page list

23 Oct 2020

CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel

  •  CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • CDC 2019-Novel Coronavirus (2019-nCoV)


  • #WakeUp
  • 👿💩👎
  • Real-Time RT-PCR Diagnostic Panel
  • For Emergency Use Only
  • Instructions for Use
  • Catalog # 2019-nCoVEUA-01
  • 1000 reactions
  • For In-vitro Diagnostic (IVD) Use
  • Rx Only
  • https://www.fda.gov/media/134922/downloadCenters for Disease Control and Prevention
  • Division of Viral Diseases
  • 1600 Clifton Rd NE
  • Atlanta GA 30329
  • 1
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Table of Contents
  • Intended Use ............................................................................................................................. 2
  • Summary and Explanation....................................................................................................... 2
  • Principles of the Procedure ..................................................................................................... 3
  • Materials Required (Provided)................................................................................................. 5
  • Materials Required (But Not Provided)................................................................................... 6
  • Warnings and Precautions .................................................................................................... 10
  • Reagent Storage, Handling, and Stability............................................................................. 11
  • Specimen Collection, Handling, and Storage....................................................................... 12
  • Specimen Referral to CDC ..................................................................................................... 13
  • Reagent and Controls Preparation........................................................................................ 13
  • General Preparation ............................................................................................................... 14
  • Nucleic Acid Extraction.......................................................................................................... 14
  • Assay Set Up........................................................................................................................... 16
  • Create a Run Template on the Applied Biosystems 7500 Fast Dx Real-time PCR
  • Instrument (Required if no template exists)......................................................................... 20
  • Defining the Instrument Settings .......................................................................................... 26
  • Running a Test........................................................................................................................ 29
  • Interpretation of Results and Reporting............................................................................... 34
  • 2019-nCoV rRT-PCR Diagnostic Panel Results Interpretation Guide ................................ 36
  • Quality Control........................................................................................................................ 37
  • Limitations .............................................................................................................................. 37
  • Conditions of Authorization for the Laboratory................................................................... 38
  • Performance Characteristics................................................................................................. 39
  • Disposal................................................................................................................................... 49
  • References .............................................................................................................................. 49
  • Revision History ..................................................................................................................... 50
  • Contact Information, Ordering, and Product Support......................................................... 50
  • Appendix A: Heat Treatment Alternative to Extraction ....................................................... 51
  • 2
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Intended Use
  • The CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is a real-time RT-PCR
  • test intended for the qualitative detection of nucleic acid from the 2019-nCoV in upper and lower
  • respiratory specimens (such as nasopharyngeal or oropharyngeal swabs, sputum, lower respiratory tract
  • aspirates, bronchoalveolar lavage, and nasopharyngeal wash/aspirate or nasal aspirate) collected from
  • individuals who meet 2019-nCoV clinical and/or epidemiological criteria (for example, clinical signs and
  • symptoms associated with 2019-nCoV infection, contact with a probable or confirmed 2019-nCoV case,
  • history of travel to geographic locations where 2019-nCoV cases were detected, or other epidemiologic
  • links for which 2019-nCoV testing may be indicated as part of a public health investigation). Testing in
  • the United States is limited to laboratories certified under the Clinical Laboratory Improvement
  • Amendments of 1988 (CLIA), 42 U.S.C. § 263a, to perform high complexity tests.
  • Results are for the identification of 2019-nCoV RNA. The 2019-nCoV RNA is generally detectable in upper
  • and lower respiratory specimens during infection. Positive results are indicative of active infection with
  • 2019-nCoV but do not rule out bacterial infection or co-infection with other viruses. The agent detected
  • may not be the definite cause of disease. Laboratories within the United States and its territories are
  • required to report all positive results to the appropriate public health authorities.
  • Negative results do not preclude 2019-nCoV infection and should not be used as the sole basis for
  • treatment or other patient management decisions. Negative results must be combined with clinical
  • observations, patient history, and epidemiological information.
  • Testing with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is intended for use by trained
  • laboratory personnel who are proficient in performing real-time RT-PCR assays. The CDC 2019-Novel
  • Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is only for use under a Food and Drug
  • Administration’s Emergency Use Authorization.
  • Summary and Explanation
  • An outbreak of pneumonia of unknown etiology in Wuhan City, Hubei Province, China was initially
  • reported to WHO on December 31, 2019. Chinese authorities identified a novel coronavirus (2019-
  • nCoV), which has resulted in millions of confirmed human infections globally. Cases of asymptomatic
  • infection, mild illness, severe illness, and deaths have been reported.
  • The CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is a molecular in vitro diagnostic test that aids in
  • the detection and diagnosis 2019-nCoV and is based on widely used nucleic acid amplification
  • technology. The product contains oligonucleotide primers and dual-labeled hydrolysis probes (TaqMan®)
  • and control material used in rRT-PCR for the in vitro qualitative detection of 2019-nCoV RNA in
  • respiratory specimens.
  • The term “qualified laboratories” refers to laboratories in which all users, analysts, and any person
  • reporting results from use of this device should be trained to perform and interpret the results from this
  • procedure by a competent instructor prior to use. 
  • 3
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Principles of the Procedure
  • The oligonucleotide primers and probes for detection of 2019-nCoV were selected from regions of the
  • virus nucleocapsid (N) gene. The panel is designed for specific detection of the 2019-nCoV (two
  • primer/probe sets). An additional primer/probe set to detect the human RNase P gene (RP) in control
  • samples and clinical specimens is also included in the panel.
  • RNA isolated and purified from upper and lower respiratory specimens is reverse transcribed to cDNA
  • and subsequently amplified in the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument with SDS
  • version 1.4 software. In the process, the probe anneals to a specific target sequence located between
  • the forward and reverse primers. During the extension phase of the PCR cycle, the 5’ nuclease activity of
  • Taq polymerase degrades the probe, causing the reporter dye to separate from the quencher dye,
  • generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved from
  • their respective probes, increasing the fluorescence intensity. Fluorescence intensity is monitored at
  • each PCR cycle by Applied Biosystems 7500 Fast Dx Real-Time PCR System with SDS version 1.4 software.
  • Detection of viral RNA not only aids in the diagnosis of illness but also provides epidemiological and
  • surveillance information.
  • 4
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Summary of Preparation and Testing Process
  • Resuspend
  • primer/probe mix,
  • aliquot and store at
  • ≤ -20°C
  • Resuspend and
  • aliquot nCoVPC,
  • store at -70°C
  • Extract sample RNA
  • and HSC RNA
  • Prepare master mix
  • (15 µL)
  • Prepare rRT-PCR
  • plate (5 µL RNA)
  • Run assay on
  • ABI 7500Fast Dx
  • Analyze data
  • Report results
  • Upon receipt of
  • rRT-PCR Panel
  • reagents
  • Upon obtaining
  • sample
  • 5
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Materials Required (Provided)
  • Note: CDC will maintain on its website a list of commercially available lots of primer and probe sets
  • and/or positive control materials that are acceptable alternatives to the CDC primer and probe set
  • and/or positive control included in the Diagnostic Panel. Only material distributed through the CDC
  • International Reagent Resource and specific lots of material posted to the CDC website are acceptable
  • for use with this assay under CDC’s Emergency Use Authorization.
  • This list of acceptable alternative lots of primer and probe materials and/or positive control materials
  • will be available at:
  • https://www.cdc.gov/coronavirus/2019-nCoV/lab/virus-requests.html
  • Primers and Probes:
  • Catalog #2019-nCoVEUA-01 Diagnostic Panel Box #1:
  • Reagent Label Part # Description Quantity /
  • Tube
  • Reactions /
  • Tube
  • 2019-nCoV_N1 RV202001
  • RV202015 2019-nCoV_N1 Combined Primer/Probe Mix 22.5 nmol 1000
  • 2019-nCoV_N2 RV202002
  • RV202016 2019-nCoV_N2 Combined Primer/Probe Mix 22.5 nmol 1000
  • RP RV202004
  • RV202018 Human RNase P Combined Primer/Probe Mix 22.5 nmol 1000
  • Positive Control (either of the following products are acceptable)
  • Catalog #2019-nCoVEUA-01 Diagnostic Panel Box #2:
  • Reagent
  • Label Part # Description Quantity Notes
  • nCoVPC RV202005
  • 2019-nCoV Positive Control (nCoVPC)
  • For use as a positive control with the CDC 2019-
  • nCoV Real-Time RT-PCR Diagnostic Panel
  • procedure. The nCoVPC contains noninfectious
  • positive control material supplied in a dried state
  • and must be resuspended before use. nCoVPC
  • consists of in vitro transcribed RNA. nCoVPC will
  • yield a positive result with each assay in the
  • 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • including RP.
  • 4 tubes
  • Provides
  • (800) 5 µL
  • test reactions 
  • 6
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Catalog #VTC-04 CDC 2019-nCoV Positive Control (nCoVPC)
  • Reagent
  • Label Part # Description Quantity Notes
  • nCoVPC RV202005
  • 2019-nCoV Positive Control (nCoVPC)
  • For use as a positive control with the CDC 2019-
  • nCoV Real-Time RT-PCR Diagnostic Panel
  • procedure. The nCoVPC contains noninfectious
  • positive control material supplied in a dried state
  • and must be resuspended before use. nCoVPC
  • consists of in vitro transcribed RNA. nCoVPC will
  • yield a positive result with each assay in the
  • 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • including RP.
  • 4 tubes
  • Provides
  • (800) 5 µL
  • test reactions
  • Materials Required (But Not Provided)
  • Human Specimen Control (HSC)
  • Description Quantity CDC Catalog No.
  • Manufactured by CDC. For use as a nucleic acid extraction procedural
  • control to demonstrate successful recovery of nucleic acid as well as
  • extraction reagent integrity. The HSC consists of noninfectious (betaPropiolactone treated) cultured human cell material supplied as a liquid
  • suspended in 0.01 M PBS at pH 7.2-7.4.
  • 10 vials x 500uL KT0189
  • Acceptable alternatives to HSC:
  • • Negative human specimen material: Laboratories may prepare a volume of human specimen
  • material (e.g., human sera or pooled leftover negative respiratory specimens) to extract and
  • run alongside clinical samples as an extraction control. This material should be prepared in
  • sufficient volume to be used across multiple runs. Material should be tested prior to use as
  • the extraction control to ensure it generates the expected results for the HSC listed in these
  • instructions for use.
  • • Contrived human specimen material: Laboratories may prepare contrived human specimen
  • materials by suspending any human cell line (e.g., A549, Hela, or 293) in PBS. This material
  • should be prepared in sufficient volume to be used across multiple runs. Material should be
  • tested prior to use as the extraction control to ensure it generates the expected results for
  • the HSC listed in these instructions for use.
  • CDC will maintain on its website a list of commercially alternative extraction controls, if applicable,
  • that are acceptable for use with this assay under CDC’s Emergency Use Authorization, at:
  • https://www.cdc.gov/coronavirus/2019-nCoV/lab/virus-requests.html
  •  
  • 7
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • rRT-PCR Enzyme Mastermix Options
  • Reagent Quantity Catalog No.
  • Quantabio qScript XLT One-Step RT-qPCR ToughMix
  • 100 x 20 μL rxns
  • (1 x 1 mL) 95132-100
  • 2000 x 20 μL rxns
  • (1 x 20 mL) 95132-02K
  • 500 x 20 μL rxns
  • (5 x 1 mL) 95132-500
  • Quantabio UltraPlex 1-Step ToughMix (4X)
  • 100 x 20 µL rxns
  • (500 µL) 95166-100
  • 500 x 20 μL rxns
  • (5 x 500 µL) 95166-500
  • 1000 x 20 μL rxns
  • (1 x 5 mL) 95166-01K
  • Promega GoTaq® Probe 1- Step RT-qPCR System
  • 200 x 20 μL rxns
  • (2 mL) A6120
  • 1250 x 20 μL rxns
  • 12.5 mL A6121
  • Thermofisher TaqPath™ 1-Step RT-qPCR Master Mix, CG
  • 1000 reactions A15299
  • 2000 reactions A15300
  • 8
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • RNA Extraction Options
  • For each of the kits listed below, CDC has confirmed that the external lysis buffer is effective for
  • inactivation of SARS-CoV-2.
  • Instrument/Manufacturer Extraction Kit Catalog No.
  • QIAGEN
  • 2
  • QIAmp DSP Viral RNA Mini Kit 50 extractions (61904)
  • 2
  • QIAamp Viral RNA Mini Kit 50 extractions (52904)
  • 250 extractions (52906)
  • QIAGEN EZ1 Advanced XL
  • 2
  • EZ1 DSP Virus Kit
  • 48 extractions (62724)
  • Buffer AVL (19073 or 19089)
  • EZ1 Advanced XL DSP Virus Card (9018703)
  • 2
  • EZ1 Virus Mini Kit v2.0
  • 48 extractions (955134)
  • Buffer AVL (19073 or 19089)
  • EZ1 Advanced XL Virus Card v2.0 (9018708)
  • Roche MagNA Pure 24 2
  • MagNA Pure 24 Total NA
  • Isolation Kit
  • 96 extractions (07 658 036 001)
  • External Lysis Buffer (06 374 913 001, 12 239 469
  • 103, 03 246 779 001 or 03 246 752 001)
  • Roche MagNA Pure 96 2
  • DNA and Viral NA Small Volume
  • Kit
  • 576 extractions (06 543 588 001)
  • External Lysis Buffer (06 374 913 001, 12 239 469
  • 103, 03 246 779 001 or 03 246 752 001)
  • 1
  • Roche MagNA Pure LC 2
  • Total Nucleic Acid Kit 192 extractions (03 038 505 001)
  • 1
  • Roche MagNA Pure Compact 2
  • Nucleic Acid Isolation Kit I 32 extractions (03 730 964 001)
  • Promega Maxwell® RSC 48 3
  • Maxwell® RSC Viral Total
  • Nucleic Acid Purification Kit
  • 48 extractions (AS1330)
  • 144 extractions (ASB1330)
  • 1
  • QIAGEN QIAcube
  • 2
  • QIAmp DSP Viral RNA Mini Kit 50 extractions (61904)
  • 2
  • QIAamp Viral RNA Mini Kit
  • 50 extractions (52904)
  • 250 extractions (52906)
  • 1, 3bioMérieux NucliSENS®
  • easyMAG®
  • and
  • 1, 3bioMérieux EMAG®
  • (Automated magnetic
  • extraction reagents sold
  • separately. Both instruments
  • use the same reagents and
  • disposables, with the
  • exception of tips.)
  • EasyMAG® Magnetic Silica (280133)
  • EasyMAG® Lysis Buffer (280134)
  • EasyMAG® Lysis Buffer, 2 mL (200292)
  • EasyMAG® Wash Buffers 1,2, and 3
  • (280130, 280131, 280132)
  • EasyMAG® Disposables (280135)
  • Biohit Pipette Tips (easyMAG® only) (280146)
  • EMAG®1000μL Tips (418922)
  • 1
  • Equivalence and performance of these extraction platforms for extraction of viral RNA were demonstrated with the CDC
  • Human Influenza Virus Real-Time RT-PCR Diagnostic Panel (K190302). Performance characteristics of these extraction
  • platforms with 2019-nCoV (SARS CoV-2) have not been demonstrated.
  • 9
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 2 CDC has confirmed that the external lysis buffer used with this extraction method is effective for inactivation of SARSCoV-2. 3 CDC has compared the concentration of inactivating agent in the lysis buffer used with this extraction method and has
  • determined the concentration to be within the range of concentrations found effective in inactivation of SARS-CoV-2.
  • Alternative to Extraction:
  • If a laboratory cannot access adequate extraction reagents to support testing demand due to the
  • global shortage of reagents, CDC has evaluated a heat treatment procedure for upper respiratory
  • specimens using the Quantabio UltraPlex 1-Step ToughMix (4X), CG. Though performance was
  • comparable, this method has been evaluated with a limited number of clinical specimens and a
  • potential reduction in sensitivity due to carryover of inhibitory substances or RNA degradation cannot
  • be ruled out. It should only be used when a jurisdiction determines that the testing need is great
  • enough to justify the risk of a potential loss of sensitivity. Heat-treated specimens generating
  • inconclusive or invalid results should be extracted with an authorized extraction method prior to
  • retesting. Details and procedure for the heat treatment alternative to extraction may be found in
  • Appendix A.
  • Equipment and Consumables Required (But Not Provided)
  •  Vortex mixer
  •  Microcentrifuge
  •  Micropipettes (2 or 10 μL, 200 μL and 1000 μL)
  •  Multichannel micropipettes (5-50 μl)
  •  Racks for 1.5 mL microcentrifuge tubes
  •  2 x 96-well -20°C cold blocks
  •  7500 Fast Dx Real-Time PCR Systems with SDS 1.4 software (Applied Biosystems; catalog
  • #4406985 or #4406984)
  •  Extraction systems (instruments): QIAGEN EZ1 Advanced XL, QIAGEN QIAcube, Roche MagNA
  • Pure 24, Roche MagNA Pure 96, Promega Maxwell® RSC 48, Roche MagNA Pure LC, Roche
  • MagNA Pure Compact, bioMérieux easyMAG, and bioMérieux EMAG
  •  Molecular grade water, nuclease-free
  •  10% bleach (1:10 dilution of commercial 5.25-6.0% hypochlorite bleach)
  •  DNAZapTM (Ambion, cat. #AM9890) or equivalent
  •  RNase AWAY™ (Fisher Scientific; cat. #21-236-21) or equivalent
  •  Disposable powder-free gloves and surgical gowns
  •  Aerosol barrier pipette tips
  •  1.5 mL microcentrifuge tubes (DNase/RNase free)
  •  0.2 mL PCR reaction plates (Applied Biosystems; catalog #4346906 or #4366932)
  •  MicroAmp Optical 8-cap Strips (Applied Biosystems; catalog #4323032)
  • Qualifying Alternative Components:
  • If a laboratory modifies this test by using unauthorized, alternative components (e.g., extraction
  • methods or PCR instruments), the modified test is not authorized under this EUA. FDA’s Policy for
  • Diagnostic Tests for Coronavirus Disease-2019 during the Public Health Emergency, updated May 11,
  • 2020, does not change this. As part of this policy, FDA does not intend to object when a laboratory
  • modifies an EUA-authorized test, which could include using unauthorized components, without 
  • 10
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • obtaining an EUA or EUA amendment, where the modified test is validated using a bridging study to the
  • EUA-authorized test .
  • Warnings and Precautions
  • • For in vitro diagnostic use (IVD).
  • • This test has not been FDA cleared or approved; this test has been authorized by FDA under
  • an EUA for use by laboratories certified under CLIA, 42 U.S.C. § 263a, to perform high
  • complexity tests.
  • • This test has been authorized only for the detection of nucleic acid from SARSCoV-2, not
  • for any other viruses or pathogens.
  • • This test is only authorized for the duration of the declaration that circumstances exist
  • justifying the authorization of emergency use of in vitro diagnostic tests for detection
  • and/or diagnosis of COVID-19 under Section 564(b)(1) of the Act, 21 U.S.C. § 360bbb3(b)(1), unless the authorization is terminated or revoked sooner.
  • • Follow standard precautions. All patient specimens and positive controls should be considered
  • potentially infectious and handled accordingly.
  • • Do not eat, drink, smoke, apply cosmetics or handle contact lenses in areas where reagents and
  • human specimens are handled.
  • • Handle all specimens as if infectious using safe laboratory procedures. Refer to Interim
  • Laboratory Biosafety Guidelines for Handling and Processing Specimens Associated with 2019-
  • nCoV https://www.cdc.gov/coronavirus/2019-nCoV/lab-biosafety-guidelines.html.
  • • Specimen processing should be performed in accordance with national biological safety
  • regulations.
  • • If infection with 2019-nCoV is suspected based on current clinical and epidemiological screening
  • criteria recommended by public health authorities, specimens should be collected with
  • appropriate infection control precautions.
  • • Performance characteristics have been determined with human upper respiratory specimens and
  • lower respiratory tract specimens from human patients with signs and symptoms of respiratory
  • infection.
  • • Perform all manipulations of live virus samples within a Class II (or higher) biological safety
  • cabinet (BSC).
  • • Use personal protective equipment such as (but not limited to) gloves, eye protection, and lab
  • coats when handling kit reagents while performing this assay and handling materials including
  • samples, reagents, pipettes, and other equipment and reagents.
  • • Amplification technologies such as PCR are sensitive to accidental introduction of PCR product
  • from previous amplifications reactions. Incorrect results could occur if either the clinical
  • specimen or the real-time reagents used in the amplification step become contaminated by
  • accidental introduction of amplification product (amplicon). Workflow in the laboratory should
  • proceed in a unidirectional manner.
  •  Maintain separate areas for assay setup and handling of nucleic acids.
  •  Always check the expiration date prior to use. Do not use expired reagents. Do not
  • substitute or mix reagents from different kit lots or from other manufacturers.
  •  Change aerosol barrier pipette tips between all manual liquid transfers.
  •  During preparation of samples, compliance with good laboratory techniques is essential
  • to minimize the risk of cross-contamination between samples and the inadvertent 
  • 11
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • introduction of nucleases into samples during and after the extraction procedure. Proper
  • aseptic technique should always be used when working with nucleic acids.
  •  Maintain separate, dedicated equipment (e.g., pipettes, microcentrifuges) and supplies
  • (e.g., microcentrifuge tubes, pipette tips) for assay setup and handling of extracted
  • nucleic acids.
  •  Wear a clean lab coat and powder-free disposable gloves (not previously worn) when
  • setting up assays.
  •  Change gloves between samples and whenever contamination is suspected.
  •  Keep reagent and reaction tubes capped or covered as much as possible.
  •  Primers, probes (including aliquots), and enzyme master mix must be thawed and
  • maintained on a cold block at all times during preparation and use.
  •  Work surfaces, pipettes, and centrifuges should be cleaned and decontaminated with
  • cleaning products such as 10% bleach, DNAZap™, or RNase AWAY™ to minimize risk of
  • nucleic acid contamination. Residual bleach should be removed using 70% ethanol.
  • • RNA should be maintained on a cold block or on ice during preparation and use to ensure
  • stability.
  • • Dispose of unused kit reagents and human specimens according to local, state, and federal
  • regulations.
  • Reagent Storage, Handling, and Stability
  • • Store all dried primers and probes and the positive control, nCoVPC, at 2-8°C until re-hydrated for
  • use. Store liquid HSC control materials at ≤ -20°C.
  • Note: Storage information is for CDC primer and probe materials obtained through the
  • International Reagent Resource. If using commercial primers and probes, please refer to the
  • manufacturer’s instructions for storage and handling.
  • • Always check the expiration date prior to use. Do not use expired reagents.
  • • Protect fluorogenic probes from light.
  • • Primers, probes (including aliquots), and enzyme master mix must be thawed and kept on a cold
  • block at all times during preparation and use.
  • • Do not refreeze probes.
  • • Controls and aliquots of controls must be thawed and kept on ice at all times during preparation
  • and use. 
  • 12
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Specimen Collection, Handling, and Storage
  • Inadequate or inappropriate specimen collection, storage, and transport are likely to yield false test
  • results. Training in specimen collection is highly recommended due to the importance of specimen
  • quality. CLSI MM13-A may be referenced as an appropriate resource.
  •  Collecting the Specimen
  • • Refer to Interim Guidelines for Collecting, Handling, and Testing Clinical Specimens from
  • Patients Under Investigation (PUIs) for 2019 Novel Coronavirus (2019-nCoV)
  • https://www.cdc.gov/coronavirus/2019-nCoV/guidelines-clinical-specimens.html
  • • Follow specimen collection device manufacturer instructions for proper collection methods.
  • • Swab specimens should be collected using only swabs with a synthetic tip, such as nylon or
  • Dacron®, and an aluminum or plastic shaft. Calcium alginate swabs are unacceptable and
  • cotton swabs with wooden shafts are not recommended. Place swabs immediately into
  • sterile tubes containing 1-3 ml of appropriate transport media, such as viral transport media
  • (VTM).
  •  Transporting Specimens
  • • Specimens must be packaged, shipped, and transported according to the current edition of
  • the International Air Transport Association (IATA) Dangerous Goods Regulation. Follow
  • shipping regulations for UN 3373 Biological Substance, Category B when sending potential
  • 2019-nCoV specimens. Store specimens at 2-8°C and ship overnight to CDC on ice pack. If a
  • specimen is frozen at -70°C or lower, ship overnight to CDC on dry ice.
  •  Storing Specimens
  • • Specimens can be stored at 2-8oC for up to 72 hours after collection.
  • • If a delay in extraction is expected, store specimens at -70oC or lower.
  • • Extracted nucleic acid should be stored at -70oC or lower.
  • 13
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Specimen Referral to CDC
  • For state and local public health laboratories:
  • • Ship all specimens overnight to CDC.
  • • Ship frozen specimens on dry ice and non-frozen specimens on cold packs.
  • • Refer to the International Air Transport Association (IATA - www.iata.org) for requirements
  • for shipment of human or potentially infectious biological specimens. Follow shipping
  • regulations for UN 3373 Biological Substance, Category B when sending potential 2019-nCoV
  • specimens.
  • • Prior to shipping, notify CDC Division of Viral Diseases (see contact information below) that
  • you are sending specimens.
  • • Send all samples to the following recipient:
  • Centers for Disease Control and Prevention
  • c/o STATT
  • Attention: Unit 66
  • 1600 Clifton Rd., Atlanta, GA 30329-4027
  • Phone: (404) 639-3931
  • The emergency contact number for CDC Emergency Operations Center (EOC) is
  • 770-488-7100.
  • All other laboratories that are CLIA certified and meet requirements to perform high complexity
  • testing:
  • • Please notify your state and/or local public health laboratory for specimen referral and
  • confirmatory testing guidance.
  • Reagent and Controls Preparation
  • NOTE: Storage information is for materials obtained through the CDC International Regent Resource.
  • If using commercial products for testing, please refer to the manufacturer’s instructions for storage,
  • handling, and preparation instructions.
  • Primer and Probe Preparation:
  • 1) Upon receipt, store dried primers and probes at 2-8°C.
  • 2) Precautions: These reagents should only be handled in a clean area and stored at appropriate
  • temperatures (see below) in the dark. Freeze-thaw cycles should be avoided. Maintain cold
  • when thawed.
  • 3) Using aseptic technique, suspend dried reagents in 1.5 mL of nuclease-free water and allow
  • to rehydrate for 15 min at room temperature in the dark.
  • 4) Mix gently and aliquot primers/probe in 300 μL volumes into 5 pre-labeled tubes. Store a
  • single, working aliquot of primers/probes at 2-8oC in the dark. Store remaining aliquots at ≤ -
  • 20oC in a non-frost-free freezer. Do not refreeze thawed aliquots (stable for up to 4 months at
  • 2-8oC).
  • 14
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 2019-nCoV Positive Control (nCoVPC) Preparation:
  • 1) Precautions: This reagent should be handled with caution in a dedicated nucleic acid handling
  • area to prevent possible contamination. Freeze-thaw cycles should be avoided. Maintain on
  • ice when thawed.
  • 2) Resuspend dried reagent in each tube in 1 mL of nuclease-free water to achieve the proper
  • concentration. Make single use aliquots (approximately 30 μL) and store at ≤ -70oC.
  • 3) Thaw a single aliquot of diluted positive control for each experiment and hold on ice until
  • adding to plate. Discard any unused portion of the aliquot.
  • Human Specimen Control (HSC) (not provided)
  • 1) Human Specimen Control (HSC) or one of the listed acceptable alternative extraction controls
  • must be extracted and processed with each specimen extraction run.
  • 2) Refer to the Human Specimen Control (HSC) package insert for instructions for use.
  • No Template Control (NTC) (not provided)
  • 1) Sterile, nuclease-free water
  • 2) Aliquot in small volumes
  • 3) Used to check for contamination during specimen extraction and/or plate set-up
  • General Preparation
  • Equipment Preparation
  • Clean and decontaminate all work surfaces, pipettes, centrifuges, and other equipment prior to use.
  • Decontamination agents should be used including 10% bleach, 70% ethanol, and DNAzap™, or RNase
  • AWAY™ to minimize the risk of nucleic acid contamination.
  • Nucleic Acid Extraction
  • Performance of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is dependent upon the amount
  • and quality of template RNA purified from human specimens. The following commercially available
  • RNA extraction kits and procedures have been qualified and validated for recovery and purity of RNA
  • for use with the panel:
  • Qiagen QIAamp® DSP Viral RNA Mini Kit or QIAamp® Viral RNA Mini Kit
  • Recommendation(s): Utilize 100 μL of sample and elute with 100 μL of buffer or utilize 140 μL of
  • sample and elute with 140 μL of buffer.
  • Qiagen EZ1 Advanced XL
  • Kit: Qiagen EZ1 DSP Virus Kit and Buffer AVL (supplied separately) for offboard lysis
  • Card: EZ1 Advanced XL DSP Virus Card
  • Recommendation(s): Add 120 μL of sample to 280 μL of pre-aliquoted Buffer AVL (total input sample
  • volume is 400 μL). Proceed with the extraction on the EZ1 Advanced XL. Elution volume is 120 μL.
  • 15
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Kit: Qiagen EZ1 Virus Mini Kit v2.0 and Buffer AVL (supplied separately) for offboard lysis
  • Card: EZ1 Advanced XL Virus Card v2.0
  • Recommendation(s): Add 120 μL of sample to 280 μL of pre-aliquoted Buffer AVL (total input sample
  • volume is 400 μL). Proceed with the extraction on the EZ1 Advanced XL. Elution volume is 120 μL.
  • Roche MagNA Pure 96
  • Kit: Roche MagNA Pure 96 DNA and Viral NA Small Volume Kit
  • Protocol: Viral NA Plasma Ext LysExt Lys SV 4.0 Protocol or Viral NA Plasma Ext Lys SV Protocol
  • Recommendation(s): Add 100 μL of sample to 350 μL of pre-aliquoted External Lysis Buffer (supplied
  • separately) (total input sample volume is 450 μL). Proceed with the extraction on the MagNA Pure 96.
  • (Internal Control = None). Elution volume is 100 μL.
  • Roche MagNA Pure 24
  • Kit: Roche MagNA Pure 24 Total NA Isolation Kit
  • Protocol: Pathogen 1000 2.0 Protocol
  • Recommendation(s): Add 100 µL of sample to 400 µL of pre-aliquoted External Lysis Buffer (supplied
  • separately) (total input sample volume is 500 µL). Proceed with the extraction on the MagNA Pure 24.
  • (Internal Control = None). Elution volume is 100 µL.
  • Promega Maxwell® RSC 48
  • Kit: Promega Maxwell® Viral Total Nucleic Acid Purification Kit
  • Protocol: Viral Total Nucleic Acid
  • Recommendation(s): Add 120 µL of sample to 330 µL of pre-aliquoted External Lysis Buffer (300 µL
  • Lysis Buffer plus 30 µL Proteinase K; supplied within the kit) (total input volume is 450 µL). Proceed
  • with the extraction on the Maxwell® RSC 48. Elution volume is 75 µL.
  • Equivalence and performance of the following extraction platforms were demonstrated with the CDC
  • Human Influenza Virus Real-Time RT-PCR Diagnostic Panel (K190302) and based on those data are
  • acceptable for use with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel.
  • QIAGEN QIAcube
  • Kit: QIAGEN QIAamp® DSP Viral RNA Mini Kit or QIAamp® Viral RNA Mini Kit
  • Recommendations: Utilize 140 μL of sample and elute with 100 μL of buffer.
  • Roche MagNA Pure LC
  • Kit: Roche MagNA Pure Total Nucleic Acid Kit
  • Protocol: Total NA External_lysis
  • Recommendation(s): Add 100 μL of sample to 300 μL of pre-aliquoted TNA isolation kit lysis buffer
  • (total input sample volume is 400 μL). Elution volume is 100 μL.
  • Roche MagNA Pure Compact
  • Kit: Roche MagNA Pure Nucleic Acid Isolation Kit I
  • Protocol: Total_NA_Plasma100_400
  • Recommendation(s): Add 100 μL of sample to 300 μL of pre-aliquoted TNA isolation kit lysis buffer
  • (total input sample volume is 400 μL). Elution volume is 100 μL. 
  • 16
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • bioMérieux NucliSENS® easyMAG® Instrument
  • Protocol: General protocol (not for blood) using “Off-board Lysis” reagent settings.
  • Recommendation(s): Add 100 μL of sample to 1000 μL of pre-aliquoted easyMAG lysis buffer (total
  • input sample volume is 1100 μL). Incubate for 10 minutes at room temperature. Elution volume is 100
  • μL.
  • bioMérieux EMAG® Instrument
  • Protocol: Custom protocol: CDC Flu V1 using “Off-board Lysis” reagent settings.
  • Recommendation(s): Add 100 μL of samples to 2000 μL of pre-aliquoted easyMAG lysis buffer (total
  • input sample volume is 2100 μL). Incubate for 10 minutes at room temperature. Elution volume is 100
  • μL. The custom protocol, CDC Flu V1, is programmed on the bioMérieux EMAG® instrument with the
  • assistance of a bioMérieux service representative. Installation verification is documented at the time of
  • installation. Laboratories are recommended to retain a record of the step-by-step verification of the
  • bioMérieux custom protocol installation procedure.
  • Manufacturer’s recommended procedures (except as noted in recommendations above) are to be
  • followed for sample extraction. HSC must be included in each extraction batch.
  • Disclaimer: Names of vendors or manufacturers are provided as examples of suitable product sources. Inclusion
  • does not imply endorsement by the Centers for Disease Control and Prevention.
  • Assay Set Up
  • Reaction Master Mix and Plate Set Up
  • Note: Plate set-up configuration can vary with the number of specimens and workday
  • organization. NTCs and nCoVPCs must be included in each run.
  • 1) In the reagent set-up room clean hood, place rRT-PCR buffer, enzyme, and primer/probes on
  • ice or cold-block. Keep cold during preparation and use.
  • 2) Mix buffer, enzyme, and primer/probes by inversion 5 times.
  • 3) Centrifuge reagents and primers/probes for 5 seconds to collect contents at the bottom of
  • the tube, and then place the tube in a cold rack.
  • 4) Label one 1.5 mL microcentrifuge tube for each primer/probe set.
  • 5) Determine the number of reactions (N) to set up per assay. It is necessary to make excess
  • reaction mix for the NTC, nCoVPC, HSC (if included in the RT-PCR run), and RP reactions and
  • for pipetting error. Use the following guide to determine N:
  • • If number of samples (n) including controls equals 1 through 14, then N = n + 1
  • • If number of samples (n) including controls is 15 or greater, then N = n + 2
  • 7) For each primer/probe set, calculate the amount of each reagent to be added for each
  • reaction mixture (N = # of reactions).
  • 17
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Thermofisher TaqPath™ 1-Step RT-qPCR Master Mix
  • Step # Reagent Vol. of Reagent Added
  • per Reaction
  • 1 Nuclease-free Water N x 8.5 µL
  • 2 Combined Primer/Probe Mix N x 1.5 µL
  • 3 TaqPathTM 1-Step RT-qPCR Master Mix (4x) N x 5.0 µL
  • Total Volume N x 15.0 µL
  • Promega GoTaq® Probe 1- Step RT-qPCR System
  • Step # Reagent Vol. of Reagent Added
  • per Reaction
  • 1 Nuclease-free Water N x 3.1 µL
  • 2 Combined Primer/Probe Mix N x 1.5 µL
  • 3 GoTaq Probe qPCR Master Mix with dUTP N x 10.0 µL
  • 4 Go Script RT Mix for 1-Step RT-qPCR N x 0.4 µL
  • Total Volume N x 15.0 µL
  • Quantabio qScript XLT One-Step RT-qPCR ToughMix
  • Step # Reagent Vol. of Reagent Added
  • per Reaction
  • 1 Nuclease-free Water N x 3.5 µL
  • 2 Combined Primer/Probe Mix N x 1.5 µL
  • 3 qScript XLT One-Step RT-qPCR ToughMix
  • (2X)
  • N x 10.0 µL
  • Total Volume N x 15.0 µL
  • Quantabio UltraPlex 1-Step ToughMix (4X)
  • Step # Reagent Vol. of Reagent Added
  • per Reaction
  • 1 Nuclease-free Water N x 8.5 µL
  • 2 Combined Primer/Probe Mix N x 1.5 µL
  • 3 UltraPlex 1-Step ToughMix (4X) N x 5.0 µL
  • Total Volume N x 15.0 µL
  • 18
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 8) Dispense reagents into each respective labeled 1.5 mL microcentrifuge tube. After addition of
  • the reagents, mix reaction mixtures by pipetting up and down. Do not vortex.
  • 9) Centrifuge for 5 seconds to collect contents at the bottom of the tube, and then place the
  • tube in a cold rack.
  • 10) Set up reaction strip tubes or plates in a 96-well cooler rack.
  • 11) Dispense 15 µL of each master mix into the appropriate wells going across the row as shown
  • below (Figure 1):
  • Figure 1: Example of Reaction Master Mix Plate Set-Up
  • 1 2 3 4 5 6 7 8 9 10 11 12
  • A N1 N1 N1 N1 N1 N1 N1 N1 N1 N1 N1 N1
  • B N2 N2 N2 N2 N2 N2 N2 N2 N2 N2 N2 N2
  • C RP RP RP RP RP RP RP RP RP RP RP RP
  • D
  • E
  • F
  • G
  • H
  • 12) Prior to moving to the nucleic acid handling area, prepare the No Template Control (NTC)
  • reactions for column #1 in the assay preparation area.
  • 13) Pipette 5 µL of nuclease-free water into the NTC sample wells (Figure 2, column 1). Securely
  • cap NTC wells before proceeding.
  • 14) Cover the entire reaction plate and move the reaction plate to the specimen nucleic acid
  • handling area.
  • Nucleic Acid Template Addition
  • 1) Gently vortex nucleic acid sample tubes for approximately 5 seconds.
  • 2) Centrifuge for 5 seconds to collect contents at the bottom of the tube.
  • 3) After centrifugation, place extracted nucleic acid sample tubes in the cold rack.
  • 4) Samples should be added to columns 2-11 (column 1 and 12 are for controls) to the specific
  • assay that is being tested as illustrated in Figure 2. Carefully pipette 5.0 µL of the first sample
  • into all the wells labeled for that sample (i.e. Sample “S1” down column #2). Keep other
  • sample wells covered during addition. Change tips after each addition.
  • 5) Securely cap the column to which the sample has been added to prevent cross contamination
  • and to ensure sample tracking.
  • 6) Change gloves often and when necessary to avoid contamination.
  • 7) Repeat steps #4 and #5 for the remaining samples.
  • 19
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 8) If necessary, add 5 µL of Human Specimen Control (HSC) extracted sample to the HSC wells
  • (Figure 2, column 11). Securely cap wells after addition. NOTE: Per CLIA regulations, HSC
  • must be tested at least once per day.
  • 9) Cover the entire reaction plate and move the reaction plate to the positive template control
  • handling area.
  • Assay Control Addition
  • 1) Pipette 5 µL of nCoVPC RNA to the sample wells of column 12 (Figure 2). Securely cap wells
  • after addition of the control RNA.
  • NOTE: If using 8-tube strips, label the TAB of each strip to indicate sample position. DO NOT
  • LABEL THE TOPS OF THE REACTION TUBES!
  • 2) Briefly centrifuge reaction tube strips for 10-15 seconds. After centrifugation return to cold
  • rack.
  • NOTE: If using 96-well plates, centrifuge plates for 30 seconds at 500 x g, 4°C.
  • Figure 2. 2019-nCoV rRT-PCR Diagnostic Panel: Example of Sample and Control Set-up
  • 1 2 3 4 5 6 7 8 9 10 11a 12
  • A NTC S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 nCoV PC
  • B NTC S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 nCoV PC
  • C NTC S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 nCoV PC
  • D
  • E
  • F
  • G
  • H
  • a
  • Replace the sample in this column with extracted HSC if necessary 
  • 20
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Create a Run Template on the Applied Biosystems 7500 Fast Dx Real-time PCR
  • Instrument (Required if no template exists)
  • If the template already exists on your instrument, please proceed to the RUNNING A TEST section.
  • 1) Launch the Applied Biosystems 7500 Fast Dx Real-time PCR Instrument by double clicking on the
  • Applied Biosystems 7500 Fast Dx System icon on the desktop.
  • 2) A new window should appear, select Create New Document from the menu.
  • Figure 3. New Document Wizard Window
  • 3) The New Document Wizard screen in Figure 3 will appear. Select:
  • a. Assay: Standard Curve (Absolute Quantitation)
  • b. Container: 96-Well Clear
  • c. Template: Blank Document
  • d. Run Mode: Standard 7500
  • e. Operator: Your Name
  • f. Comments: SDS v1.4
  • g. Plate Name: Your Choice
  • 4) After making selections click Next at the bottom of the window.
  • Make sure to change
  • Run Mode to
  • STANDARD 7500
  • 21
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Figure 4. Creating New Detectors
  • NOTE: ROX is the default passive reference. This will be changed to “none” in step 12.
  • 5) After selecting next, the Select Detectors screen (Figure 4) will appear.
  • 6) Click the New Detector button (see Figure 4).
  • 7) The New Detector window will appear (Figure 5). A new detector will need to be defined for
  • each primer and probe set. Creating these detectors will enable you to analyze each primer and
  • probe set individually at the end of the reaction.
  • 22
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Figure 5. New Detector Window
  • 8) Start by creating the N1 Detector. Include the following:
  • a. Name: N1
  • b. Description: leave blank
  • c. Reporter Dye: FAM
  • d. Quencher Dye: (none)
  • e. Color: to change the color of the detector indicator do the following:
  • ⇒ Click on the color square to reveal the color chart
  • ⇒ Select a color by clicking on one of the squares
  • ⇒ After selecting a color click OK to return to the New Detector screen
  • f. Click the OK button of the New Detector screen to return to the screen shown in Figure 4.
  • 9) Repeat step 6-8 for each target in the panel.
  • Name Reporter Dye Quencher Dye
  • N1 FAM (none)
  • N2 FAM (none)
  • RP FAM (none)
  • 23
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 10) After each Detector is added, the Detector Name, Description, Reporter and Quencher fields
  • will become populated in the Select Detectors screen (Figure 6).
  • 11) Before proceeding, the newly created detectors must be added to the document. To add the
  • new detectors to the document, click ADD (see Figure 6). Detector names will appear on the
  • right-hand side of the Select Detectors window (Figure 6).
  • Figure 6. Adding New Detectors to Document
  • 12) Once all detectors have been added, select (none) for Passive Reference at the top right-hand
  • drop-down menu (Figure 7).
  • Figure 7. Select Passive Reference
  • Passive reference should be set to “(none)” as described above.
  • 24
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 13)Click Next at the bottom of the Select Detectors window to proceed to the Set Up Sample Plate
  • window (Figure 8).
  • 14)In the Set Up Sample Plate window (Figure 8), use your mouse to select row A from the lower
  • portion of the window, in the spreadsheet (see Figure 8).
  • 15)In the top portion of the window, select detector N1. A check will appear next to the detector
  • you have selected (Figure 8). You will also notice the row in the spreadsheet will be populated
  • with a colored “U” icon to indicate which detector you’ve selected.
  • 16)Repeat step 14-15 for each detector that will be used in the assay.
  • Figure 8. Sample Plate Set-up
  • 17) Select Finish after detectors have been assigned to their respective rows. (Figure 9).
  • Figure 9. Finished Plate Set-up
  • 25
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 18) After clicking “Finish”, there will be a brief pause allowing the Applied Biosystems 7500 Fast Dx
  • to initialize. This initialization is followed by a clicking noise. Note: The machine must be turned
  • on for initialization.
  • 19) After initialization, the Plate tab of the Setup (Figure 10) will appear.
  • 20)Each well of the plate should contain colored U icons that correspond with the detector labels
  • that were previously chosen. To confirm detector assignments, select Tools from the file menu,
  • then select Detector Manager.
  • Figure 10. Plate Set-up Window
  • 26
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 21) The Detector Manager window will appear (Figure 11).
  • Figure 11. Detector Manager Window
  • 22)Confirm all detectors are included and that each target has a Reporter set to FAM and the
  • Quencher is set to (none).
  • 23)If all detectors are present, select Done. The detector information has been created and
  • assigned to wells on the plate.
  • Defining the Instrument Settings
  • 1) After detectors have been created and assigned, proceed to instrument set up.
  • 2) Select the Instrument tab to define thermal cycling conditions.
  • 3) Modify the thermal cycling conditions as follows (Figure 12):
  • Thermofisher TaqPath™ 1-Step RT-qPCR Master Mix, CG
  • a. In Stage 1, Set to 2 min at 25°C; 1 Rep.
  • b. In Stage 2, Set to 15 min at 50°C; 1 Rep.
  • c. In Stage 3, Set to 2 min at 95°C, 1 Rep.
  • d. In Stage 4, Step 1 set to 3 sec at 95°C.
  • e. In Stage 4, Step 2 set to 30 sec at 55.0°C.
  • f. In Stage 4, Reps should be set to 45.
  • g. Under Settings (Figure 12), bottom left-hand box, change volume to 20 µL.
  • h. Under Settings, Run Mode selection should be Standard 7500.
  • i. Step 2 of Stage 4 should be highlighted in yellow to indicate data collection (see Figure
  • 12).
  • OR
  • 27
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Quantabio qScriptTM XLT One-Step RT-qPCR ToughMix or UltraPlex 1-Step ToughMix (4X)
  • a. In Stage 1, Set to 10 min at 50°C; 1 Rep.
  • b. In Stage 2, Set to 3 min at 95°C, 1 Rep.
  • c. In Stage 3, Step 1 set to 3 sec at 95°C.
  • d. In Stage 3, Step 2 set to 30 sec at 55.0°C.
  • e. In Stage 3, Reps should be set to 45.
  • f. Under Settings (Figure 12), bottom left-hand box, change volume to 20 µL.
  • g. Under Settings, Run Mode selection should be Standard 7500.
  • h. Step 2 of Stage 3 should be highlighted in yellow to indicate data collection (see Figure
  • 12).
  • OR
  • Promega GoTaq® Probe 1-Step RT-qPCR System
  • a. In Stage 1, Set to 15 min at 45°C; 1 Rep.
  • b. In Stage 2, Set to 2 min at 95°C, 1 Rep.
  • c. In Stage 3, Step 1 set to 3 sec at 95°C.
  • d. In Stage 3, Step 2 set to 30 sec at 55.0°C.
  • e. In Stage 3, Reps should be set to 45.
  • f. Under Settings (Figure 12), bottom left-hand box, change volume to 20 µL.
  • g. Under Settings, Run Mode selection should be Standard 7500.
  • h. Step 2 of Stage 3 should be highlighted in yellow to indicate data collection (see Figure
  • 12).
  • 28
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Figure 12. Instrument Window
  • 4) After making changes to the Instrument tab, the template file is ready to be saved. To save the template,
  • select File from the top menu, then select Save As. Since the enzyme options have different instrument
  • settings, it is recommended that the template be saved with a name indicating the enzyme option.
  • 5) Save the template as 2019-nCoV Dx Panel TaqPath or 2019-nCoV Dx Panel Quanta or 2019-nCoV Dx
  • Panel Promega as appropriate in the desktop folder labeled “ABI Run Templates” (you must create this
  • folder). Save as type should be SDS Templates (*.sdt) (Figure 13).
  • Figure 13. Saving Template
  • 29
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Running a Test
  • 1) Turn on the ABI 7500 Fast Dx Real-Time PCR Instrument.
  • 2) Launch the Applied Biosystems 7500 Fast Dx Real-time PCR System by double clicking on the
  • 7500 Fast Dx System icon on the desktop.
  • 3) A new window should appear, select Open Existing Document from the menu.
  • 4) Navigate to select your ABI Run Template folder from the desktop.
  • 5) Double click on the appropriate template file (2019-nCoV Dx Panel TaqPath or 2019-nCoV Dx
  • Panel Quanta or 2019-nCoV Dx Panel Promega)
  • 6) There will be a brief pause allowing the Applied Biosystems 7500 Fast Dx Real-Time PCR
  • Instrument to initialize. This initialization is followed by a clicking noise. Note: The machine must
  • be turned on for initialization.
  • Figure 14. Plate Set-up Window
  • 7) After the instrument initializes, a plate map will appear (Figure 14). The detectors and controls
  • should already be labeled as they were assigned in the original template.
  • 30
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 8) Click the Well Inspector icon from the top menu.
  • 9) Highlight specimen wells of interest on the plate map.
  • 10) Type sample identifiers to Sample Name box in the Well Inspector window (Figure 15).
  • Figure 15. Labeling Wells
  • 11) Repeat steps 9-10 until all sample identifiers are added to the plate setup.
  • 31
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 12) Once all specimen and control identifiers are added click the Close button on the Well Inspector
  • window to return to the Plate set up tab.
  • 13) Click the Instrument tab at the upper left corner.
  • 14) The reaction conditions, volumes, and type of 7500 reaction should already be loaded (Figure
  • 16).
  • Figure 16. Instrument Settings
  • 15) Ensure settings are correct (refer to the Defining Instrument Settings).
  • 16) Before proceeding, the run file must be saved; from the main menu, select File, then Save As.
  • Save in appropriate run folder designation.
  • 17) Load the plate into the plate holder in the instrument. Ensure that the plate is properly aligned
  • in the holder.
  • 18) Once the run file is saved, click the Start button. Note: The run should take approximately 1 hour
  • and 20 minutes to complete.
  • 32
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Data Analysis
  • 1) After the run has completed, select the Results tab at the upper left corner of the software.
  • 2) Select the Amplification Plot tab to view the raw data (Figure 17).
  • Figure 17. Amplification Plot Window
  • 3) Start by highlighting all the samples from the run; to do this, click on the upper left-hand box (a)
  • of the sample wells (Figure 17). All the growth curves should appear on the graph.
  • 4) On the right-hand side of the window (b), the Data drop down selection should be set to Delta
  • Rn vs. Cycle.
  • 5) Select N1 from (c), the Detector drop down menu, using the downward arrow.
  • a. Please note that each detector is analyzed individually to reflect different
  • performance profiles of each primer and probe set.
  • 6) In the Line Color drop down (d), Detector Color should be selected.
  • 7) Under Analysis Settings select Manual Ct (e).
  • b. Do not change the Manual Baseline default numbers.
  • 8) Using the mouse, click and drag the red threshold line until it lies within the exponential phase
  • of the fluorescence curves and above any background signal (Figure 18).
  • c
  • c
  • a
  • b
  • c
  • d
  • e
  • 33
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Figure 18. Amplification Plot
  • 9) Click the Analyze button in the lower right corner of the window. The red threshold line will turn
  • to green, indicating the data has been analyzed.
  • 10) Repeat steps 5-9 to analyze results generated for each set of markers (N1, N2, RP).
  • 11) Save analysis file by selecting File then Save As from the main menu.
  • 12) After completing analysis for each of the markers, select the Report tab above the graph to
  • display the Ct values (Figure 19). To filter report by sample name in ascending or descending
  • order, simply click on Sample Name in the table.
  • Figure 19. Report
  • Exponential
  • PCR Phase
  • Background noise
  • Threshold adjusted
  • to fall within the
  • PCR exponential
  • phase. 
  • 34
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Interpretation of Results and Reporting
  • Extraction and Positive Control Results and Interpretation
  • No Template Control (NTC)
  • The NTC consists of using nuclease-free water in the rRT-PCR reactions instead of RNA. The NTC
  • reactions for all primer and probe sets should not exhibit fluorescence growth curves that cross the
  • threshold line. If any of the NTC reactions exhibit a growth curve that crosses the cycle threshold, sample
  • contamination may have occurred. Invalidate the run and repeat the assay with strict adherence to the
  • guidelines.
  • 2019-nCoV Positive Control (nCoVPC)
  • The nCoVPC consists of in vitro transcribed RNA. The nCoVPC will yield a positive result with the
  • following primer and probe sets: N1, N2, and RP.
  • Human Specimen Control (HSC) (Extraction Control)
  • When HSC is run with the CDC 2019-nCoV rRT-PCR Diagnostic Panel (see previous section on Assay Set
  • Up), the HSC is used as an nucleic acid extraction procedural control to demonstrate successful recovery
  • of nucleic acid as well as extraction reagent integrity. The HSC control consists of noninfectious cultured
  • human cell (A549) material. Purified nucleic acid from the HSC should yield a positive result with the RP
  • primer and probe set and negative results with all 2019-nCoV markers.
  • Expected Performance of Controls Included in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Control
  • Type
  • External
  • Control
  • Name
  • Used to
  • Monitor
  • 2019
  • nCoV_N1
  • 2019
  • nCoV_N2 RP Expected Ct
  • Values
  • Positive nCoVPC
  • Substantial
  • reagent failure
  • including
  • primer and
  • probe integrity
  • + + + < 40.00 Ct
  • Negative NTC
  • Reagent and/or
  • environmental
  • contamination
  • - - - None
  • detected
  • Extraction HSC
  • Failure in lysis
  • and extraction
  • procedure,
  • potential
  • contamination
  • during
  • extraction
  • - - + < 40.00 Ct
  • 35
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • If any of the above controls do not exhibit the expected performance as described, the assay may have
  • been set up and/or executed improperly, or reagent or equipment malfunction could have occurred.
  • Invalidate the run and re-test.
  • RNase P (Extraction Control)
  •  All clinical samples should exhibit fluorescence growth curves in the RNase P reaction that cross
  • the threshold line within 40.00 cycles (< 40.00 Ct), thus indicating the presence of the human
  • RNase P gene. Failure to detect RNase P in any clinical specimens may indicate:
  • − Improper extraction of nucleic acid from clinical materials resulting in loss of RNA and/or
  • RNA degradation.
  • − Absence of sufficient human cellular material due to poor collection or loss of specimen
  • integrity.
  • − Improper assay set up and execution.
  • − Reagent or equipment malfunction.
  •  If the RP assay does not produce a positive result for human clinical specimens, interpret as
  • follows:
  • − If the 2019-nCoV N1 and N2are positive even in the absence of a positive RP, the result
  • should be considered valid. It is possible, that some samples may fail to exhibit RNase P
  • growth curves due to low cell numbers in the original clinical sample. A negative RP signal
  • does not preclude the presence of 2019-nCoV virus RNA in a clinical specimen.
  • − If all 2019-nCoV markers AND RNase P are negative for the specimen, the result should be
  • considered invalid for the specimen. If residual specimen is available, repeat the extraction
  • procedure and repeat the test. If all markers remain negative after re-test, report the
  • results as invalid and a new specimen should be collected if possible.

  • 2019-nCoV Markers (N1 and N2)
  • • When all controls exhibit the expected performance, a specimen is considered negative if all
  • 2019-nCoV marker (N1, N2) cycle threshold growth curves DO NOT cross the threshold line
  • within 40.00 cycles (< 40.00 Ct) AND the RNase P growth curve DOES cross the threshold line
  • within 40.00 cycles (< 40.00 Ct).
  • • When all controls exhibit the expected performance, a specimen is considered positive for 2019-
  • nCoV if all 2019-nCoV marker (N1, N2) cycle threshold growth curves cross the threshold line
  • within 40.00 cycles (< 40.00 Ct). The RNase P may or may not be positive as described above, but
  • the 2019-nCoV result is still valid.
  • • When all controls exhibit the expected performance and the growth curves for the 2019-nCoV
  • markers (N1, N2) AND the RNase P marker DO NOT cross the cycle threshold growth curve within
  • 40.00 cycles (< 40.00 Ct), the result is invalid. The extracted RNA from the specimen should be retested. If residual RNA is not available, re-extract RNA from residual specimen and re-test. If the
  • re-tested sample is negative for all markers and RNase P, the result is invalid and collection of a
  • new specimen from the patient should be considered.
  • • When all controls exhibit the expected performance and the cycle threshold growth curve for any
  • one marker (N1 or N2, but not both markers) crosses the threshold line within 40.00 cycles (<
  • 40.00 Ct) the result is inconclusive. The extracted RNA should be retested. If residual RNA is not
  • available, re-extract RNA from residual specimen and re-test. If the same result is obtained, 
  • 36
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • report the inconclusive result. Consult with your state public health laboratory or CDC, as
  • appropriate, to request guidance and/or to coordinate transfer of the specimen for additional
  • analysis.
  • • If HSC is positive for N1 or N2, then contamination may have occurred during extraction or
  • sample processing. Invalidate all results for specimens extracted alongside the HSC. Re-extract
  • specimens and HSC and re-test.
  • 2019-nCoV rRT-PCR Diagnostic Panel Results Interpretation Guide
  • The table below lists the expected results for the 2019-nCoV rRT-PCR Diagnostic Panel. If a laboratory
  • obtains unexpected results for assay controls or if inconclusive or invalid results are obtained and cannot
  • be resolved through the recommended re-testing, please contact CDC for consultation and possible
  • specimen referral. See pages 13 and 50 for referral and contact information.
  • 2019
  • nCoV_N1
  • 2019
  • nCoV_N2 RP Result
  • Interpretationa Report Actions
  • + + ±
  • 2019-nCoV
  • detected Positive 2019-nCoV
  • Report results to CDC and
  • sender.
  • If only one of the two
  • targets is positive ±
  • Inconclusive
  • Result Inconclusive
  • Repeat testing of nucleic acid
  • and/or re-extract and repeat
  • rRT-PCR. If the repeated result
  • remains inconclusive, contact
  • your State Public Health
  • Laboratory or CDC for
  • instructions for transfer of the
  • specimen or further guidance.
  • - - +
  • 2019-nCoV not
  • detected Not Detected
  • Report results to sender.
  • Consider testing for other
  • respiratory viruses.b
  • - - - Invalid Result Invalid
  • Repeat extraction and rRT-PCR.
  • If the repeated result remains
  • invalid, consider collecting a
  • new specimen from the patient. aLaboratories should report their diagnostic result as appropriate and in compliance with their specific reporting
  • system.
  • bOptimum specimen types and timing for peak viral levels during infections caused by 2019-nCoV have not been
  • determined. Collection of multiple specimens from the same patient may be necessary to detect the virus. The
  • possibility of a false negative result should especially be considered if the patient’s recent exposures or clinical
  • presentation suggest that 2019-nCoV infection is possible, and diagnostic tests for other causes of illness (e.g.,
  • other respiratory illness) are negative. If 2019-nCoV infection is still suspected, re-testing should be considered
  • in consultation with public health authorities.
  • 37
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Quality Control
  • • Quality control requirements must be performed in conformance with local, state, and federal
  • regulations or accreditation requirements and the user’s laboratory’s standard quality control
  • procedures. For further guidance on appropriate quality control practices, refer to 42 CFR
  • 493.1256.
  • • Quality control procedures are intended to monitor reagent and assay performance.
  • • Test all positive controls prior to running diagnostic samples with each new kit lot to ensure all
  • reagents and kit components are working properly.
  • • Good laboratory practice (cGLP) recommends including a positive extraction control in each
  • nucleic acid isolation batch.
  • • Although HSC is not included with the 2019-nCov rRT-PCR Diagnostic Panel, the HSC extraction
  • control must proceed through nucleic acid isolation per batch of specimens to be tested.
  • • Always include a negative template control (NTC) and the appropriate positive control (nCoVPC)
  • in each amplification and detection run. All clinical samples should be tested for human RNase P
  • gene to control for specimen quality and extraction.
  • Limitations
  • • All users, analysts, and any person reporting diagnostic results should be trained to perform this
  • procedure by a competent instructor. They should demonstrate their ability to perform the test
  • and interpret the results prior to performing the assay independently.
  • • Performance of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel has only been established
  • in upper and lower respiratory specimens (such as nasopharyngeal or oropharyngeal swabs,
  • sputum, lower respiratory tract aspirates, bronchoalveolar lavage, and nasopharyngeal
  • wash/aspirate or nasal aspirate).
  • • Negative results do not preclude 2019-nCoV infection and should not be used as the sole basis
  • for treatment or other patient management decisions. Optimum specimen types and timing for
  • peak viral levels during infections caused by 2019-nCoV have not been determined. Collection of
  • multiple specimens (types and time points) from the same patient may be necessary to detect
  • the virus.
  • • A false-negative result may occur if a specimen is improperly collected, transported or handled.
  • False-negative results may also occur if amplification inhibitors are present in the specimen or if
  • inadequate numbers of organisms are present in the specimen.
  • • Positive and negative predictive values are highly dependent on prevalence. False-negative test
  • results are more likely when prevalence of disease is high. False-positive test results are more
  • likely when prevalence is moderate to low.
  • • Do not use any reagent past the expiration date.
  • • If the virus mutates in the rRT-PCR target region, 2019-nCoV may not be detected or may be
  • detected less predictably. Inhibitors or other types of interference may produce a false-negative
  • result. An interference study evaluating the effect of common cold medications was not
  • performed.
  • • Test performance can be affected because the epidemiology and clinical spectrum of infection
  • caused by 2019-nCoV is not fully known. For example, clinicians and laboratories may not know 
  • 38
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • the optimum types of specimens to collect, and, during the course of infection, when these
  • specimens are most likely to contain levels of viral RNA that can be readily detected.
  • • Detection of viral RNA may not indicate the presence of infectious virus or that 2019-nCoV is the
  • causative agent for clinical symptoms.
  • • The performance of this test has not been established for monitoring treatment of 2019-nCoV
  • infection.
  • • The performance of this test has not been established for screening of blood or blood products
  • for the presence of 2019-nCoV.
  • • This test cannot rule out diseases caused by other bacterial or viral pathogens.
  • Conditions of Authorization for the Laboratory
  • The CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel Letter of Authorization, along with the
  • authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for Patients, and authorized
  • labeling are available on the FDA website:
  • https://www.fda.gov/MedicalDevices/Safety/EmergencySituations/ucm161496.htm
  • Use of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel must follow the procedures outlined in
  • these manufacturer’s Instructions for Use and the conditions of authorization outlined in the Letter of
  • Authorization. Deviations from the procedures outlined are not permitted under the Emergency Use
  • Authorization (EUA). To assist clinical laboratories running the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel, the relevant Conditions of Authorization are listed verbatim below, and are required to
  • be met by laboratories performing the EUA test.
  • • Authorized laboratories1 will include with reports of the results of the CDC 2019-nCoV Real-Time
  • RT-PCR Diagnostic Panel, all authorized Fact Sheets. Under exigent circumstances, other
  • appropriate methods for disseminating these Fact Sheets may be used, which may include mass
  • media.
  • • Authorized laboratories will perform the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel as
  • outlined in the CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use. Deviations from the authorized procedures, including the authorized RT-PCR
  • instruments, authorized extraction methods, authorized clinical specimen types, authorized
  • control materials, authorized other ancillary reagents and authorized materials required to
  • perform the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel are not permitted. 2
  • • Authorized laboratories that receive the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel must
  • notify the relevant public health authorities of their intent to run the test prior to initiating
  • testing.
  •  1Authorized Laboratories: For ease of reference, the Letter of Authorization refers to “laboratories certified under the Clinical
  • Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. § 263a, to perform high complexity tests” as “authorized
  • laboratories.”
  • 2If an authorized laboratory is interested in implementing changes to the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • that are not in the scope (Section II) of this letter of authorization FDA recommends you discuss with FDA after considering
  • the policy outlined in Immediately in Effect Guidance for Clinical Laboratories and Food and Drug Administration Staff: Policy
  • for Diagnostics Testing in Laboratories Certified to Perform High Complexity Testing under CLIA prior to Emergency Use
  • Authorization for Coronavirus Disease-2019 during the Public Health Emergency
  • (https://www.fda.gov/media/135659/download).
  • 39
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • • Authorized laboratories will have a process in place for reporting test results to healthcare
  • providers and relevant public health authorities, as appropriate.
  • • Authorized laboratories will collect information on the performance of the test and report to
  • DMD/OHT7-OIR/OPEQ/CDRH (via email: CDRH-EUA-Reporting@fda.hhs.gov) and CDC
  • (respvirus@cdc.gov) any suspected occurrence of false-positive or false-negative results and
  • significant deviations from the established performance characteristics of the test of which they
  • become aware.
  • • Authorized laboratories will report adverse events, including problems with test performance or
  • results, to MedWatch by submitting the online FDA Form 3500
  • (https://www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home) or by
  • calling 1-800-FDA-1088
  • • All laboratory personnel using the test must be appropriately trained in RT-PCR techniques and
  • use appropriate laboratory and personal protective equipment when handling this kit and use
  • the test in accordance with the authorized labeling.
  • • CDC, IRR, manufacturers and distributors of commercial materials identified as acceptable on the
  • CDC website, and authorized laboratories will ensure that any records associated with this EUA
  • are maintained until otherwise notified by FDA. Such records will be made available to FDA for
  • inspection upon request.
  • Performance Characteristics
  • Analytical Performance:
  • Limit of Detection (LoD):
  • LoD studies determine the lowest detectable concentration of 2019-nCoV at which approximately 95%
  • of all (true positive) replicates test positive. The LoD was determined by limiting dilution studies using
  • characterized samples.
  • The analytical sensitivity of the rRT-PCR assays contained in the CDC 2019 Novel Coronavirus (2019-
  • nCoV) Real-Time RT-PCR Diagnostic Panel were determined in Limit of Detection studies. Since no
  • quantified virus isolates of the 2019-nCoV are currently available, assays designed for detection of the
  • 2019-nCoV RNA were tested with characterized stocks of in vitro transcribed full length RNA (N gene;
  • GenBank accession: MN908947.2) of known titer (RNA copies/µL) spiked into a diluent consisting of a
  • suspension of human A549 cells and viral transport medium (VTM) to mimic clinical specimen. Samples
  • were extracted using the QIAGEN EZ1 Advanced XL instrument and EZ1 DSP Virus Kit (Cat# 62724) and
  • manually with the QIAGEN DSP Viral RNA Mini Kit (Cat# 61904). Real-Time RT-PCR assays were
  • performed using the ThemoFisher Scientific TaqPath™ 1-Step RT-qPCR Master Mix, CG (Cat# A15299) on
  • the Applied Biosystems™ 7500 Fast Dx Real-Time PCR Instrument according to the CDC 2019-nCoV RealTime RT-PCR Diagnostic Panel instructions for use.
  • 40
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • A preliminary LoD for each assay was determined testing triplicate samples of RNA purified using each
  • extraction method. The approximate LoD was identified by extracting and testing 10-fold serial dilutions
  • of characterized stocks of in vitro transcribed full-length RNA. A confirmation of the LoD was determined
  • using 3-fold serial dilution RNA samples with 20 extracted replicates. The LoD was determined as the
  • lowest concentration where ≥ 95% (19/20) of the replicates were positive.
  • Table 4. Limit of Detection Confirmation of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • with QIAGEN EZ1 DSP
  • Targets 2019-nCoV_N1 2019-nCoV_N2
  • RNA Concentration1 10 0.5 10 0.0 10 -0.5 10 0.5 10 0.0 10 -0.5
  • Positives/Total 20/20 19/20 13/20 20/20 17/20 9/20
  • Mean Ct2 32.5 35.4 NA 35.8 NA NA
  • Standard Deviation
  • (Ct) 0.5 0.8 NA 1.3 NA NA
  • 1 Concentration is presented in RNA copies/µL
  • 2 Mean Ct reported for dilutions that are ≥ 95% positive. Calculations only include positive results.
  • NA not applicable
  • Table 5. Limit of Detection Confirmation CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel with
  • QIAGEN QIAmp DSP Viral RNA Mini Kit
  • Targets 2019-nCoV_N1 2019-nCoV_N2
  • RNA Concentration1 10 0.5 10 0.0 10 -0.5 10 0.5 10 0.0 10 -0.5 10 -1.0
  • Positives/Total 20/20 20/20 6/20 20/20 20/20 20/20 8/20
  • Mean Ct2 32.0 32.8 NA 33.0 35.4 36.2 NA
  • Standard Deviation
  • (Ct) 0.7 0.8 NA 1.4 0.9 1.9 NA
  • 1 Concentration is presented in RNA copies/µL
  • 2 Mean Ct reported for dilutions that are ≥ 95% positive. Calculations only include positive results.
  • NA not applicable
  • Table 6. Limit of Detection of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Virus Material
  • Limit of Detection (RNA copies/µL)
  • QIAGEN EZ1
  • Advanced XL
  • QIAGEN DSP Viral
  • RNA Mini Kit
  • 2019 Novel
  • Coronavirus
  • N Gene RNA
  • Transcript 100.5 100
  • FDA Sensitivity Evaluation: The analytical sensitivity of the test will be further assessed by evaluating an
  • FDA-recommended reference material using an FDA developed protocol if applicable and/or when
  • available. 
  • 41
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • In Silico Analysis of Primer and Probe Sequences:
  • The oligonucleotide primer and probe sequences of the CDC 2019 nCoV Real-Time RT-PCR Diagnostic
  • Panel were evaluated against 31,623 sequences available in the Global Initiative on Sharing All Influenza
  • Data (GISAID, https://www.gisaid.org) database as of June 20, 2020, to demonstrate the predicted
  • inclusivity of the 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Nucleotide mismatches in the
  • primer/probe regions with frequencies > 0.1% are shown below. With the exception of one nucleotide
  • mismatch with frequency > 1% (2.00%) at the third position of the N1 probe, the frequency of all
  • mismatches was < 1%, indicating that prevalence of the mismatches were sporadic. Only one sequence
  • (0.0032%) had two nucleotide mismatches in the N1 probe, and one other sequence from a different
  • isolate (0.0032%) had two nucleotide mismatches in the N1 reverse primer. No sequences were found to
  • have more than one mismatch in any N2 primer/probe region. The risk of these mismatches resulting in
  • a significant loss in reactivity causing a false negative result is extremely low due to the design of the
  • primers and probes, with melting temperatures > 60°C and with annealing temperature at 55°C that can
  • tolerate up to two mismatches.
  • Table 7. In Silico Inclusivity Analysis of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel Among
  • 31,623 Genome Sequences Available from GISAID as of June 20, 2020
  • Primer/probe N1 probe N1 reverse N2 probe
  • Location (5'>3') 3 15 21 13
  • Mismatch Nucleotide C>T G>T T>C C>T
  • Mismatch No. 632 34 71 46
  • Mismatch Frequency (%) 2.00 0.11 0.22 0.15
  • Specificity/Exclusivity Testing: In Silico Analysis
  • BLASTn analysis queries of the 2019-nCoV rRT-PCR assays primers and probes were performed against
  • public domain nucleotide sequences. The database search parameters were as follows: 1) The nucleotide
  • collection consists of GenBank+EMBL+DDBJ+PDB+RefSeq sequences, but excludes EST, STS, GSS, WGS,
  • TSA, patent sequences as well as phase 0, 1, and 2 HTGS sequences and sequences longer than 100Mb;
  • 2) The database is non-redundant. Identical sequences have been merged into one entry, while
  • preserving the accession, GI, title and taxonomy information for each entry; 3) Database was updated on
  • 10/03/2019; 4) The search parameters automatically adjust for short input sequences and the expect
  • threshold is 1000; 5) The match and mismatch scores are 1 and -3, respectively; 6) The penalty to create
  • and extend a gap in an alignment is 5 and 2 respectively.
  • 2019-nCoV_N1 Assay:
  • Probe sequence of 2019-nCoV rRT-PCR assay N1 showed high sequence homology with SARS
  • coronavirus and Bat SARS-like coronavirus genome. However, forward and reverse primers showed no
  • sequence homology with SARS coronavirus and Bat SARS-like coronavirus genome. Combining primers
  • and probe, there is no significant homologies with human genome, other coronaviruses or human
  • microflora that would predict potential false positive rRT-PCR results.
  • 42
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 2019-nCoV_N2 Assay:
  • The forward primer sequence of 2019-nCoV rRT-PCR assay N2 showed high sequence homology to Bat
  • SARS-like coronaviruses. The reverse primer and probe sequences showed no significant homology with
  • human genome, other coronaviruses or human microflora. Combining primers and probe, there is no
  • prediction of potential false positive rRT-PCR results.
  • In summary, the 2019-nCoV rRT-PCR assay N1 and N2, designed for the specific detection of 2019-nCoV,
  • showed no significant combined homologies with human genome, other coronaviruses, or human
  • microflora that would predict potential false positive rRT-PCR results.
  • In addition to the in silico analysis, several organisms were extracted and tested with the CDC 2019-nCoV
  • Real-Time RT-PCR Diagnostic Panel to demonstrate analytical specificity and exclusivity. Studies were
  • performed with nucleic acids extracted using the QIAGEN EZ1 Advanced XL instrument and EZ1 DSP
  • Virus Kit. Nucleic acids were extracted from high titer preparations (typically ≥ 105 PFU/mL or ≥ 106
  • CFU/mL). Testing was performed using the ThemoFisher Scientific TaqPath™ 1-Step RT-qPCR Master Mix,
  • CG on the Applied Biosystems™ 7500 Fast Dx Real-Time PCR instrument. The data demonstrate that the
  • expected results are obtained for each organism when tested with the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel.
  • 43
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Table 8. Specificity/Exclusivity of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Virus Strain Source
  • 2019-
  • nCoV_
  • N1
  • 2019-
  • nCoV_
  • N2
  • Final
  • Result
  • Human coronavirus 229E Isolate 0/3 0/3 Neg.
  • Human coronavirus OC43 Isolate 0/3 0/3 Neg.
  • Human coronavirus NL63 clinical specimen 0/3 0/3 Neg.
  • Human coronavirus HKU1 clinical specimen 0/3 0/3 Neg.
  • MERS-coronavirus Isolate 0/3 0/3 Neg.
  • SARS-coronavirus Isolate 0/3 0/3 Neg.
  • bocavirus - clinical specimen 0/3 0/3 Neg.
  • Mycoplasma pneumoniae Isolate 0/3 0/3 Neg.
  • Streptococcus Isolate 0/3 0/3 Neg.
  • Influenza A(H1N1) Isolate 0/3 0/3 Neg.
  • Influenza A(H3N2) Isolate 0/3 0/3 Neg.
  • Influenza B Isolate 0/3 0/3 Neg.
  • Human adenovirus, type 1 Ad71 Isolate 0/3 0/3 Neg.
  • Human metapneumovirus - Isolate 0/3 0/3 Neg.
  • respiratory syncytial virus Long A Isolate 0/3 0/3 Neg.
  • rhinovirus Isolate 0/3 0/3 Neg.
  • parainfluenza 1 C35 Isolate 0/3 0/3 Neg.
  • parainfluenza 2 Greer Isolate 0/3 0/3 Neg.
  • parainfluenza 3 C-43 Isolate 0/3 0/3 Neg.
  • parainfluenza 4 M-25 Isolate 0/3 0/3 Neg.
  • Endogenous Interference Substances Studies:
  • The CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel uses conventional well-established nucleic acid
  • extraction methods and based on our experience with CDC’s other EUA assays, including the CDC Novel
  • Coronavirus 2012 Real-time RT-PCR Assay for the presumptive detection of Middle East Respiratory
  • Syndrome Coronavirus (MERS-CoV) and the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic
  • Panel-Influenza A/H7 (Eurasian Lineage) Assay for the presumptive detection of novel influenza A (H7N9)
  • virus that are both intended for use with a number of respiratory specimens, we do not anticipate
  • interference from common endogenous substances.
  • Specimen Stability and Fresh-frozen Testing:
  • To increase the likelihood of detecting infection, CDC recommends collection of lower respiratory and
  • upper respiratory specimens for testing. If possible, additional specimen types (e.g., stool, urine) should
  • be collected and should be stored initially until decision is made by CDC whether additional specimen
  • sources should be tested. Specimens should be collected as soon as possible once a PUI is identified
  • regardless of symptom onset. Maintain proper infection control when collecting specimens. Store
  • specimens at 2-8°C and ship overnight to CDC on ice pack. Label each specimen container with the
  • patient’s ID number (e.g., medical record number), unique specimen ID (e.g., laboratory requisition
  • number), specimen type (e.g., nasal swabs) and the date the sample was collected. Complete a CDC
  • Form 50.34 for each specimen submitted.
  • 44
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Clinical Performance:

  • As of February 22, 2020, CDC has tested 2071 respiratory specimens from persons under investigation
  • (PUI) in the U.S. using the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Specimen types include
  • bronchial fluid/wash, buccal swab, nasal wash/aspirate, nasopharyngeal swab, nasopharyngeal/throat
  • swab, oral swab, sputum, oropharyngeal (throat) swab, swab (unspecified), and throat swab.
  • Table 9: Summary of CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel Data
  • Generated by Testing Human Respiratory Specimens Collected from PUI Subjects in the U.S.
  • Specimen Type
  • 2019 nCoV
  • Negative
  • 2019 nCoV
  • Positive Inconclusive Invalid Total
  • Bronchial
  • fluid/wash 2 0 0 0 2
  • Buccal swab 5 1 0 0 6
  • Nasal
  • wash/aspirate 6 0 0 0 6
  • Nasopharyngeal
  • swab 927 23 0 0 950
  • Nasopharyngeal
  • swab/throat
  • swab 4 0 0 0 4
  • Oral swab 476 9 0 0 485
  • Pharyngeal
  • (throat) swab 363 10 0 1 374
  • Sputum 165 5 0 0 170
  • Swab
  • (unspecified)1 71 1 0 0 72
  • Tissue (lung) 2 0 0 0 2
  • Total 2021 49 0 1 2071
  • 1Actual swab type information was missing from these upper respiratory tract specimens.
  • Two thousand twenty-one (2021) respiratory specimens of the 2071 respiratory specimens tested
  • negative by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Forty-nine (49) of the 2071
  • respiratory specimens tested positive by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. Only
  • one specimen (oropharyngeal (throat) swab) was invalid. Of the 49 respiratory specimens that tested
  • positive by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel, seventeen (17) were confirmed by
  • genetic sequencing and/or virus culture (positive percent agreement = 17/17, 95% CI: 81.6%-100%)
  • During the early phase of the testing, a total of 117 respiratory specimens collected from 46 PUI subjects
  • were also tested with two analytically validated real-time RT-PCR assays that target separate and
  • independent regions of the nucleocapsid protein gene of the 2019-nCoV, N4 and N5 assays. The
  • nucleocapsid protein gene targets for the N4 and N5 assays are different and independent from the
  • nucleocapsid protein gene targets for the two RT-PCR assays included in the CDC 2019-nCoV Real-Time
  • RT-PCR Diagnostic Panel, N1 and N2. Any positive result from the N4 and/or the N5 assay was further
  • investigated by genetic sequencing.
  • 45
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Performance of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel testing these 117 respiratory
  • specimens was estimated against a composite comparator. A specimen was considered comparator
  • negative if both the N4 and the N5 assays were negative. A specimen was considered comparator
  • positive when the N4 and/or the N5 assay generated a positive result, and the comparator positive
  • result(s) were further investigated and confirmed to be 2019-nCoV RNA positive by genetic sequencing.
  • Table 10: Percent Agreement of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel with the
  • Composite Comparator
  • CDC 2019-nCoV
  • Panel Result
  • Composite Comparator Result
  • Positive Negative
  • Positive 131 0
  • Inconclusive 0 0
  • Negative 0 104
  • 1Composite comparator results were available for 13 of 49 CDC 2019-nCoV Panel positive specimens
  • only.
  • Positive percent agreement = 13/13 = 100% (95% CI: 77.2% - 100%)
  • Negative percent agreement = 104/104 = 100% (95% CI: 96.4% - 100%)
  • Enzyme Master Mix Evaluation:
  • The limit of detection equivalence between the ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix and
  • the following enzyme master mixes was evaluated: Quantabio qScript XLT One-Step RT-qPCR ToughMix,
  • Quantabio UltraPlex 1-Step ToughMix (4X), and Promega GoTaq® Probe 1- Step RT-qPCR System. Serial
  • dilutions of 2019 novel coronavirus (SARS CoV-2) transcript were tested in triplicate with the CDC 2019-
  • nCoV Real-time RT-PCR Diagnostic Panel using all four enzyme master mixes. Both manufactured
  • versions of oligonucleotide probe, BHQ and ZEN, were used in the comparison. The lowest detectable
  • concentration of transcript at which all replicates tested positive using the Quantabio qScript XLT OneStep RT-qPCR ToughMix and Quantabio UltraPlex 1-Step ToughMix (4X) was similar to that observed for
  • the ThemoFisher TaqPath™ 1-Step RT-qPCR Master Mix. The lowest detectable concentration of
  • transcript when using the Promega GoTaq® Probe 1- Step RT-qPCR System was one dilution above that
  • observed for the other candidates when evaluated with the BHQ version of the CDC assays. The
  • candidate master mixes all performed equivalently or at one dilution below the ThemoFisher TaqPath™
  • 1-Step RT-qPCR Master Mix when evaluated with the ZEN version of the CDC assays. 
  • 46
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Table 11: Limit of Detection Comparison for Enzyme Master Mixes – BHQ Probe Summary Results
  • Copy Number
  • ThemoFisher TaqPath™
  • 1-Step RT-qPCR Master
  • Mix
  • Quantabio qScript XLT
  • One-Step RT-qPCR
  • ToughMix
  • Quantabio UltraPlex 1-
  • Step ToughMix (4X)
  • Promega GoTaq® Probe
  • 1- Step RT-qPCR System
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 102 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
  • 101 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
  • 100 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 2/3
  • 10-1 copies µL 2/3 0/3 1/3 1/3 1/3 1/3 0/3 0/3
  • Table 12: Limit of Detection Comparison for Enzyme Master Mixes – ZEN Probe Summary Results
  • Copy Number
  • ThemoFisher TaqPath™
  • 1-Step RT-qPCR Master
  • Mix
  • Quantabio qScript XLT
  • One-Step RT-qPCR
  • ToughMix
  • Quantabio UltraPlex 1-
  • Step ToughMix (4X)
  • Promega GoTaq® Probe
  • 1- Step RT-qPCR System
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 2019-
  • nCoV_N1
  • 2019-
  • nCoV_N2
  • 102 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
  • 101 copies/µL 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
  • 100 copies/µL 3/3 2/3 3/3 3/3 3/3 2/3 3/3 3/3
  • 10-1 copies µL 1/3 1/3 0/3 0/3 0/3 1/3 1/3 1/3
  • Retrospective positive (18) and negative (17) clinical respiratory specimens were extracted using the
  • QIAGEN EZ1 Advanced XL instrument and EZ1 DSP Virus Kit and were tested with the CDC 2019-nCoV
  • Real-time RT-PCR Diagnostic Panel using the Quantabio qScript XLT One-Step RT-qPCR ToughMix,
  • Quantabio UltraPlex 1-Step ToughMix (4X), and Promega GoTaq® Probe 1- Step RT-qPCR System master
  • mixes. All three enzyme master mixes performed equivalently, demonstrating 100% positive and 100%
  • negative agreement with expected results and a 95% confidence interval of 82.4%-100% and 81.6%-
  • 100%, respectively.
  • Table 13: Clinical Comparison – Retrospective Study Summary Results
  • CDC 2019-nCoV
  • Real-time RTPCR Diagnostic
  • Panel Result
  • Quantabio qScript XLT
  • One-Step RT-qPCR
  • ToughMix
  • Quantabio UltraPlex 1-Step
  • ToughMix (4X)
  • Promega GoTaq® Probe 1-
  • Step RT-qPCR System
  • Positive Negative Positive Negative Positive Negative
  • Positive 18 0 18 0 18 0
  • Negative 0 17 0 17 0 17
  • 47
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Roche MagNA Pure 24 and MagNA Pure 96 Extraction Platform Evaluation:
  • Performance of the 2019-CoV Real-time RT-PCR Diagnostic Panel using the Roche MagNA Pure 24 and
  • MagNA Pure 96 extraction platforms was compared to performance with an authorized extraction
  • method. Serial dilutions of quantified inactivated SARS-CoV-2 virus (USA-WA1/2020; 100 RNA copies/µL)
  • in lysis buffer were added to pooled negative upper respiratory tract specimen matrix. Five samples of
  • each dilution were extracted in parallel with the QIAGEN EZ1 Advanced XL (EZ1 DSP Virus Kit Cat# 62724)
  • and the Roche MagNA Pure 24 (MagNA Pure 24 Total NA Isolation Kit Cat# 07658036001) and Roche
  • MagNA Pure 96 (MagNA Pure 96 DNA and Viral Nucleic Acid Small Volume Kit Cat# 06543588001)
  • extraction platforms and evaluated using the 2019-nCoV Real-Time RT-PCR Diagnostic Panel and
  • ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix. The observed LoD was defined as the lowest
  • concentration at which 100% (5 out of 5 total) of all replicates tested positive for both primer/probe sets
  • (N1 and N2) in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. The acceptance criteria for
  • equivalence were defined as demonstrating an observed LoD either at the same endpoint or within a 3-
  • fold dilution. The results showed that both the MagNA Pure 24 and MagNA Pure 96 extraction platforms
  • performed equivalently or within one 3-fold dilution of the LoD observed when using the QIAGEN EZ1
  • Advanced XL extraction platform.
  • Table 14. Limit of Detection Comparison between the QIAGEN EZ1 Advanced XL, Roche MagNA Pure
  • 96, and Roche MagNA Pure 24 Extraction Platforms using the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel
  • Platform Parameter 2019-nCoV_N1 Assay 2019-nCoV_N2 Assay Observed
  • LoD1
  • QIAGEN EZ1
  • Advanced XL
  • RNA copies/µL 101.0 100.5 100.0 101.0 100.5 100.0
  • 100.5 # pos./total 5/5 5/5 5/5 5/5 5/5 3/5
  • Mean Ct2 34.0 35.0 36.3 33.9 36.6 NA
  • Std. Deviation 0.2 0.8 0.2 0.4 0.9 NA
  • Roche MagNA Pure
  • 96
  • RNA copies/µL 101.0 100.5 100.0 101.0 100.5 100.0
  • 100.5 # pos./total 5/5 5/5 5/5 5/5 5/5 2/5
  • Mean Ct2 33.3 34.6 36.1 33.2 35.7 NA
  • Std. Deviation 0.5 0.5 0.3 0.3 0.4 NA
  • Roche MagNA Pure
  • 24
  • RNA copies/µL 101.0 100.5 100.0 101.0 100.5 100.0
  • 101.0 # pos./total 5/5 3/5 3/5 5/5 5/5 5/5
  • Mean Ct2 34.4 NA NA 35.2 36.9 36.2
  • Std. Deviation 0.6 NA NA 0.5 1.0 0.8
  • 1
  • Concentration is presented in RNA copies/µL. The observed LoD is the lowest concentration where both assays showed 100%
  • positive detection.
  • 2
  • Mean Ct reported for dilutions that show 100% positivity. Calculations only include positive results.
  • NA = not applicable
  • Previously characterized clinical remainder specimens (14 positive and 15 negative) were extracted using
  • both the Roche MagNA Pure 96 and MagNA Pure 24 extraction platforms and evaluated using the 2019-
  • nCoV Real-Time RT-PCR Diagnostic Panel and ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix.
  • Acceptance criteria for clinical equivalence was defined as demonstrating 100% concurrence with
  • qualitative results shown with the authorized comparator method (QIAGEN EZ1 Advanced XL). Results
  • from this study showed 100% concurrence with the comparator method for both the Roche MagNA Pure 
  • 48
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • 96 and Roche MagNA Pure 24 extraction platforms when used with the CDC 2019-nCoV Real-Time RTPCR Diagnostic panel.
  • Table 15. Clinical Comparison Results – Retrospective Study Results
  • Test Platform
  • Test
  • Platform
  • Result
  • QIAGEN EZ1
  • Advanced XL Result Positive % Agreement (CI)1 Negative % Agreement (CI)1
  • Positive Negative
  • Roche MagNA
  • Pure 96
  • Positive 14 0 100.0 (78.5 – 100.0) 100.0 (79.6 – 100.0) Negative 0 15
  • Roche MagNA
  • Pure 24
  • Positive 14 0 100.0 (78.5 – 100.0) 100.0 (79.6 – 100.0) Negative 0 15 1 CI = 95% confidence interval
  • Promega Maxwell® RSC 48 Extraction Platform Evaluation:
  • Performance of the 2019-CoV Real-time RT-PCR Diagnostic Panel using the Promega Maxwell® RSC 48
  • extraction platform was compared to performance with an authorized extraction method. Serial
  • dilutions of quantified inactivated SARS-CoV-2 virus (USA-WA1/2020; 100 RNA copies/µL) in VTM were
  • added to pooled negative upper respiratory tract specimen matrix. Five samples of each dilution were
  • extracted in parallel with the QIAGEN EZ1® Advanced XL (EZ1 DSP Virus Kit Cat# 62724) and the Promega
  • Maxwell® RSC 48 (Promega Maxwell® Viral Total Nucleic Acid Purification Kit Cat# AS1330) extraction
  • platforms and evaluated using the 2019-nCoV Real-Time RT-PCR Diagnostic Panel and ThermoFisher
  • TaqPath™ 1-Step RT-qPCR Master Mix. The observed LoD was defined as the lowest concentration at
  • which 100% (5 out of 5 total) of all replicates tested positive for both primer/probe sets (N1 and N2) in
  • the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel. The acceptance criteria for equivalence were
  • defined as demonstrating an observed LoD either at the same endpoint or within a 3-fold dilution. The
  • results showed that the performance of the Maxwell® RSC 48 extraction platform performed
  • equivalently or within one 3-fold dilution of the LoD observed when using the QIAGEN EZ1® Advanced XL
  • extraction platform.
  • Table 16. Limit of Detection Comparison Between the QIAGEN EZ1® Advanced XL and Promega
  • Maxwell® RSC 48 Extraction Platforms Using the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Platform Parameter 2019-nCoV_N1 Assay 2019-nCoV_N2 Assay Observed
  • LoD1
  • QIAGEN EZ1®
  • Advanced XL
  • RNA copies/µL 100.5 100.0 10-0.5 100.5 100.0 10-0.5
  • 10 # pos./total 5/5 5/5 0/5 5/5 5/5 3/5 0.0
  • Mean Ct2 32.27 33.80 NA 35.13 36.41 NA
  • Std. Deviation 0.81 0.40 NA 0.81 0.40 NA
  • Promega Maxwell®
  • RSC 48
  • RNA copies/µL 100.5 100.0 10-0.5 100.5 100.0 10-0.5
  • 10 # pos./total 5/5 5/5 3/5 5/5 5/5 5/5 0.0
  • Mean Ct2 31.11 32.97 NA 31.89 33.95 35.17
  • Std. Deviation 0.24 0.34 NA 0.24 0.35 0.65
  • 1
  • Concentration is presented in RNA copies/µL. The observed LoD is the lowest concentration where both assays showed 100%
  • positive detection.
  • 2
  • Mean cycle threshold (Ct) reported for dilutions that show 100% positivity. Calculations only include positive results.
  • NA = not applicable
  • 49
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Previously characterized clinical remainder specimens (15 positive and 15 negative) were extracted using
  • the Promega Maxwell® RSC 48 extraction platform alongside the currently authorized QIAGEN EZ1®
  • Advanced XL extraction platform and evaluated using the 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • and ThermoFisher TaqPath™ 1-Step RT-qPCR Master Mix. Results from the Maxwell® RSC 48 were
  • compared with the QIAGEN EZ1® Advanced XL extraction performed in parallel showing 100% (15/15)
  • qualitative concurrence on positive samples and 93.3% (14/15) qualitative concurrence on negative
  • samples. This evaluation showed that two originally negative (QIAGEN QIAamp® DSP Viral RNA Mini Kit)
  • specimens (Specimens 16 and 24) yielded an inconclusive result after extraction using the QIAGEN EZ1®
  • Advanced XL. Repeat of the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel resolved one of the two
  • specimens (Specimen 24, negative result). The second specimen (Specimen 16) remained inconclusive.
  • Both these specimens yielded a negative result on the Maxwell® RSC 48.
  • Table 17. Clinical Comparison Results – Retrospective Study Results
  • Test Platform
  • Promega Maxwell® RSC 48 Positive %
  • Agreement (CI)1
  • Negative %
  • Agreement (CI)1
  • Result Positive Negative Inconclusive
  • QIAGEN EZ1®
  • Advanced XL
  • Positive 15 0 0
  • 100.0
  • (79.6-100.0)
  • 93.3
  • (70.2-98.9) Negative 0 14 0
  • Inconclusive 0 1 0
  • 1 CI = 95% confidence interval
  • Disposal
  • Dispose of hazardous or biologically contaminated materials according to the practices of your
  • institution.
  • References
  • 1. Ballew, H. C., et al. “Basic Laboratory Methods in Virology,” DHHS, Public Health Service 1975
  • (Revised 1981), Centers for Disease Control and Prevention, Atlanta, Georgia 30333.
  • 2. Clinical Laboratory Standards Institute (CLSI), “Collection, Transport, Preparation and Storage of
  • Specimens for Molecular Methods: Proposed Guideline,” MM13-A
  • 3. Lieber, M., et al. "A Continuous Tumor Cell Line from a Human Lung Carcinoma with Properties of
  • Type II Alveolar Epithelial Cells." International Journal of Cancer 1976, 17(1), 62-70.
  • 50
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Revision History
  • Revision # Effective Date Summary of Revisions
  • 1 February 4, 2020 Original Instructions for Use
  • 2 March 15, 2020 • Intended use update
  • • Removal of N3 primer and probe set from Diagnostic Panel
  • • Performance data update
  • • Addition of alternative nucleic acid extraction platforms
  • • Addition of acceptable alternatives to HSC and addition of
  • QIAGEN RUO extraction reagents
  • • Positive results no longer presumptive. No confirmation of
  • positive results required
  • 3 March 30, 2020 • Addition of alternative enzyme master mix options
  • 4 June 12, 2020 • Addition of MagNA Pure 24 extraction method
  • • Addition of performance data for the MagNA Pure 96
  • extraction method with SARS-CoV-2
  • • Addition of heat treatment alternative to specimen
  • extraction
  • • Addition of Roche and QIAGEN external lysis buffer
  • alternatives
  • • Acknowledgment of FDA policy permitting end users to
  • qualify alternative components without seeking an EUA or
  • EUA amendment
  • 5 July 13, 2020 • Addition of Promega Maxwell® RSC 48 extraction method
  • • Update to in silico inclusivity analyses
  • Contact Information, Ordering, and Product Support
  • For technical and product support, contact the CDC Division of Viral Diseases directly.
  • Send email to: respvirus@cdc.gov
  • Note: If your laboratory is using reagents sourced from someone other than the CDC International
  • Reagent Resource, please refer to the manufacturer’s instructions provided with the commercial
  • materials.
  • 51
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Appendix A: Heat Treatment Alternative to Extraction
  • UltraPlex 1-Step ToughMix (4X)
  • This procedure is only for use by public health laboratories.
  • Purpose:
  • In response to a global shortage of nucleic acid extraction reagents causing significant delays in testing,
  • the CDC has investigated the use of a heat treatment method requiring minimal reagents as a specimen
  • processing alternative to nucleic acid extraction for use with the 2019-nCoV Real-Time RT-PCR Diagnostic
  • Panel.
  • Where possible, laboratories should use qualified RNA or total nucleic acid extraction methods for
  • processing of specimens for subsequent testing by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic
  • Panel. Extraction removes inhibitory substances from specimens that could negatively impact PCR
  • performance.
  • This procedure for use of heat treatment for specimen processing is only recommended when a
  • shortage of qualified extraction reagents is a limiting factor in a laboratory’s ability to meet urgent
  • COVID-19 testing demand.
  • Precautions/Warnings/Limitations:
  • • CDC has evaluated this heat treatment process and has determined that this process is effective
  • for inactivation of SARS-CoV-2 in patient specimens.
  • • Performance was evaluated with only upper respiratory specimens. Heat treatment of lower
  • respiratory specimens for subsequent testing by the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel has not been evaluated.
  • • This procedure for heat treatment of specimens is only for use with the Quantabio UltraPlex 1-
  • Step ToughMix (4X).
  • • Heat treatment should only be conducted when a lab is ready to test the specimens by PCR.
  • Testing of heat-treated specimens must be conducted the same day.
  • Acceptable Specimens:
  • • Upper respiratory specimens
  • Note: Do not use heat treatment to process specimens that appear bloody or that contain
  • particulate matter. Such specimens should be extracted using a qualified RNA or TNA extraction
  • method prior to testing.
  • Materials Required (not provided):
  • • 70% ethanol
  • • 10% bleach, freshly prepared
  • • 96-well PCR reaction plates (Applied Biosystems catalog # 4346906, 4366932, 4346907, or
  • equivalent)
  • • Optical strip caps (Applied Biosystems 4323032, or equivalent)
  • • 1.5 mL Sarstedt tubes or equivalent
  • 52
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • • Aerosol resistant micropipette tips
  • • Micropipettes
  • • 96-well cold block
  • • Cold blocks for 1.5 mL - 2.0 mL tubes
  • • Vortex mixer
  • • 96-well plate centrifuge or equivalent
  • • Thermal cycler or equivalent
  • • Class II Biological Safety Cabinet (BSC)
  • Procedure:
  • Sample Preparation
  • 1) Decontaminate BSC with 10% bleach followed by 70% ethanol.
  • 2) If samples are frozen, thaw on ice or at 4°C. Wipe the outside of the sample tube with 70%
  • ethanol. Place thawed sample on cold rack or ice in BSC.
  • 3) Pulse vortex each sample and briefly spin down in a centrifuge to collect the liquid at the
  • bottom of the tube.
  • Heat Treatment
  • 1) Place a thermal cycler in the BSC, turn on, and program for 95°C for 1 min followed by 4°C
  • hold.
  • 2) Place a 96-well PCR plate onto a cold rack or ice in the BSC.
  • 3) Transfer 100 µL of each sample to the 96-well PCR plate and securely cap each well using
  • optical strip caps.
  • NOTE: Ensure that an HSC extraction control is included in each batch run as required under
  • CLIA.
  • 4) Place this 96-well PCR plate on the pre-heated thermal cycler and start run. Leave plate on
  • thermal cycler at 4°C, or place on ice or a cold block.
  • 5) Remove plate and centrifuge for 1 minute at 500 x g to pellet cellular debris.
  • 6) Place plate on a cold rack or ice and proceed to testing the supernatant by rRT-PCR.
  • 7) Testing of heat-treated specimens must be conducted the same day heat treatment is
  • performed. For long term storage, keep the original specimen at ≤-70°C.
  • Special Testing Considerations for Heat Treated Specimens:
  • • Enzyme Master Mix
  • Testing of specimens that have been processed with heat treatment should be conducted
  • with the Quantabio UltraPlex 1-Step ToughMix (4X), which demonstrated the best
  • performance with heat treated specimens. PCR testing of heat-treated specimens should
  • follow the instructions in the main body of this Instructions for Use document.
  • • Resolution of Inconclusive and Invalid Results
  • Retesting of heat-treated specimens that generated an inconclusive or invalid result must
  • include extraction of the original specimen with a qualified RNA or total nucleic acid (TNA)
  • extraction method, if available. Do not re-test the heat-treated specimen material to resolve
  • inconclusive or invalid test results.
  • 53
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Verification:
  • CDC recommends performance of verification studies for the heat treatment method prior to diagnostic
  • use that includes side-by-side preparation of a panel of positive and negative clinical specimens using a
  • qualified extraction method and this heat treatment method with subsequent testing by the CDC 2019-
  • nCoV Real-Time RT-PCR Diagnostic Panel.
  • Performance Characteristics:
  • Quantabio UltraPlex 1-Step ToughMix (4X)
  • Limit of Detection Comparison
  • Serial dilutions of inactivated SARS-CoV-2 [SARS-CoV-2 USA-WA1/2020] were prepared in simulated
  • specimen material (human A549 cells suspended in viral transport medium). Each concentration was
  • prepared side-by-side five times by both EZ1 extraction and by heat treatment. Each extracted or heattreated sample was subsequently tested by the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel using
  • the Quantabio UltraPlex 1-Step ToughMix (4X) on the Applied Biosystems 7500 Fast Dx instrument.
  • Observed detection was similar between the two specimen preparation methods.
  • Table B1: UltraPlex Limit of Detection Comparison between QIAGEN EZ1 Advanced XL extraction and heat
  • treatment (95°C for 1 min) method – Summary Results
  • Enzyme Platform Parameter 2019-nCoV_N1 Assay 2019-nCoV_N2 Assay Observed
  • LoD1
  • Quantabio UltraPlex 1-
  • Step ToughMix (4X)
  • 5 µL Template Addition
  • QIAGEN
  • EZ1
  • Advanced
  • XL
  • RNA copies/µL 101.0 100.5 100.0 10-0.5 10-1.0 101.0 100.5 100.0 10-0.5 10-1.0
  • 100.5 # pos./total 5/5 5/5 4/5 4/5 3/5 5/5 5/5 5/5 2/5 2/5
  • Mean Ct2 34.11 34.59 NA NA NA 32.97 33.76 34.70 NA NA
  • Std. Deviation 0.75 0.99 NA NA NA 0.33 0.72 0.98 NA NA
  • Heat
  • Treatment
  • 95°C
  • for 1 min
  • RNA copies/µL 101.0 100.5 100.0 10-0.5 10-1.0 101.0 100.5 100.0 10-0.5 10-1.0
  • 100.5 # pos./total 5/5 5/5 4/5 5/5 1/5 5/5 5/5 4/5 2/5 1/5
  • Mean Ct2 33.41 34.32 NA 36.73 NA 33.45 35.25 NA NA NA
  • Std. Deviation 0.62 0.40 NA 0.82 NA 0.40 0.80 NA NA NA
  • 1Concentration is presented in RNA copies/µL. The observed LoD is the lowest concentration where both assays showed 100%
  • positive detection.
  • 2Mean Ct reported for dilutions that show 100% positivity. Calculations only include positive results.
  • NA = not applicable
  • 54
  • CDC-006-00019, Revision: 05 CDC/DDID/NCIRD/ Division of Viral Diseases Effective: 07/13/2020
  • Clinical Comparison
  • A panel of 39 upper respiratory specimens were tested side-by-side using extraction with the Qiagen EZ1
  • extraction instrument and heat treatment. Extracted and heat-treated specimens were subsequently
  • tested with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel using the Quantabio UltraPlex 1-Step
  • ToughMix (4X). Qualitative results were compared to demonstrate agreement.
  • Table B2: Clinical Comparison Results Summary – Heat Treatment versus QIAGEN EZ1 Advanced XL
  • Test Result Heat Treatment Total
  • Positive %
  • Agreement (CI)1
  • Negative %
  • Agreement (CI)1
  • Positive Inconclusive Negative
  • QIAGEN EZ1
  • Advanced XL
  • Positive 18 1 0 19
  • Inconclusive 0 0 0 0 94.7 (75.4-99.1) 100 (83.9-100)
  • Negative 0 0 20 20
  • Total 18 1 20 39
  • 1 CI = 95% confidence interval
  • Questions and Comments:
  • If you have questions or comments about this procedure, please send by email to: respvirus@cdc.gov
  • Division of Viral Diseases / Respiratory Viruses Branch
  • CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
  • Requirements
  • *** DO NOT DISCARD: Important product-specific information ***
  • Document #: CDC-006-00005 Revision #: 05 Effective Date: 07/13/2020 Page 1 of 4
  • CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel –
  • Verification Requirements
  • Please consult the following guidance from the Centers for Medicare & Medicaid Services
  • (CMS) regarding diagnostic tests under Emergency Use Authorization (EUA):
  • https://www.cms.gov/Medicare/Provider-Enrollment-andCertification/SurveyCertificationGenInfo/Policy-and-Memos-to-States-and-RegionsItems/QSO18-19-CLIA
  • INTENDED USE
  • The CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is a realtime RT-PCR test intended for the qualitative detection of nucleic acid from the 2019-nCoV in
  • upper and lower respiratory specimens (such as nasopharyngeal or oropharyngeal swabs,
  • sputum, lower respiratory tract aspirates, bronchoalveolar lavage, and nasopharyngeal
  • wash/aspirate or nasal aspirate) collected from individuals who meet 2019-nCoV clinical
  • and/or epidemiological criteria (for example, clinical signs and symptoms associated with
  • 2019-nCoV infection, contact with a probable or confirmed 2019-nCoV case, history of travel
  • to a geographic locations where 2019-nCoV cases were detected, or other epidemiologic links
  • for which 2019-nCoV testing may be indicated as part of a public health investigation). Testing
  • in the United States is limited to laboratories certified under the Clinical Laboratory
  • Improvement Amendments of 1988 (CLIA), 42 U.S.C. § 263a, to perform high complexity
  • tests.
  • Results are for the identification of 2019-nCoV RNA. The 2019-nCoV RNA is generally
  • detectable in upper and lower respiratory specimens during infection. Positive results are
  • indicative of active infection with 2019-nCoV but do not rule out bacterial infection or coinfection with other viruses. The agent detected may not be the definite cause of disease.
  • Laboratories within the United States and its territories are required to report all positive
  • results to the appropriate public health authorities.
  • Negative results do not preclude 2019-nCoV infection and should not be used as the sole basis
  • for treatment or other patient management decisions. Negative results must be combined with
  • clinical observations, patient history, and epidemiological information.
  • Testing with the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel is intended for use by
  • trained laboratory personnel who are proficient in performing real-time RT-PCR assays. The
  • CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel is only for use
  • under a Food and Drug Administration’s Emergency Use Authorization.
  • REQUIRED MATERIALS
  • The 2019 novel coronavirus positive control (nCoVPC) is provided with the CDC 2019-nCoV
  • Real-Time RT-PCR Diagnostic Panel and should be prepared according to the Instructions for
  • Use. The nCoVPC consists of an RNA transcript of the 2019-nCoV N gene as well as human
  • RNase P gene segment. nCoVPC will yield a positive result with the following primer and probe
  • sets: 2019-nCoV_N1, 2019-nCoV_N2, and RP.
  • Approximately 2 mL of an upper respiratory specimen (e.g. nasopharyngeal swab (NPS) in
  • transport media) are needed for testing. Specimens may be pooled if less than 2 mL of one
  • specimen is available.
  • Refer to CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel package insert (manufacturer
  • instructions) for additional reagents, materials, and instructions.
  • PRECAUTIONS
  • This reagent should be handled in an approved biosafety level 2 (BSL-2) handling area to
  • avoid contamination of laboratory equipment and reagents that could cause false positive 
  • Division of Viral Diseases / Respiratory Viruses Branch
  • CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
  • Requirements
  • *** DO NOT DISCARD: Important product-specific information ***
  • Document #: CDC-006-00005 Revision #: 05 Effective Date: Page 2 of 4
  • results. This product is an RNA transcript and is non-infectious. However, the nCoVPC should
  • be handled in accordance with Good Laboratory Practices.
  • Store reagent at appropriate temperatures (see Instructions for Use) and hold on ice when
  • thawed.
  • Please use standard precautions when handling respiratory specimens.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE
  • QIAamp® DSP VIRAL RNA MINI KIT OR THE QIAamp® VIRAL RNA MINI KIT
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 560 µL of lysis buffer into each of nine tubes labeled 1-9.
  • • Add 140 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • • To prepare samples at a moderate concentration, spike 14 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 14 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 14 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE QIAGEN
  • EZ1® ADVANCED XL
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 280 µL of lysis buffer into each of nine tubes labeled 1-9.
  • • Add 120 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE ROCHE
  • MagNA PURE TOTAL NUCLEIC ACID KIT OR THE ROCHE MagNA PURE NUCLEIC ACID
  • ISOLATION KIT I
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 300 µL of lysis buffer into each of nine tubes labeled 1-9.
  • • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • Division of Viral Diseases / Respiratory Viruses Branch
  • CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
  • Requirements
  • *** DO NOT DISCARD: Important product-specific information ***
  • Document #: CDC-006-00005 Revision #: 05 Effective Date: Page 3 of 4
  • • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE ROCHE
  • MagNA PURE 24 AND TOTAL NUCLEIC ACID ISOLATION KIT
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 400 µL of lysis buffer into each of nine tubes labeled 1-9.
  • • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE ROCHE
  • MagNA PURE 96 DNA AND VIRAL NA SMALL VOLUME KIT
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 350 µL of lysis buffer into each of nine tubes labeled 1-9.
  • • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE
  • PROMEGA MAXWELL® RSC 48
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 330 µL of lysis buffer (300 µL of lysis buffer + 30 µL Proteinase K, included in the
  • kit) into each of nine tubes labeled 1-9.
  • Division of Viral Diseases / Respiratory Viruses Branch
  • CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel – Verification
  • Requirements
  • *** DO NOT DISCARD: Important product-specific information ***
  • Document #: CDC-006-00005 Revision #: 05 Effective Date: Page 4 of 4
  • • Add 120 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • INSTRUCTIONS FOR PREPARING SAMPLES BEFORE EXTRACTION WITH THE
  • BIOMÉRIEUX NucliSENS easyMAG OR THE BIOMÉRIEUX EMAG
  • • Refer to the 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for
  • reconstitution of the materials for use. RNA should be kept cold during preparation and use.
  • • Make a 1/10 dilution of nCoVPC by adding 5 µL of nCoVPC into 45 µL of nuclease-free water
  • or 10 mM Tris.
  • • Aliquot 1000 μL or 2000 µL of pre-aliquoted easyMAG lysis buffer into each of nine tubes
  • labeled 1-9 for the easyMAG or eMAG, respectively.
  • • Add 100 µL of upper respiratory specimen (e.g. NPS in viral transport media) into each of
  • the nine labeled tubes with lysis buffer.
  • • To prepare samples at a moderate concentration, spike 12 µL of undiluted nCoVPC
  • (rehydrated as described in the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel
  • Instructions for Use) into each tube labeled 1-3 containing lysis buffer and specimen.
  • • To prepare samples at a low concentration, spike 12 µL of 1/10 dilution of nCoVPC into
  • each tube labeled 4-6 containing lysis buffer and specimen.
  • • To prepare negative samples, spike 12 µL of nuclease-free water into each tube labeled 7-9
  • containing lysis buffer and specimen.
  • • Perform extractions of all nine samples according to the CDC 2019-nCoV Real-Time RT-PCR
  • Diagnostic Panel Instructions for Use.
  • PROCEDURE
  • Follow the CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel Instructions for Use for testing
  • the nine extracted samples at least once.
  • EXPECTED RESULTS
  • Moderate nCoVPC samples should be positive for 2019-nCoV.
  • Low nCoVPC samples should be positive for 2019-nCoV.
  • Negative upper respiratory samples should be negative for 2019-nCoV.
  • ≥90% of test results should be in agreement with the expected results. If test results are
  • <90% in agreement with expected results, contact CDC at respvirus@cdc.gov.
  • QUESTIONS
  • Please send questions or comments by email to respvirus@cdc.gov.
  • DISTRIBUTION
  • Distributed to qualified laboratories by Centers for Disease Control and Prevention, 1600
  • Clifton Road, Atlanta, GA, 30329 USA