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
- reconstitution of the materials for use. RNA should be kept cold during preparation and use.
- • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
- or 10 mM Tris.
- • Aliquot 1000 μL or 2000 µL of pre-aliquoted easyMAG lysis buffer into each of nine tubes
- labeled 1-9 for the easyMAG or eMAG, respectively.
- • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
- the nine labeled tubes with lysis buffer.
- • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
- (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
- Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
- • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
- each tube labeled 4-6 containing lysis buffer and specimen.
- • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
- containing lysis buffer and specimen.
- • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
- Diagnostic Panel Instructions for Use.
- PROCEDURE
- Follow the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for testing
- the nine extracted samples at least once.
- EXPECTED RESULTS
- Moderate nCoVPC samples should be positive for 2019-nCoV.
- Low nCoVPC samples should be positive for 2019-nCoV.
- Negative upper respiratory samples should be negative for 2019-nCoV.
- ≥90% of test results should be in agreement with the expected results. If test results are
- <90% in agreement with expected results, contact CDC at respvirus@cdc.gov.
- QUESTIONS
- Please send questions or comments by email to respvirus@cdc.gov.
- DISTRIBUTION
- Distributed to qualified laboratories by Centers for Disease Control and Prevention, 1600
- Clifton Road, Atlanta, GA, 30329 USA
page list
▼